Collaborative investigation of the AutoMicrobic System Enterobacteriaceae biochemical card.

The Enterobacteriaceae biochemical card used in six separate laboratories to identify 170 representatives of Enterobacteriaceae. The AutoMicrobic System (Vitek Systems, Inc.) performed with an accuracy of 97.8% as compared with 98.1% by the standard method selected and 97.6% by a commerically prepared manual system approach. During this time, 5,450 clinical isolates belonging to Enterobacteriaceae were analyzed. Compared with the routine methods used in the various laboratories, the AutoMicrobic System identified 96.4% correctly

Bacteriological Techniques↗

Clinical laboratory evaluation of automated microbial detection/identification system in analysis of clinical urine specimens.

More than 4,000 clinical urine specimens were evaluated with an automated microbial detection/identification system compared to a standarized manual analysis and the routine modalities used in five peer-group laboratories. The comparison indicates that the automated system recognizes the nine groups of significant microorganisms in urinary tract infections in hospitalized patients with the same efficiency as a standarized manual method. The automated system's ability to enumerate the bacterial populations in the original clinical specimen attained a high degree of accuracy.

Automation↗

Multi-laboratory evaluation of an automated microbial detection/identification system.

An automated and computerized system (Automicrobic System [AMS]) for the detection of frequently encountered bacteria in clinical urine specimens was tested in a collaborative study among six laboratories. The sensitivity, specificity, reliability, and reproducibility of the AMS were determined, and the system was compared with conventional detection and identification systems. In this study, pure cultures and mixtures of pure cultures were used to simulate clinical urine specimens. With pure cultures, the sensitivity of the AMS in identifying the nine groups of organisms most commonly found in urine averaged 92.8%. The specificity averaged 99.4%, and the reliability of a positive result averaged 92.1%. The latter value was strongly influenced by a relatively high occurrence of false positive Escherichia coli results. The AMS was capable of detecting growth of most organisms, including those which it was not designed to identify. However, it identified some of these incorrectly as common urinary tract flora. Reproducibility of results, both within laboratories and among different laboratories, was high. Fast-growing organisms, such as E. coli and Klebsiella/Enterobacter species, were detected often at cell populations well below the AMS enumeration threshold of 70,000/ml. In mixed culture studies, high levels of sensitivity and specificity were maintained but when Serratia species were present in mixtures with other organisms, there was often a false positive report of E. coli. The overall performance of the AMS was considered satisfactory under the test conditions used.

Bacteria↗