Unipolar defibrillator?

An unusual case of "unipolar" pacing and myopotential over-sensing leading to an inappropriate ICD shock in a patient with an implanted defibrillator is reported. The reasons for unipolar behavior in a system using a committed bipolar device are discussed.

Algorithms↗

Pacing alternans: an unusual pattern of pacemaker behavior.

We report an unusual pattern of pacemaker function related to the "autocapture" feature of a recently released pacemaker model. The electrocardiogram reveals pacing alternans. This report discusses the differential diagnosis and the correct explanation.

Algorithms↗

An electrocardiographic masquerade.

A 60-year-old woman underwent successful pulmonary embolectomy for a massive pulmonary embolism. On postoperative day 2, while receiving intravenous dopamine for hypotension, she developed the tachycardia, and a 12-lead ECG was obtained (Fig. 1). Figure 2 shows the 12-lead ECG obtained on the previous day, when she was in sinus rhythm with frequent premature

Bundle-Branch Block↗

Histopathologic effects of radiofrequency catheter ablation in previously infarcted human myocardium.

INTRODUCTION: The use of catheter-based radiofrequency (RF) ablation for the treatment of ventricular tachyarrhythmias due to previous myocardial infarction has been steadily increasing. The histopathologic changes caused by this technique are not well described in humans. METHODS AND RESULTS: Three patients with hemodynamically tolerated ventricular tachycardias (VTs) due to previous myocardial infarction underwent endocardial mapping and catheter based RF ablation. All patients received between 5 and 11 RF lesions each of 60-second duration. One patient underwent myocardial resection of a left ventricular aneurysm 1 day following RF ablation, one expired 7 days after RF ablation, and one expired 9 months after RF ablation. None of the deaths occurred as a result of RF ablation. Pathologic specimens obtained early after RF ablation revealed areas of focal acute inflammation and fibrin deposition. Later specimens revealed several focal areas of fibrosis and granulation tissue. Specimens obtained late after RF ablation revealed a dense band of fibrosis, measuring 17 x 17 x 5 mm (1,250 mm3). CONCLUSION: Catheter-based RF ablation of ischemic VT in humans causes lesions that initially resemble coagulation necrosis. This is followed by the development of an inflammatory infiltrate and, finally, the development of fibrosis. Repeated application of RF ablation may result in much larger lesions than have been previously reported.

Aged↗

Cardiac death and stored electrograms in patients with third-generation implantable cardioverter-defibrillators.

OBJECTIVES: We sought to utilize terminal stored intracardiac electrograms (EGMs) to study the electrophysiologic events that accompany mortality in patients with third-generation implantable cardioverter-defibrillators (ICDs). BACKGROUND: Despite their ability to effectively terminate ventricular tachyarrhythmias, cardiac mortality in patients with ICDs remains high. The mechanisms and modes of death in these patients are not well understood. METHODS: We retrospectively analyzed clinical data and stored EGMs from patients enrolled in the clinical trial of the Ventritex Cadence ICD. Of the 1,729 patients 119 died during 6 years of follow-up. The final recorded EGM was reviewed. Postimplant EGMs as well as 50 control EGMs were used to define normal EGM characteristics. RESULTS: There were 36 noncardiac deaths (30%) and 83 cardiac deaths (70%). Of the cardiac deaths, 55 (66%) were nonsudden and 28 (34%) were sudden. When cardiac deaths were analyzed, 46 (55%) had no stored EGMs within 1 h of death, implying that the deaths were not directly related to tachyarrhythmias. In 37 cardiac deaths (18 nonsudden, 19 sudden), stored EGMs were present within 1 h of death. In these 37 deaths, the final EGM recorded was wide (>158 ms) in 33 (89%). Wide EGMs were interpreted as ventricular tachycardia in 27 and ventricular fibrillation in 6. In 13 of the 33 patients (39%) with wide EGMs, therapy was not delivered by the ICD, as it incorrectly detected a spontaneous termination of the arrhythmia. EGMs were significantly wider if recorded within 1 h, as compared with those recorded from 1 to 48 h before death (261+/-124 vs. 181+/-93 ms, p=0.04). CONCLUSIONS: Only 37 patients (31%) who died after placement of an ICD had a stored EGM within 1 h of the time of death, suggesting that the majority of deaths (69%) were not the immediate result of a tachyarrhythmia. When EGMs were recorded, they were wide in 89% of patients. These wide EGMs most likely represent intracardiac recordings of electromechanical dissociation. Thus, of the 119 deaths, 112 (94%) were not the immediate result of a tachyarrhythmia.

Aged↗

Sustained bundle branch reentry in a patient with hypertrophic cardiomyopathy and nondilated left ventricle.

Ventricular tachycardia is a well-known complication in patients with hypertrophic cardiomyopathy. We report the case of a patient with hypertrophic cardiomyopathy with easily inducible monomorphic ventricular tachycardia. Electrophysiology study demonstrated that bundle branch reentry was the mechanism of the tachycardia. The tachycardia was rendered non-inducible by radiofrequency ablation of the right bundle branch.

Adult↗

The yield of head-up tilt testing is not significantly increased by repeating the baseline test.

BACKGROUND AND HYPOTHESIS: Head-up tilt is widely used in evaluating patients with syncope. Data suggest that baseline tilt without isoproterenol is associated with few (< or = 10%) positive results in normals. However, there is considerable day-to-day variability in patients with syncope, resulting in low reproducibility. We hypothesized that repeating the baseline tilt would increase the total number of positive responses. METHODS: In all, 193 patients with syncope of unknown etiology prospectively underwent serial baseline tilt (Tilt-1 and Tilt-2) at 70 degrees for 25 min on two consecutive days. RESULTS: Of 193 patients, 39 (20%) had a positive Tilt-1. Of these, 19 had a positive Tilt-2, (positive concordance rate: 19/39 = 49%). Of 154 patients with a negative Tilt-1, 145 had a negative Tilt-2 (negative concordance rate: 142/151 = 94%). Only 9 of 154 patients (6%) with a negative Tilt-1 had a positive Tilt-2. Thus, the yield increased from 20% (39/193) after Tilt-1, and to 25% (48/193) after Tilt-1 plus Tilt-2, p = NS. CONCLUSIONS: (1) Reproducibility of a positive Tilt-1 is moderate. (2) Reproducibility of a negative Tilt-1 is high. (3) Repetition of tilt in patients with a negative Tilt-1 does not significantly increase the number of positive results.

Blood Pressure↗

Vagal modulation of RR intervals during head-up tilt and the infusion of isoproterenol.

The objective of this study was to characterize the autonomic effects of 2 interventions, head-up tilt and isoproterenol infusion, which alter autonomic balance by different mechanisms but produce the same RR intervals. We compared the effect of head-up tilt with the effect of isoproterenol on autonomic balance as measured by power spectral analysis of RR variability. Fifteen normal subjects had baseline measurements and then underwent head-up tilt. After return to baseline supine values, isoproterenol was infused at a rate of 1 microgram/min (low-dose isoproterenol), which was then increased to 2.1 +/- 0.5 microgram/min (high-dose isoproterenol). All subjects underwent a second tilt during high-dose isoproterenol, and 9 subjects completed this second tilt study. During the experiment, normal RR intervals were recorded and 5-minute segments were used to calculate power spectra. High-frequency (HF) power (0.15 to 0.40 Hz) was used as a measure of vagal activity. The effects of head-up tilt were compared with the effects of low-dose isoproterenol. Despite nearly identical mean RR intervals (784 ms with tilt vs 792 ms with low-dose isoproterenol, p = NS), there was significantly (p < 0.05) less HF power during head-up tilt in the drug-free state (172 ms2) than during low-dose isoproterenol in the supine position (307 ms2). A second head-up tilt was performed during the infusion of high-dose isoproterenol. During high-dose isoproterenol, tilt caused a decrease in RR intervals (from 573 to 491 ms; p < 0.01) and a decrease in HF power (from 68 to 28 ms2; p < 0.05).(ABSTRACT TRUNCATED AT 250 WORDS)

Adult↗

Prevalence and prognostic significance of atrial arrhythmias after orthotopic cardiac transplantation.

OBJECTIVES: We studied the duration and prognostic significance of atrial arrhythmias in the denervated transplanted heart, specifically the occurrence of atrial fibrillation in the absence of vagal modulation. BACKGROUND: Substantial animal data indicate that vagally induced dispersion of atrial refractoriness plays a central role in the induction and maintenance of atrial fibrillation. METHODS: We studied the occurrence of atrial arrhythmias in the denervated hearts of 88 consecutive orthotopic transplantations in 85 patients by means of continuous telemetry and all available electrocardiographic tracings. RESULTS: Fifty percent of recipients (44 of 88) developed at least one atrial arrhythmia. Atrial fibrillation occurred 23 times (21 recipients), atrial flutter 39 times (26 recipients), ectopic atrial tachycardia 3 times (3 recipients) and supraventricular tachycardia 18 times (11 recipients). The number of atrial fibrillation and atrial flutter episodes did not differ (23 vs. 39, p = 0.072), but the mean duration of atrial flutter was longer than that of atrial fibrillation (37.0 +/- 10 vs. 6.6 +/- 3.6 h, p = 0.014). Atrial fibrillation was associated with an increased risk of subsequent death (10 of 21 recipients with vs. 15 of 67 without atrial fibrillation, risk ratio 3.15 +/- 0.18, p = 0.005 by Cox proportional hazards model). All 5 recipients who developed "late" atrial fibrillation (> 2 weeks after transplantation) died versus 5 of 16 who developed atrial fibrillation within the first 2 weeks (p = 0.007). Causes of death included rejection (three recipients), allograft failure (two recipients), infection (three recipients) and multiorgan failure (two recipients). Atrial fibrillation was not associated with age, gender, ischemic time, reason for transplantation, echocardiographic variables, invasive hemodynamic variables or biopsy grade. Mean time from atrial arrhythmia to echocardiography was 2.7 +/- 3.3 days; that to biopsy was 4.8 +/- 6.3 days. Atrial flutter was not associated with subsequent death. Only 7 (15.9%) of 44 recipients demonstrated moderate or severe allograft rejection at the time of the arrhythmia. CONCLUSIONS: Atrial arrhythmias occur frequently in the denervated transplanted heart, often in the absence of significant rejection. Late atrial fibrillation may be associated with an increased all-cause mortality.

Actuarial Analysis↗