Effect of fabrication method and resin type on performance of tibial bearings.

Polyethylene has been used successfully for more than 30 years as an orthopedic bearing material. During this time, several polyethylene resins and fabrication methods have been used to produce bearings. Some bearings fail prematurely due to fatigue, which has been linked to oxidation and degradation of mechanical properties resulting from gamma sterilization in air. Fabrication method and/or resin have been hypothesized to govern whether oxidative degradation occurs in gamma-sterilized bearings. This study evaluates the effect of fabrication (machining/direct compression molding) and resin type on oxidation and the resulting mechanical properties for a large series of never-implanted bearings. While many molded bearings studied exhibit lower oxidation than machined bearings, fabrication method is not a significant predictor of oxidation. Resin type and shelf-age are found to be significant predictors of oxidation. Bearings fabricated from Himont 1900 exhibit lower oxidation than those from GUR 415/412 at comparable times after gamma in air. However, Himont 1900 bearings lose strength and elongation at lower oxidation levels than GUR 415/412 bearings. But since Himont 1900 oxidizes more slowly, Himont 1900 bearings retain mechanical properties for longer shelf times than comparable GUR 415/412 bearings. These effects are seen in retrievals as well.

Biocompatible Materials↗

Current sterilization and packaging methods for polyethylene.

Gamma sterilization in an air environment can induce oxidation in polyethylene. Oxidation can lead to polyethylene embrittlement, compromising mechanical integrity and clinical performance of polyethylene bearings. For these reasons, orthopaedic manufacturers have modified their methods of sterilizing and packaging polyethylene. Two alternative approaches have emerged: sterilization by non-radiation methods and sterilization by gamma irradiation in inert environments. The current study presents a prognosis for clinical performance of polyethylene sterilized with new methods, based on material property analyses (oxidation levels, mechanical properties, crosslink density) of never implanted and retrieved bearings. Data from bearings that were never implanted which were sterilized with the new methods and shelf aged as many as 3 years, show negligible oxidation, ductility above 400%, and ultimate tensile strength near 50 MPa, all exceeding specifications of the American Society for Testing and Materials. There are significant differences in crosslink density (swell ratio) depending on the sterilization method. Retrievals indicate that bearings sterilized with these new methods are performing well clinically and that the majority are not changing with time. The current study suggests that the shelf oxidation problem has been addressed by these new sterilization techniques and that clinical performance at short followup is acceptable. However, long-term clinical performance must be evaluated in the future.

Drug Packaging↗

In vitro simulation of contact fatigue damage found in ultra-high molecular weight polyethylene components of knee prostheses.

An in vitro simulation of fatigue loading of ultra-high molecular weight polyethylene (UHMWPE) knee components was carried out on a knee simulator and on a rolling and sliding wear tester. Tibial components for the knee simulator were gamma-sterilized, implantable components taken from manufacturing inventory. The rolling/sliding UHMWPE discs were machined from bar stock and either gamma sterilized in air and accelerated aged, or left as non-sterilized (controls). Cracking and delamination of samples that had been gamma sterilized in air and aged were observed in both types of tests. Contact fatigue damage was visible in as few as 150,000 cycles using the knee simulator at loads of 122 N (275 1b). The rolling/sliding samples showed signs of damage in as few as 130,000 cycles with an estimated stress of 15 MPa and 25 per cent sliding. However, cracking and delamination were not generated in the never-sterilized or recently sterilized controls. UHMWPE that has been gamma sterilized in air and aged is shown to be susceptible to contact fatigue damage. These results are important to the interpretation of in vitro total knee replacement simulations used to assess the performance of tibial bearings.

Compressive Strength↗

An analysis of hylamer and polyethylene bearings from retrieved acetabular components.

Hylamer and conventional polyethylene acetabular liners of the same design, revised for a variety of reasons, were examined and compared to assess the performance of Hylamer as a bearing material. Clinical damage modes, linear wear rates, oxidation levels, and mechanical properties were measured. In both series, many liners were retrieved for dislocation. Wear/osteolysis was the most common reason for retrieval in the Hylamer series, while none of the conventional polyethylene liners were retrieved for this reason. Nearly all liners exhibited abrasion, burnishing, scratching, and creep. The Hylamer liners had more cracking, delamination, and pitting. The Hylamer liners had an average linear wear rate of 0.32 mm/year, while the conventional polyethylene liners had an average wear rate of 0.20 mm/year. Due to sample size, no statistical difference in wear rate was noted between the two groups. In general, both the Hylamer and conventional polyethylene showed oxidation peaks subsurface, resulting from their exposure to gamma radiation in air. Liners with elevated oxidation had decreased ultimate tensile strength, elongation, and toughness. For given oxidation levels, the corresponding mechanical properties of Hylamer appeared lower than those of conventional polyethylene. The ultimate tensile strength values ranged from 14 to 33 MPa for Hylamer and 19 to 32 MPa for conventional polyethylene. Elongation ranges were 19% to 350% (Hylamer) and 80% to 375% (conventional). The Hylamer retrievals in this study gave initial indications of performance; Hylamer appeared to behave similarly, but not superiorly, to conventional polyethylene, in the early functional period with respect to clinical wear and clinical performance. Both Hylamer and conventional polyethylene liners were degraded by gamma sterilization in air, with Hylamer liners demonstrating greater property changes.

Chemistry, Pharmaceutical↗

Shelf life and in vivo duration. Impacts on performance of tibial bearings.

Polyethylene has been used for more than 30 years as an orthopaedic bearing material; however, recently concern has been focused on the early failure of some polyethylene bearings. The damage seen in some bearings has been linked to gamma radiation sterilization performed in an air environment. Gamma sterilization in air has been documented to cause an increase in oxidation and degradation of mechanical properties that continue with time. However, not all retrieved bearings that are gamma sterilized in air exhibit the elevated oxidation and mechanical property degradation that lead to early component failure. Bearings that are gamma sterilized in air oxidize while sitting in inventory before implantation. Shelf oxidation rate was estimated based on analysis of a series of never implanted tibial bearings. This shelf oxidation rate allowed estimation of in vivo oxidation for retrieved tibial bearings of known sterilization date. Bearings with less than 1 year of shelf life after gamma sterilization in air had lower in vivo oxidation and better in vivo performance than did those with longer shelf life before implantation. Shelf time before implantation appears to be a significant factor in the success or failure of bearings that are gamma sterilized in air.

Aged↗

Overview of polyethylene as a bearing material: comparison of sterilization methods.

Polyethylene has been used for more than 30 years as an orthopaedic bearing material; however, there has been recent concern regarding the early failure of a small percentage of the polyethylene bearings. The damage seen in some retrieved polyethylene components has been linked to gamma radiation sterilization in air, which was widely used by the industry for years. Gamma radiation in air has been documented to cause an increase in oxidation and degradation of mechanical properties with time. The degradation of polyethylene initiated by gamma sterilization in air has led the orthopaedic industry toward alternative sterilization methods, including gamma radiation in an inert gas or vacuum environment, ethylene oxide gas sterilization, and gas plasma sterilization. For many of these alternative techniques, little clinical performance data exist. This study is a comparative evaluation of sterilization methods using the same analytic techniques that have been used to document the effects of gamma sterilization in air on polyethylene. Fourier transform infrared spectroscopy, electron spin resonance, and uniaxial tensile testing are used to compare, respectively, the oxidation levels, free radical concentration, and mechanical properties of material sterilized by each method. The polyethylene is evaluated before sterilization, poststerilization, and postartificial aging. All examined alternative sterilization methods, when compared with gamma sterilization in air, caused less material degradation during a component's preimplantation shelf life.

Electron Spin Resonance Spectroscopy↗

The Otto Aufranc Award. Impact of gamma sterilization on clinical performance of polyethylene in the hip.

Despite studies to determine their causes, significant variations in polyethylene acetabular component wear rates, radial cracking of component rims, and occasional delamination cannot be explained. A subsurface white band frequently occurs in such damaged components. These damaged components often are gamma sterilized. To date, the origin of the band and its effect on polyethylene chemical and mechanical properties, and hence, clinical performance, have not been confirmed, and correlations between radiation sterilization and clinical wear have not been made. By developing techniques for polyethylene retrieval testing and rating, chemical analysis, and mechanical analysis, this research has determined that gamma sterilization in air alters the chemical and mechanical properties of polyethylene over time, resulting in high subsurface oxidation, reduced ductility, and reduced strength. Gamma sterilization-induced oxidation is found to be most severe in the subsurface region of components, and coincides with zones of significantly reduced strength and ductility. This chemical and mechanical property degradation is time dependent and is not typically visible until after 3 years' postirradiation. The presence of the subsurface white band significantly correlates with clinical cracking and delamination observed in retrieved components. Wear of the retrieved components often is observed to have progressed into this heavily oxidized, weakened, and embrittled zone. A method for accelerated aging shows that irradiating in air causes oxidation damage in polyethylene components that is not seen with other sterilization methods. Modifications of gamma sterilization techniques to minimize this damage are discussed.

Biomechanical Phenomena↗

The tradeoffs associated with modular hip prostheses.

In an effort to gain greater insight into the tradeoffs associated with modular hip prostheses, 2 approaches were taken. A questionnaire was sent to each of the orthopaedic implant manufacturing companies asking specific questions regarding modular components, and a series of retrieved prostheses, both modular and nonmodular, were examined to determine the potential sources of problems associated with modular connections. The respondents to the questionnaire generally agreed that it was more expensive to produce modular prostheses due to the required tolerances at the modular connections, and that the increased flexibility provided by the modularity was important to surgical outcome. There was less consensus on whether inventories were reduced and little data to support any improvement in surgical outcome caused by modularity. The most frequent problems associated with modular connections were fretting and corrosion. Easily observable significant fretting occurred in 4% of 701 head/neck tapers. Corrosion was observed in > 30% of the mixed-alloy head/stem combinations, in < 10% of all-titanium-alloy modular components, and in < 6% of all-cobalt-alloy devices. In 1 series of retrieved modular femoral components (15 titanium alloy and 15 cobalt alloy) with both sets having approximately the same duration of implantation, 7% of the all-cobalt-alloy components had corrosion, whereas 33% of the mixed-alloy components had corrosion.

Alloys↗