PubMed Health⌕ Search

Biomedical subjects

A R van der Wiel

Publications and source records attributed to A R van der Wiel.

3 recordsLinked to original sources

Pulse oximetry and transcutaneous oxygen tension in hypoxemic neonates and infants with bronchopulmonary dysplasia.

The reliability of pulse oximetry and transcutaneous oxygen tension (tcPO2) was investigated in hypoxemic neonates and older infants with chronic hypoxemia due to bronchopulmonary dysplasia (BPD). It was found that during severe hypoxemia (tcPO2 less than 40 mmHg and saturation less than 80%) pulse oximetry showed a better correlation with arterial saturation than tcPO2 with arterial oxygen tension. During mild hypoxemia and normoxemia (tcPO2 40-90 mmHg and saturation 80-95%) tcPO2 and pulse oximetry both showed a good correlation with arterial values. Above 95% saturation and a corresponding tcPO2 of 70-120 mmHg, the correlation between arterial and transcutaneous PO2 was better than that between pulse oximetric and arterial saturation. Computer recording and analysis of tcPO2 and pulse oximetry improves the quality of both noninvasive oxygenation parameters in older infants with BPD.

Birth Weight↗

Multichannel recording and analysis of physiological data using a personal computer.

A flexible data acquisition system was designed for an Apple IIe computer. For implementation, we used a microcomputer capable of recording 4 channels simultaneously with a selectable sample rate; this information was saved on a floppy disc and the recordings could be evaluated later. This evaluation is accomplished by reproducing the analog signals on display and calculating a histogram after deleting artefacts in the signal.

Blood Gas Monitoring, Transcutaneous↗