Standardization of the Bausch & Lomb Autoplot tangent screen.

A method is described of modifying the Bausch & Lomb Autoplot to obtain background and test target luminances of 10.02 cd/m2 and 32.8 cd/m2, respectively. Together with the use of a 2-mm target, these test conditions closely correspond to those of the Goldmann perimeter. These conditions can be maintained to provide more standardized perimetry than is currently available with the Autoplot.

Calibration↗

Perimetry: a need for standardization.

Luminance measurements of targets and backgrounds revealed differences for various types of perimetric instruments. A lack of uniform testing conditions for instruments of the same design was also noted. Of the instruments measured, hemispheric-projection perimeters seemed to provide the most repeatable test conditions.

Humans↗