PubMed Health⌕ Search

Biomedical subjects

A Schuchert

Publications and source records attributed to A Schuchert.

At least 19 recordsLinked to original sources

Adjustment of the evoked response sensitivity after hospital discharge in pacemaker patients with automatic ventricular threshold tracking activated.

Automatic threshold tracking in cardiac pacemakers allows ventricular capture verification and self-adaptive pacing output regulation. The Autocapture algorithm detects the evoked response (ER) signal immediately after the pacing pulse to verify the efficacy of ventricular pacing. Before hospital delivery, the ER sensitivity must be programmed individually so that the pacemaker detects the ER signal adequately without sensing lead polarization. The aims of the study were to assess the frequency of patients in whom Autocapture could be activated and whether the ER sensitivity had to be adjusted after hospital discharge. The study included 44 patients who received the VVIR pacemaker Regency SR+ (St. Jude Medical) connected to the model 1450 T pacing lead. ER signal, lead polarization, and ER sensitivity were evaluated before hospital discharge and 1, 3, and 6 months after implantation. The system recommended activating Autocapture in 42 of 44 patients. The mean ER signal was 8.4+/-1.2 mV at discharge, 9.0+/-3.9 mV at month 1, 8.9+/-4.9 mV at month 3, and 9.3+/-4.5 mV at month 6. Polarization was 1.0+/-0.1 mV at discharge, 1.1+/-0.5 mV at month 1, 1.1+/-0.2 mV at month 3, and 1.1+/-0.5 mV at month 6. Mean ER sensitivity was 3.7+/-1.8 mV at discharge, 4.0+/-1.8 mV after 1, 4.1+/-2.2 mV after 3, and 4.1+/-1.8 mV after 6 months. ER sensitivity could remain unadjusted in 14 patients. Programming to a less sensitive ER setting from 2.9+/-1.2 mV to 4.3+/-1.5 mV was possible in 21 patients. Programming to a more sensitive ER setting from 4.1+/-1.1 mV to 2.5+/-0.9 mV was required in nine patients because of the decrease of the ER signal. The automatic threshold tracking algorithm Autocapture could be activated in 95% of patients. Programming to more sensitive ER settings was recommended in 21% of the patients after hospital discharge. Therefore, ER signal and polarization must be checked at each follow-up, as a decrease in ER signal amplitude can make reprogramming of the ER sensitivity necessary. There is no risk for the patient if the ER is not sensed, as high voltage backup stimulation is present.

Aged↗

A randomized study on the effects of pacemaker programming to a lower output on projected pulse generator longevity.

The programmability of cardiac pacemakers enables the physician at follow-up to adjust the pacing pulse under consideration of the 100% safety margin with respect to the individual pacing threshold. The purpose of reducing the output is to prolong pacemaker longevity. The aims of this prospective, randomized trial were to compare the effects of nominal output versus a lower output on projected pacemaker longevity in single and dual chamber pacemakers. The secondary aim was to assess how many patients can be programmed to 2.5 V/0.4 ms instead of the nominal 3.5-V setting with > or = 100% safety margin. The patients received the same types of VVI or DDD pacemakers that were connected in the ventricle to the steroid-eluting, high impedance pacing lead. At the 3-month follow-up, patients with ventricular pacing thresholds < or = 0.15 ms at 2.5-V pulse amplitude were randomized to 3.5 V or 2.5 V amplitude at 0.4-ms pulse duration. Lead function and projected device longevity were assessed with the pacemaker's telemetry 6 and 12 months after implantation. Of patients implanted with a VVI pacemaker, at the 3-month follow-up, 3 patients had pacing thresholds > 0.15 ms at 2.5 V and 139 patients could be randomized. A reprogramming to a higher output was necessary in one patient. The mean percentage of ventricular pacing was about 40% throughout the study time. The programming to 2.5-V output resulted in an insignificant increase of device longevity from 117.9 +/- 18.7 months in the nominal group to 123.7 +/- 11.9 months at the 12-month follow-up (P = 0.16). Of patients implanted with a DDD pacemaker, 166 patients underwent randomization. The mean percentage of ventricular pacing was 85% in the ventricle and 35% in the atrium. The 2.5-V setting significantly prolonged pacemaker longevity from 98.1 +/- 21.3 to 112.0 +/- 13.6 months (P < 0.0001). In three (1%) patients a late increase of the pacing threshold was observed. Due to the low ventricular pacing thresholds, the 2.5-V/0.4-ms setting provided, 3 months after implantation, a > or = 100% safety margin in 99% of the patients. Programming to a lower output slightly increased projected pacemaker longevity compared to the nominal 3.5-V setting. Longevity increasedfor 5% in patients with single andfor 14% in dual chamber pulse generators.

Aged↗

Multicenter experiences with a single lead electrode for dual chamber ICD systems.

Monitoring of atrial signals improves the accuracy in identifying supraventricular tachyarrhythmias to prevent inappropriate therapies in patients with implantable ICDs. Since difficulties due to the additional atrial lead were found in dual chamber ICD systems with two leads, the authors designed a single pass VDD lead for use with dual chamber ICDs. After a successful animal study, the prototype VDD lead (single coil defibrillation lead with two additional fractally coated rings for bipolar sensing in the atrium) was temporarily used in 30 patients during a German multicenter study. Atrial and ventricular signals were recorded during sinus rhythm (SR), atrial flutter, AF, and VT or VF. The implantation of the lead was successful in 27 of 30 patients. Mean atrial pacing threshold was 2.5 +/- 0.9 V/0.5 ms, mean atrial impedance was 213 +/- 31 ohms. Atrial amplitudes were greater during SR (2.7 +/- 1.6 mV) than during atrial flutter (1.46 +/- 0.3 mV, P < 0.05) or AF (0.93 +/- 0.37 mV, P < 0.01). During VF atrial "sinus" signals had significantly (P < 0.01) lower amplitudes (1.4 +/- 0.52 mV) than during SR. The mean ventricular sensing was 13.3 +/- 7.9 mV and mean ventricular impedance was 577 +/- 64 ohms. Defibrillation was successful with a 20-J shock in all patients. In addition, 99.6% of P waves could be detected in SR and 84.4% of flutter waves during atrial flutter. During AF, 56.6% of atrial signals could be detected without modification of the signal amplifier. In conclusion, a new designed VDD dual chamber lead provides stable detection of atrial and ventricular signals during SR and atrial flutter. Reliable detection of atrial signals is possible without modification of the ICD amplifier.

Aged↗

Autocapture compatibility in patients with the MembraneEX lead and Affinity pulse generators.

The first Autocapture generation worked well with all recommended leads. The newer Autocapture generation provides a more sensitive resolution for evoked response testing and its implementation in a dual-chamber device. The purpose of the study was to evaluate the performance of the Affinity SR/DR pacemaker with the new Autocapture algorithm in combination with the small surface area pacing lead MembraneEX in 129 patients. Autocapture ventricular threshold, sensing threshold, lead impedance, evoked response (ER) and polarization signals were determined at implantation and discharge, as well as after 1 and 3 months. Autocapture recommendation rate was based on the ER sensitivity test. The median pacing threshold was 0.38, 0.50, 0.75, 0.75 V at implant, discharge, 1 and 3 months post-implant, respectively. The respective data for median lead impedance were 744, 605, 649 and 691 ohms; median sensing threshold was 12.5 mV at all visits. The median ER amplitude was 9.0, 10.1, 9.9 and 10.1 mV and the median polarization signal 0.39 mV at all visits. The frequency of recommended Autocapture activation was 98.3%, 99.2%, 98.3% and 96.2% of all patients at implant, at discharge, 1 and 3 months post-implant respectively. In conclusion, the studied pulse generator enabled, in combination with this pacing lead, in >95% of all patients activation of Autocapture.

Aged↗

Psychiatric conditions in patients with recurrent unexplained syncope.

AIMS: The relationship between syncope and psychiatric disorders is little investigated. This study evaluated the prevalence of psychiatric diseases and prognostic outcome in patients with recurrent unexplained syncope. METHODS AND RESULTS: After an inconclusive standard diagnostic work-up for syncope, including head-up tilt testing, a psychiatric evaluation was offered to 50 consecutive patients with recurrent syncope. The evaluation was accepted by 26 patients (77% females, 36 +/- 16 years) and refused by 24 (63% females, 50 +/- 19 years). A psychiatric disorder was diagnosed in 21 (81%) patients: 12 had depression, four panic attacks, two general anxiety, and three a somatization disorder. Only five patients showed normal psychosocial function. Of the patients with psychiatric disorders four accepted psychiatric care, such as psychotherapy and/or pharmacotherapy; 17 patients refused treatment. During 6 months of follow-up no patient under psychiatric care had syncope, while all patients without psycho- or pharmacotherapy had recurrent syncopal events. In these patients the median of syncopal episodes was three in a 6 months interval before and after clinical assessment. Patients who refused both psychiatric evaluation and therapy continued to experience syncope as before. CONCLUSIONS: In patients with recurrent unexplained syncope psychiatric alteration is common. However, patients seldom accepted a psychiatric evaluation and treatment.

Adolescent↗

Effects of a thin-sized lead body of a transvenous single coil defibrillation lead on ICD implantation. Kainox RV Study Group.

In the interest of patients receiving implantable cardioverter defibrillators (ICDs), the clinical benefits of newer and thinner transvenous defibrillation leads have to be determined. The aims of this study were to evaluate the ICD procedure duration and the frequency of lead dislocation at the 3-month follow-up of a new defibrillation lead with a thin-sized lead body and its conventional-sized predecessor. The thin-sized single coil defibrillation lead (Kainox RV, Biotronik; lead body 6.7 Fr) was implanted in 61 patients and the conventional-sized defibrillation lead (SPS, Biotronik; lead body 7.8 Fr) in 60 patients. Both leads were connected to a left-sided, prepectorally implanted Phylax ICD (Biotronik) with active housing. The lead implantation time and total procedure duration were determined. Lead implantation time was defined as the time from lead insertion to the end of the pacing measurements. The total procedure duration spanned skin incision to closure. The incidence of lead repositioning during the lead implantation time and during ventricular fibrillation conversion testing was also assessed. The frequency of lead dislocations was recorded at the 3-month follow-up. Mean lead implantation time and total procedure duration of the thin-sized lead (23 +/- 22 minutes 76 +/- 37 minutes) were not statistically different from the time needed for the conventional-sized lead (22 +/- 20 minutes 81 +/- 34 minutes). The number of lead repositionings during the lead implantation time was similar (thin-sized lead: 1.4 +/- 2.4; conventional-sized lead: 1.1 +/- 1.9). An additional lead repositioning was not necessary during ventricular fibrillation conversion testing in 93.4% of the patients with thin-sized and in 94.4% with conventional-sized leads (not significant). At the 3-month follow-up, there were four (6.6%) lead dislocations in the thin-sized and four (6.7%) in the conventional-sized lead group. In conclusion, the down-sized lead body of the new defibrillation lead influenced neither ICD procedure duration nor the incidence of lead dislocation during follow-up.

Case-Control Studies↗

Efficacy of single lead VDD pacing in patients with impaired and normal left ventricular function.

Atrial synchronous ventricular pacing seems to be the best pacing mode for patients with advanced AV block and impaired LV function. The long-term follow-up of single lead VDD pacing was studied in 33 patients with impaired LV function and compared to 42 patients with normal LV function. All patients received the same VDD lead and VDDR pacemaker. The lead model with 13-cm AV spacing between the atrial and ventricular electrode was implanted in 89% of the patients. Follow-ups were 1, 3, 6, and 12 months after implantation. The percentage of atrial sensing and the P wave amplitude were determined at each follow-up. Minimal P wave amplitude at implantation was 2.0 +/- 1.4 mV in patients with impaired and 1.7 +/- 0.9 mV with normal LV function (not significant). At the 12-month follow-up, 33 patients with normal and 23 patients with depressed LV function remained paced in the VDD mode. The remaining patients died in five (impaired LV function) and seven cases (normal LV function) or their pacemakers were programmed to the VVI/VVIR pacing mode in four (impaired LV function) and three cases (normal LV function). P wave amplitude did not differ in the two groups (e.g., at month 12: impaired: 1.17 +/- 0.42 mV; normal: 1.09 +/- 0.49 mV). The atrial sensitivity was programmed in most patients to sensitive settings with no differences between the two groups (e.g., at month 12: impaired: 0.13 +/- 0.06 mV; normal: 0.13 +/- 0.05 mV). The diagnostic counters indicated nearly permanent atrial sensing (e.g., at month 12: impaired: 99.3 +/- 2.2%; normal: 99.0 +/- 1.0 mV). In conclusions, single lead VDD pacing restored AV synchronous ventricular pacing in patients with normal and with impaired LV function indicating that it could be an alternative to DDD pacemakers, but not to dual-chamber pacing.

Aged↗

Automatic threshold tracking activation without the intraoperative evaluation of the evoked response amplitude. AUTOCAP Investigators.

Automatic threshold tracking (Autocapture) controls the amplitude of the pacing pulse and adjusts it to the actual pacing threshold. The algorithm is based on the proper detection of the evoked response (ER) amplitude after the pacing pulse. For this reason an intraoperative evaluation of ER and polarization is recommended. The aims of the study were to evaluate the ER signal and polarization and the performance of automatic threshold tracking without any intraoperative testing of the ER signal. In addition, the ER amplitude was correlated with the pacing threshold, pacing impedance, spontaneous R wave amplitude, and with the clinical data. The study included 60 patients who received the VVIR pacemaker Regency connected to the Membrane E 1450/1452 pacing lead (St. Jude-Pacesetter). At implantation, a pacing threshold < 0.7 V at 0.5 ms was achieved in all patients. ER and polarization were assessed for the first time at hospital predischarge testing. Follow-up measurements were conducted at month 1, 3, and 6. The ER amplitude at hospital discharge was 8.4 +/- 4.2 mV and increased to 9.4 +/- 4.8 mV at the 6-month follow-up. The pacemaker recommended not to program automatic threshold tracking on in one patient permanently and in three patients intermittently. The ER amplitudes were not differently distributed in men compared with women or in right-sided compared to left-sided implants. The correlation between age and the evoked response was r = 0.15. The correlation between ER amplitude and pacing threshold was r = -0.08, with pacing impedance r = 0.02, and with R wave amplitude r = 0.44. In conclusion, despite no operative evaluation of the ER amplitude being performed, the mean ER amplitude was about 9 mV at 6-month follow-up. Automatic threshold tracking could be programmed on in 93% of the patients throughout the time. Neither the clinical data nor the conventional electrical parameters help to predict patients who will have low ER amplitude or to optimize the ER signal at implantation.

Aged↗

Induction of atrial fibrillation with rapid high voltage ventricular pacing for ventricular fibrillation conversion testing. The Ventak AV II DR Study.

OBJECTIVE: To assess whether rapid high voltage ventricular pacing can also induce atrial fibrillation, and whether the induction of atrial fibrillation during ventricular fibrillation conversion testing is related to the patient's heart disease. DESIGN: Prospective study of 50 patients who received the dual chamber implantable cardioverter-defibrillator (ICD) Ventak AV II DR (Guidant) as a first implant. This device can record atrial activity even during a ventricular fibrillation episode and can induce atrial fibrillation by rapid atrial bursts. MAIN OUTCOME MEASURES: Frequency of atrial fibrillation after induction of ventricular fibrillation; clinical characteristics of patients with and without induced atrial fibrillation; frequency of atrial fibrillation induced by rapid atrial bursts during predischarge testing. RESULTS: Atrial fibrillation was observed in 40 of the 217 ventricular fibrillation episodes (18%) that could be detected immediately after delivery of high voltage pacing. The biphasic ICD shock for termination of ventricular fibrillation also terminated the atrial fibrillation in all cases. The 40 episodes of simultaneous atrial and ventricular fibrillation occurred in 18 patients (36%). The distribution of the clinical characteristics of the patients and the inducibility of atrial fibrillation during predischarge testing were similar in those with and without induced atrial fibrillation. CONCLUSIONS: Rapid high voltage ventricular pacing frequently induces atrial fibrillation, which was terminated by the subsequent biphasic ICD shock. The induction of atrial fibrillation seems to be a non-specific phenomenon, unrelated to the clinical status of the patient.

Adult↗

Acute effects of a class IA antiarrhythmic drug on the ventricular evoked response amplitude in patients with cardiac pacemakers.

Some newer cardiac pacemakers are able to control the efficacy of the ventricular pacing pulse beat by beat and to adjust the ventricular output to the actual pacing threshold. This capture verification is based on the detection of the ventricular evoked response amplitude, which has to be detected immediately after the pacing pulse. The sensitivity of the pacemaker to detect the evoked response amplitude must be adjusted individually to avoid the simultaneous detection of lead polarization. The aim of the present study was to evaluate the acute effects of a class IA antiarrhythmic drug on the evoked response amplitude and polarization in 13 pacemaker patients. The implanted pacemaker was the VVIR pacemaker Regency (St. Jude Medical), which provides the automatic capture verification algorithm Autocapture. The patients received 50 mg of ajmaline intravenously within 1 min. The evoked response amplitude and polarization were measured before and 2, 4, 6 and 8 min after ajmaline injection. The evoked response amplitude significantly decreased from 8.0 +/- 4.0 mV to a minimum value of 6.4 +/- 3.1 mV 2 min after drug administration. The decrease remained significant from the end of the application up to 6 min. The recommended sensitivity setting for the evoked response significantly (p < 0.05) decreased from 4.0 +/- 2.3 mV before to 3.1 +/- 1.3 mV 2 min after administration. No significant changes were observed for polarization. After the ajmaline application in 2 patients, the pacemaker recommended the deactivation of Autocapture for 9 min in 1 patient and 12 min in the other. The reasons were a decrease in the evoked response amplitude from 3.1 to 1.9 mV and from 9.0 to 5.7 mV, respectively, with a polarization ranging to about 3.0 mV. In conclusion, the ajmaline injection decreased the evoked response amplitude for some minutes. These findings indicate that antiarrhythmic drugs can alter the automatic capture verification function.

Aged↗

[Comparison of permanent and paroxysmal atrial fibrillation on the clinical outcome following radiofrequency AV node catheter ablation].

UNLABELLED: The aim of the study was to compare the clinical outcome after AV node radiofrequency current ablation and pacemaker implantation in patients with permanent and paroxysmal atrial fibrillation. The study consecutively included 19 patients with permanent (male n = 15; 55 +/- 13 years) and 45 patients with paroxysmal atrial fibrillation (male n = 31; 60 +/- 13 years). Clinical evaluation before and 1 year after ablation included the symptoms and the self-assessed state of health. Before ablation, both groups reported with a similar frequency about tachycardia (permanent n = 15; paroxysmal n = 2) and palpitations (permanent n = 5; paroxysmal n = 19). Patients with permanent (n = 7) atrial fibrillation had more often breathlessness than patients with paroxysmal atrial fibrillation (n = 2), whereas dizziness/syncope was more often reported by patients with paroxysmal (n = 35) than with permanent arrhythmia (n = 8). Ablation improved symptoms in patients with permanent (89%) as well as with paroxysmal atrial fibrillation (84%). The self-assessed state of health described as good had not patient before ablation and 63% patients with permanent and 47% with paroxysmal atrial fibrillation after ablation. There were no differences for the studied parameters between patients with permanent and paroxysmal atrial fibrillation. CONCLUSION: AV node ablation and pacemaker implantation improved the symptoms and the state of health in with paroxysmal and paroxysmal atrial fibrillation. No differences were detected between patients with permanent and paroxysmal atrial fibrillation.

Adult↗

Hemostasis with the ultrasonically activated scalpel. Effective substitute for electrocautery in surgical patients with pacemakers.

In patients with implanted pacemaker/cardioverter defibrillator (ICD), the use of electrocautery can lead to serious pacemaker dysfunction. The ultrasonically activated scalpel, however, which has been introduced mainly for the use in laparoscopic surgery, could potentially avoid the outlined problem, since no electrical current flows while in use. This hypothesis was tested in a pacemaker patient undergoing laparoscopic cholecystectomy. During the procedure, no abnormal rhythms or ECG interferences were detected while working in close vicinity to the device. Thus, the ultrasonically activated scalpel provides adequate hemostasis and does not bear the risk of pacemaker dysfunction.

Cholecystectomy, Laparoscopic↗

Prehospital testing for troponin T in patients with suspected acute myocardial infarction.

BACKGROUND: Cardiac troponin T (TnT) is a highly sensitive and specific marker for myocardial damage and can be detected early after myocardial injury. Our hypothesis was to use TnT as an objective marker to verify acute myocardial infarction before hospital admission. METHODS AND RESULTS: We evaluated the sensitivity of a rapid qualitative assay for serum TnT for the detection of acute myocardial infarction in the ambulance and assessed the predictive value of a positive prehospital TnT test for death and myocardial infarction during 6-months of follow-up. The study, conducted in an urban area, included 158 consecutive patients with suspected acute myocardial infarction (93 men aged 69 +/- 13 years). A myocardial infarction was confirmed in 40 and excluded in 118 patients. The prehospital TnT test was positive in 11 patients, of whom 7 had acute myocardial infarction. Fifty-three patients had a positive test result at hospital admission, with evidence of myocardial infarction in 39 of them. The sensitivity to acute myocardial infarction was 18% for the prehospital and 98% for the in-hospital test with 78% and 88% specificity, respectively. During follow-up, patients with a positive prehospital TnT test result had cardiac events more often (9 of 11) than patients with a negative result (26 of 147; P <.0001). CONCLUSIONS: In areas with short transport times to the patient the rapid TnT test performed at the point of care identified only a minority of the patients with acute myocardial infarction. A positive prehospital TnT test result seems to be an objective marker for a worse outcome in patients presenting with suspected acute myocardial infarction.

Aged↗

Effects of body position and exercise on evoked response signal for automatic threshold activation.

The Autocapture function controls and optimizes the output of the ventricular pulse amplitude automatically. For this reason an automatic test has to be performed during follow-up to measure the evoked response signal and lead polarization for the calculation of the appropriate evoked response sensitivity setting. The aim of the study was to assess whether body position and exercise influence the evoked response and polarization. Both parameters were determined in the supine and upright position and subsequently during supine and upright symptom-limited ergometry. The study included 14 patients with the VVIR pacemaker Regency SR+ who had received the ventricular pacing leads Membrane E 1450 T (n = 8), CapSure Z 5034 (n = 4), or SX 60 (n = 2). The evoked response signal was 7.4 +/- 3.3 mV during supine and increased to 9.7 +/- 5.6 mV (+35%) during upright position (P < 0.05). The exercise tests were terminated at 105 +/- 36 W (supine) and 110 +/- 34 W (upright). There was a gradual insignificant decrease of the evoked response during each exercise test with a mean decrease of -1.1 +/- 0.9 mV (-15%; supine) and -1.6 +/- 2.1 mV (-16%; upright). The evoked response increased within 5 minutes during recovery to the initial values. Polarization remained unchanged during both tests. The pacemaker did not recommend activating autocapture in four patients who all had received high-ohmic pacing leads. In conclusions, the measurement of the evoked response in supine position seems to represent the worst case. Physical activities did not effect autocapture function in patients with the recommended lead, but the pacemaker did not always recommend Autocapture activation in some patients with high-ohmic pacing leads.

Aged↗

Impact of long-term ECG recording on the detection of paroxysmal atrial fibrillation in patients after an acute ischemic stroke.

An ECG recording time of 24 hours has a low yield to detect atrial arrhythmias in patients after an acute ischemic stroke. The present study investigated whether a recording time of 72 instead of 24 hours detects paroxysmal atrial fibrillation in more patients. The study prospectively included 82 consecutive patients 2-3 weeks after an acute ischemic stroke. All patients had sinus rhythm in the resting ECGs and no history of atrial fibrillation or flutter. The frequency of atrial fibrillation was assessed after 24, 48, and 72 hours of ambulatory ECG monitoring. An ECG monitoring time of 72 hours documented paroxysmal atrial fibrillation in five (6%) patients. The episode of paroxysmal atrial fibrillation occurred in only one patient within 24 hours. The other patients had their first episode of atrial fibrillation between 24 and 48 hours (n = 2) and between 48 and 72 hours (n = 2). These five patients were older (age = 70 +/- 5 years), whereas the mean age of the remaining patients was 59 +/- 13 years. All five patients had cardiovascular disease in comparison to 36 of 77 patients and reported palpitations in comparison to 6 of 77 of the remaining patients. In conclusion, ambulatory ECG monitoring over 72 hours detected after the first recording day four of five patients in whom paroxysmal atrial fibrillation could be documented for the first time. The 72-hour recording time improved, compared to the 24-hour period, the detection of paroxysmal atrial fibrillation in patients after an ischemic stroke. It seems to be more efficient to perform prolonged ECG recording mainly in older patients with a cardiovascular disease and/or a history of palpitations.

Acute Disease↗

Uni- and bipolar pacing threshold measurements with capturecontrol, a new automatic pacemaker function for capture verification. Comparison of the automatic and the manual pacing threshold measurement.

It is mandatory in pacemaker patients to determine the pacing threshold at each follow-up visit. To facilitate the pacing threshold measurements, complete automatic pacemaker tests are being developed. A new pacemaker algorithm for automatic capture verification (Capturecontrol) detects the presence of the evoked response signal 70-100 ms after the pacing pulse. The aim of this study was to determine the uni- and bipolar pacing thresholds using this automatic pacemaker function and compare them with the manually determined threshold. The study included 14 patients with the DDD pacemaker Logos (Biotronik) connected to the high-ohmic ventricular pacing lead Synox SX 60-BP (Biotronik). At predischarge and 8, 14, 20 and 26 weeks after implantation the uni- and bipolar pacing thresholds were assessed manually and with the automatic pacemaker function at 0.4 ms duration. Mean pacing thresholds determined with the automatic pacemaker function were not different from the manually measured values. This applied to uni- and bipolar pacing. Seventy percent of all unipolar and 67% of all bipolar measurements had no deviation. A deviation of 0.1 V between manual and automatic threshold measurement was observed in 25% (unipolar) and 28% (bipolar), respectively. A 0.2-volt difference occurred in 3% in the unipolar measurements. Deviations >/=0.3 V were found in 2% of all unipolar and in 5% of all bipolar measurements. In conclusion, automatic pacing threshold measurements using the Capturecontrol algorithm were similar compared to the manually determined thresholds. The excellent agreement between the two methods was observed for bipolar as well as unipolar pacing and on condition that all patients had a high-ohmic ventricular pacing lead. Therefore, fewer requirements are necessary for Capturecontrol than for presently available systems. Such pacemaker functions can help to speed up the measurements during follow-up visits.

Aged↗

[Pacemaker therapy in patients with atrial fibrillation].

Pacemaker therapy in patients with atrial fibrillation means the best current pacemaker therapy for patients with bradycardias with the aim to avoid the onset of atrial fibrillation and to establish DDD pacing despite of a history of atrial tachyarrhythmias. The newer application of pacing is the suppression of atrial arrhythmias in patients with medical refractory atrial tachyarrhythmias. Patients with slow ventricular rates and permanent atrial fibrillation should receive a VVI-pacemaker, if the bradycardias causes syncope, dizziness or a decrease of their exercise tolerance. In case of chronotropic incompetence the pacemaker should provide rate responsive pacing. Patients with sick sinus syndrome should receive an atrial (AAI) or dual-chamber (DDD) pacemaker, because patients with these in contrast to VVI-pacemakers develop less often atrial fibrillation and subsequent complications such as atrial thromboembolism. A dual-chamber or VDD-pacemaker--the latter connected to a VDD-single-lead--is indicated in patients with advanced AV-block. Atrial fibrillation occurs in 3 to 6% of the patients with no history of arrythmia and is, if pacemakers have no automatic mode switch, an often reason to program the devices to the VVI-pacing mode. Nowadays, most DDD(R)-pacemakers provide an automatic mode switch: During an atrial tachycardia the pacemaker switches to a VVI/VVIR mode and restores the initial DDD(R)-pacing mode with termination of the arrhythmia. In respect to the newer applications, one approach to prevent atrial tachyarrhythmias is permanent atrial pacing. As lower pacing rates of 80 to 90 ppm are usually needed and many patients hardly tolerate these pacing rates, new algorithms are under clinical investigation. Another approach is the simultaneous depolarization of the right and left atrium. Biatrial pacing is performed with one lead in the high right atrium and another lead in the coronary sinus. Another solution is bifocal atrial pacing with leads placed in the high right atrium and in the coronary sinus ostium. One effect of the new pacing techniques is to shorten interatrial conduction times. Therefore, biatrial pacing has become a therapy to prevent atrial arrhythmias deriving from delayed interatrial conduction times. As atrial reentry circuits seem to be important in atrial fibrillation, multisite atrial pacing is also performed in patients with medical refractory paroxysmal atrial fibrillation. Preliminary results suggest a more effective prevention of atrial fibrillation; nevertheless, these techniques should be still restricted to patients enrolled in clinical studies.

Atrial Fibrillation↗