Dynamic simulation of the natural and replaced human ankle joint.

Disappointing results for total ankle replacement have been explained by poor knowledge of the mechanics of the intact and replaced joints. Dynamic simulation tools have the capacity to simulate dynamic conditions that occur in human joints. The Working Model 2D tool was used to simulate the mechanics of the intact and replaced ankle joints, based on previously validated mathematical models. Elementary objects were used to model ligaments, articular surfaces, retinacula and muscle-tendon units. The performance of several pairs of prosthetic articular surfaces was also analysed. According to the results of these simulations, rolling as well as sliding motion occurs in the natural ankle, governed by a ligamentous linkage. Elongation of the tibiocalcaneal and calcaneofibular ligaments was found to be 1.5% and 4.8%, respectively. A 13% change in lever arm length occurred for both the tibialis anterior and gastrocnemius muscles during ankle flexion. Unlike the currently available three-component designs, the newly proposed convex-tibial ligament-compatible prosthesis was found to be able to restore the original mobility and physiological function of the ligaments. This prosthesis combines freedom from restraint with congruity of the components throughout the range of flexion.

Ankle Joint↗

Computer-assisted design of the sagittal shapes of a ligament-compatible total ankle replacement.

The poor results of total ankle replacement have been attributed to the inability of designers to restore adequately the critical mutual function of the ligaments and the articular surfaces. The purpose of this study was to design sagittal shapes of the articular surfaces for a new ankle prosthesis to be compatible with the geometry of the retained ligamentous structures. Several ligament-compatible pairs of articular surfaces were tested using a computerised version of a four-bar linkage model. The kinematics of the ankle when replaced by non-conforming two-component and by fully conforming three-component designs with either flat, concave or convex tibial surfaces were assessed by the model. A ligament-compatible convex-tibia fully-congruent three-component prosthesis showed the best features. The three-component prosthesis allows complete congruence over the entire range of flexion. A convex shape for the tibial arc was preferred because of the better degree of entrapment of the meniscal bearing. A 5 cm convex-tibia arc radius gave 2 mm entrapment together with 9.8 mm of tibial bone cut. Ligament elongation imposed by full congruence of the articular surfaces was less than 0.03% of the original length. The original patterns of joint kinematics and ligament tensioning are closely restored in the joint replaced by the proposed prosthesis.

Ankle Joint↗

Mobility of the subtalar joint in the intact ankle complex.

A previous study by these authors showed that the calcaneus follows a unique path of unresisted coupled motion relative to the tibia during passive flexion and that most of this motion occurred at the ankle level. Subtalar motion in the intact ankle complex was observed only when perturbations from this path were induced by the application of force to the calcaneus. Relative motion of the bones of the ankle complex was tracked by stereophotogrammetry in seven specimens. Anatomical landmarks, reference frames and joint angles were defined by standard techniques. Sequential moments were applied to the calcaneus about the long axis of the tibia. Measured movements at subtalar level demonstrated plantarflexion coupled to supination and internal rotation (inversion) and dorsiflexion coupled to pronation and external rotation (eversion). These movements were resisted and were fully recovered when the external load was removed. Subtalar motion diminished as the ankle approached maximal dorsi- and plantarflexion. Two clearly distinguished mean axes of rotation were observed for inversion and eversion runs. The axes of inversion and eversion of the subtalar complex changed orientation along a preferred and repeatable path. The subtalar joint complex occupied only a single stable position in the unloaded state and with no range of unresisted motion. It is inferred that mobility was possible only by the stretching and lengthening of the ligaments and the indentation of the articular surfaces, requiring the application of loads. The subtalar joint complex behaves like a flexible structure.

Ankle Joint↗

Geometry and mechanics of the human ankle complex and ankle prosthesis design.

The main objective of the study was to develop a model of the intact human ankle complex. It was also aimed at designing total ankle replacement which would better reproduce the physiological function of the joint. Passive flexion was analysed in seven lower-leg preparations with a stereophotogrammetric system. The articular surfaces and fibres within the calcaneofibular and tibiocalcaneal ligaments prescribed the changing positions of the axis of rotation. Joint motion included rolling as well as sliding. A computer-based model elucidated the observed kinematics at the intact joint. The experimental evidence and the geometrical model gave the basis for the design of models of replaced ankle in the sagittal plane. A three-component, convex-tibia prosthesis was eventually selected with articular surface shapes compatible with the geometry of the ligaments. It was demonstrated that in intact ankle joint, the geometry of the articular surfaces is strictly related to that of the ligaments and that current prosthesis designs do not restore physiological pattern of ligament tensioning. Careful reconstruction of the ligaments is recommended in any ankle surgery for maintenance of the normal kinematics and mechanics. A proposed novel design based on ligament/shape compatibility may improve total ankle replacement results.

Ankle Joint↗

Integrated pressure-force-kinematics measuring system for the characterisation of plantar foot loading during locomotion.

Plantar pressure, ground reaction force and body-segment kinematics measurements are largely used in gait analysis to characterise normal and abnormal function of the human foot. The combination of all these data together provides a more exhaustive, detailed and accurate view of foot loading during activities than traditional measurement systems alone do. A prototype system is presented that integrates a pressure platform, a force platform and a 3D anatomical tracking system to acquire combined information about foot function and loading. A stereophotogrammetric system and an anatomically based protocol for foot segment kinematics is included in a previously devised piezo-dynamometric system that combines pressure and force measurements. Experimental validation tests are carried out to check for both spatial and time synchronisation. Misalignment of the three systems is found to be within 6.0, 5.0 and 1.5 mm for the stereophotogrammetric system, force platform and pressure platform, respectively. The combination of position and pressure data allows for a more accurate selection of plantar foot subareas on the footprint. Measurements are also taken on five healthy volunteers during level walking to verify the feasibility of the overall experimental protocol. Four main subareas are defined and identified, and the relevant vertical and shear force data are computed. The integrated system is effective when there is a need for loading measurements in specific plantar foot subareas. This is attractive both in clinical assessment and in biomechanics research.

Biomechanical Phenomena↗

Effects of hip joint centre mislocation on gait analysis results.

Methods to determine the hip joint centre (HJC) location are necessary in gait analysis. It has been demonstrated that the methods proposed in the literature involve large mislocation errors. The choice should be made according to the extent by which HJC location errors distort the estimates of angles and resultant moments at the hip and knee joints. This study aimed at quantifying how mislocation errors propagate to these gait analysis results. Angles and moments at the hip and knee joint were calculated for five able-bodied subjects during level walking. The nominal position of the HJC was determined as the position of the pivot point of a 3D movement of the thigh relative to the pelvis. Angles and moments were then re-calculated after having added to HJC co-ordinates errors in the range of +/-30 mm. Angles and moments at both hip and knee joints were affected by HJC mislocation. The hip moments showed the largest propagation error: a 30 mm HJC anterior mislocation resulted in a propagated error into flexion/extension component of about -22%. The hip abduction/adduction moment was found the second largest affected quantity: a 30 mm lateral HJC mislocation produced a propagated error of about -15%. Finally, a 30 mm posterior HJC mislocation produced a delay of the flexion-to-extension timing in the order of 25% of the stride duration. HJC estimation methods with minimum antero-posterior error should therefore be preferred.

Adult↗

Cruciate ligament forces in the human knee during rehabilitation exercises.

OBJECTIVE: To determine the cruciate ligament forces occurring during typical rehabilitation exercises.Design. A combination of non-invasive measurements with mathematical modelling of the lower limb.Background. Direct measurement of ligament forces has not yet been successful in vivo in humans. A promising alternative is to calculate the forces mathematically. METHODS: Sixteen subjects performed isometric and isokinetic or squat exercises while the external forces and limb kinematics were measured. Internal forces were calculated using a geometrical model of the lower limb and the "dynamically determinate one-sided constraint" analysis procedure. RESULTS: During isokinetic/isometric extension, peak anterior cruciate ligament forces, occurring at knee angles of 35-40 degrees, may reach 0.55x body-weight. Peak posterior cruciate ligament forces are lower and occur around 90 degrees. During isokinetic/isometric flexion, peak posterior cruciate forces, which occur around 90 degrees, may exceed 4x body-weight; the anterior cruciate is not loaded. During squats, the anterior cruciate is lightly loaded at knee angles up to 50 degrees, after which the posterior cruciate is loaded. Peak posterior cruciate forces occur near the lowest point of the squat and may reach 3.5x body-weight. CONCLUSIONS: For anterior cruciate injuries, squats should be safer than isokinetic or isometric extension for quadriceps strengthening, though isokinetic or isometric flexion may safely be used for hamstrings strengthening. For posterior cruciate injuries, isokinetic extension at knee angles less than 70 degrees should be safe but isokinetic flexion and deep squats should be avoided until healing is well-advanced. RELEVANCE: Good rehabilitation is vital for a successful outcome to cruciate ligament injuries. Knowledge of ligament forces can aid the physician in the design of improved rehabilitation protocols.

Adult↗

The role of the passive structures in the mobility and stability of the human ankle joint: a literature review.

The mobility and stability of the ankle joint have been extensively investigated, but many critical important issues still need to be elucidated. However, there seems to be a general agreement on several important observations. A more isometric pattern of rotation for the calcaneofibular and the tibiocalcaneal ligaments with respect to all the others has been reported. Many recent studies have found changing positions of the instantaneous axis of rotation, suggesting that the hinge joint concept is an oversimplification for the ankle joint. A few recent works have also claimed anterior shift of the contact area at the tibial mortise during dorsiflexion, which would imply combined rolling and sliding motion at this joint. Many findings from the literature support the view of a close interaction between the geometry of the ligaments and the shapes of the articular surfaces in guiding and stabilizing motion at the ankle joint.

Ankle Joint↗

Validation of a functional method for the estimation of hip joint centre location.

The present study assesses the accuracy with which the subject specific coordinates of the hip joint centre (HJC) in a pelvic anatomical frame can be estimated using different methods. The functional method was applied by calculating the centre of the best sphere described by the trajectory of markers placed on the thigh during several trials of hip rotations. Different prediction methods, proposed in the literature and in the present investigation, which estimate the HJC of adult subjects using regression equations and anthropometric measurements, were also assessed. The accuracy of each of the above-mentioned methods was investigated by comparing their predictions with measurements obtained on a sample of 11 male adult able-bodied volunteers using roentgen stereophotogrammetric analysis (RSA), assumed to provide the true HJC locations. Prediction methods estimated the HJC location at an average rms distance of 25-30 mm. The functional method performed significantly better and estimated HJCs within a rms distance of 13 mm on average. This result may be confidently generalised if the photogrammetric experiment is carefully conducted and an optimal analytical approach used. The method is therefore suggested for use in motion analysis when the subject's hip range of motion is not limited. In addition, the facts that it is not an invasive technique and that it has relatively small and un-biased errors, make it suitable for regression equations identification with no limit to sample size and population typology.

Adult↗

Kinematics of the human ankle complex in passive flexion; a single degree of freedom system.

The restoration of original range and pattern of motion is the primary goal of joint replacement and ligament reconstruction. The objective of the present work is to investigate whether or not a preferred path of joint motion at the intact human ankle complex is exhibited during passive flexion. A rig was built to move the ankle complex through its range of flexion while applying only the minimum necessary load to drive ankle flexion. Joint motion was constrained only by the articular surfaces and the ligaments. The movements of the calcaneus, talus and fibula relative to the stationary tibia in seven cadaveric specimens were tracked with a stereophotogrammetric system. It was shown that the calcaneus follows a unique path of unresisted coupled motion relative to the tibia and that most of the motion occurred at the ankle, with little motion at the subtalar level. The calcaneofibular and the tibiocalcaneal ligaments showed near-isometric pattern of rotations. All specimens showed motion of the axis of rotation relative to the bones. Deviations from the unique path due to the application of load involved mostly subtalar motion and were resisted. The ankle complex exhibits one degree of unresisted freedom, the ankle behaving as a single degree of freedom mechanism and the subtalar as a flexible structure. We deduced that the calcaneofibular and tibiocalcaneal ligaments together with the articular surfaces guide ankle passive motion, other ligaments limit but do not guide motion.

Ankle Joint↗

A geometric model of the human ankle joint.

A two-dimensional four-bar linkage model of the ankle joint is formulated to describe dorsi/plantarflexion in unloaded conditions as observed in passive tests on ankle complex specimens. The experiments demonstrated that the human ankle joint complex behaves as a single-degree-of-freedom system during passive motion, with a moving axis of rotation. The bulk of the movement occurred at the level of the ankle. Fibres within the calcaneofibular and tibiocalcaneal ligaments remained approximately isometric. The experiments showed that passive kinematics of the ankle complex is governed only by the articular surfaces and the ligaments. It was deduced that the ankle is a single-degree-of-freedom mechanism where mobility is allowed by the sliding of the articular surfaces upon each other and the isometric rotation of two ligaments about their origins and insertions, without tissue deformation. The linkage model is formed by the tibia/fibula and talus/calcaneus bone segments and by the calcaneofibular and tibiocalcaneal ligament segments. The model predicts the path of calcaneus motion, ligament orientations, instantaneous axis of rotation, and conjugate talus surface profile as observed in the experiments. Many features of ankle kinematics such as rolling and multiaxial rotation are elucidated. The geometrical model is a necessary preliminary step to the study of ankle joint stability in response to applied loads and can be used to predict the effects of changes to the original geometry of the intact joint. Careful reconstruction of the original geometry of the ligaments is necessary after injury or during total ankle replacement.

Ankle Injuries↗

An anatomically based protocol for the description of foot segment kinematics during gait.

OBJECTIVE: To design a technique for the in vivo description of ankle and other foot joint rotations to be applied in routine functional evaluation using non-invasive stereophotogrammetry. DESIGN: Position and orientation of tibia/fibula, calcaneus, mid-foot, 1st metatarsal and hallux segments were tracked during the stance phase of walking in nine asymptomatic subjects. Rigid clusters of reflective markers were used for foot segment pose estimation. Anatomical landmark calibration was applied for the reconstruction of anatomical landmarks. BACKGROUND: Previous studies have analysed only a limited number of joints or have proposed invasive techniques. METHODS: Anatomical landmark trajectories were reconstructed in the laboratory frame using data from the anatomical calibration procedure. Anatomical co-ordinate frames were defined using the obtained landmark trajectories. Joint co-ordinate systems were used to calculate corresponding joint rotations in all three anatomical planes. RESULTS: The patterns of the joint rotations were highly repeatable within subjects. Consistent patterns between subjects were also exhibited at most of the joints. CONCLUSION: The method proposed enables a detailed description of ankle and other foot joint rotations on an anatomical base. Joint rotations can therefore be expressed in the well-established terminology necessary for their clinical interpretation. RELEVANCE: Functional evaluation of patients affected by foot diseases has recently called for more detailed and non-invasive protocols for the description of foot joint rotations during gait. The proposed method can help clinicians to distinguish between normal and pathological pattern of foot joint rotations, and to quantitatively assess the restoration of normal function after treatment.

Adult↗

A model-based method for the reconstruction of total knee replacement kinematics.

A better knowledge of the kinematics behavior of total knee replacement (TKR) during activity still remains a crucial issue to validate innovative prosthesis designs and different surgical strategies. Tools for more accurate measurement of in vivo kinematics of knee prosthesis components are therefore fundamental to improve the clinical outcome of knee replacement. In the present study, a novel model-based method for the estimation of the three-dimensional (3-D) position and orientation (pose) of both the femoral and tibial knee prosthesis components during activity is presented. The knowledge of the 3-D geometry of the components and a single plane projection view in a fluoroscopic image are sufficient to reconstruct the absolute and relative pose of the components in space. The technique is based on the best alignment of the component designs with the corresponding projection on the image plane. The image generation process is modeled and an iterative procedure localizes the spatial pose of the object by minimizing the Euclidean distance of the projection rays from the object surface. Computer simulation and static/dynamic in vitro tests using real knee prosthesis show that the accuracy with which relative orientation and position of the components can be estimated is better than 1.5 degrees and 1.5 mm, respectively. In vivo tests demonstrate that the method is well suited for kinematics analysis on TKR patients and that good quality images can be obtained with a carefully positioning of the fluoroscope and an appropriate dosage. With respect to previously adopted template matching techniques, the present method overcomes the complete segmentation of the components on the projected image and also features the simultaneous evaluation of all the six degrees of freedom (DOF) of the object. The expected small difference between successive poses in in vivo sequences strongly reduces the frequency of false poses and both the operator and computation time.

Algorithms↗

Analysis of function after intra-articular fracture of the os calcis.

Fractures of the os calcis with intra-articular involvement can cause difficulty in treatment and can result in abnormal function throughout the ankle-foot complex. We evaluated the function of the ankle-foot complex and the overall gait pattern after fracture of the os calcis, and we compared the results of surgical and nonsurgical treatment. Those in whom the geometry of the os calcis and joint was restored by reconstructive surgery had better compensation of gait and a better clinical-functional score. Complex disturbances in gait were found in the group of patients that did not undergo open reduction or internal fixation.

Adult↗

Data management in gait analysis for clinical applications.

OBJECTIVE: To study the reliability of gait analysis data obtained using the Calibrated Anatomical System Technique (CAST) protocol and to verify the suitability and repeatability of the extraction of a number of parameters from the waveforms obtained. DESIGN: The experimental protocol and the parametric analysis technique were applied on a population of able-bodied subjects. BACKGROUND: The clinical interpretation process of gait data still needs a more accurate analysis of the reliability and repeatability of the measurements and a suitable procedure for data reduction useful for data comparison. METHODS: Gait analysis was performed in 20 able-bodied subjects using a stereophotogrammetric system and a forceplate. 124 parameters relative to time-distance, kinematic and kinetic variables were calculated by means of an automatic procedure and statistically analysed. RESULTS: Most of the parameters were found to be normally distributed with relatively small range of variation. Few of them showed poor repeatability, mostly due to the experimental inaccuracies introduced. Correlation of several gait parameters with age, sex, and speed of progression was also identified.

Journal Article↗

Application of stereophotogrammetry to total body three-dimensional analysis of human tremor.

This work aims at verifying the possibility of investigating joint rotations, with amplitude and frequency ranges typical of pathological human tremor, using commercially available stereophotogrammetric systems together with signal processing techniques. A rotating disk is used as a mechanical tremor simulator to test the ability of the system to track known marker trajectories both in a large and a small calibrated volume. The performances of standard discrete Fourier transform (DFT) and autoregressive techniques are also evaluated and compared in the signal spectrum estimation. Results obtained from a pathological test subject and their dependence on the processing techniques adopted are also presented. The use of stereophotogrammetry and of the proposed signal spectrum estimation technique allow to quantify both frequency and amplitude content of three-dimensional (3-D) rotations of many human joints simultaneously and therefore to isolate the contribution of each joint to the whole body tremor.

Aged↗

Optimization and smoothing techniques in movement analysis.

This paper deals with accurate estimation of the degrees of freedom (DOF) of a body segment, starting from the trajectories of clusters of markers acquired with a suitable measurement system. The most commonly employed estimation procedures involve two sequential steps, a trajectory smoothing algorithm and a DOF reconstruction routine. Three optimized smoothing and reconstruction schemes are described, analyzed and tested and their performances are compared with each other and with those of a more traditional technique which contains no optimization criteria. All three schemes include an iterative, weighted-least-squares DOF reconstruction routine and a self-tuning, zero-phase-shift, 4th-order Butterworth filter. Both routines are extensively described and validated on the basis of numerically-simulated marker trajectories. Test results, analyzed on a statistical basis, show that the use of an optimization routine provides a visible improvement in DOF reconstruction. This performance has also been confirmed using stereophotogrammetric data collected on a subject wearing an external fracture fixation device which provides reference values for the bone DOF. Angular DOF estimated applying the optimized method to skin technical clusters are much closer to the reference values than the non-optimized values. Smoothing of data further improves the reconstruction accuracy while a far less crucial role is played by the order in which smoothing and reconstruction routines are applied.

Algorithms↗

Position and orientation in space of bones during movement: experimental artefacts.

This paper deals with the experimental problems related to the reconstruction of the position and orientation of the lower limb bones in space during the execution of locomotion and physical exercises. The inaccuracies associated with the relative movement between markers and underlying bone are analysed. Quantitative information regarding this movement was collected by making experiments on subjects who had suffered fractures and were wearing either femoral or tibial external fixators. These latter devices provided frames that were reliably rigid with the bone involved, and hence the possibility of assessing the relative movement between markers mounted on the skin and this bone. Anatomical frames associated with thigh and shank were reconstructed using technical frames based on different clusters of skin markers and their rotation with respect to the relevant bone evaluated. Marker movement was also assessed in subjects with intact musculoskeletal structures using digital videofluoroscopy.

Journal Article↗