Gated cardiac computed tomography with a motion phantom.

CT images of the heart are significantly degraded by the effects of motion during the scanning interval. The use of electrocardiographic (ECG) gating to provide "stop-action" cardiac images remains limited by artifacts. A motion phantom has been constructed to allow systematic study of the artifact structure of ECG-gated images and to isolate the origins of these artifacts. "Stop-action" reconstructed images are presented demonstrating two classes of artifacts: (a) pinwheel artifacts that appear at the edges of high-contrast moving objects; and (b) linear streaks occurring in relation to missing views when an incomplete angular set of projections is used for image reconstruction with the convolution back-projection algorithm. These results underscore the use of the motion phantom for the analysis of image artifacts and stress the need for new CT reconstruction algorithms which are optmized for "stop-action" scanning.

Heart↗

Stop-action cardiac computed tomography.

Computed tomographic (CT) cardiac imaging in vivo has been hampered by motion of the heart during its cardiac cycle. A technique of post data-acquisition correlation of the angular projection data using the electrocardiogram as a reference signal is described. This method produced seven "stop action" images of the heart and resulted in delineating morphological detail not recognizable on the conventional CT scan.