PubMed Health⌕ Search

Biomedical subjects

Chien-Yuan Kuo

Publications and source records attributed to Chien-Yuan Kuo.

4 recordsLinked to original sources

Solid-phase extraction and large-volume sample stacking with an electroosmotic flow pump in capillary electrophoresis for determination of methotrexate and its metabolites in human plasma.

This paper describes approaches for large-volume sample stacking (LVSS) with an EOF pumpin CE for the determination of methotrexate (MTX) and its metabolites in human plasma. After pretreatment of plasma through a SPE cartridge, a large sample volume was loaded by hydrodynamic injection (3 psi, 70 s) into the capillary filled with phosphate buffer (70 mM, pH 6.0) containing 0.01% polyethylene oxide. Following removal of a large plug of sample matrix from the capillary using polarity switching (-25 kV), the separation of anionic analytes was subsequently performed without changing polarity again, achieving an improvement of sensitivity of around a 100-fold. The method was applied to therapeutic drug monitoring of MTX in one acute lymphoblastic leukemia patient. This study is one of very few applications showing the feasibility of LVSS in analysis of biological samples by CE.

Antimetabolites, Antineoplastic↗

High-performance liquid chromatography with electrochemical detection for analysis of gliclazide in plasma.

A sensitive HPLC-electrochemical detection method was developed for the analysis of gliclazide (GL) in human plasma. After deproteination of 100 microL of plasma by acetonitrile, evaporation, and reconstitution, GL was separated on a C18 column (150 mm x 4.6 mm) by the mobile phase (70 mM disodium tetraborate, pH 7.5, containing 26.5% of acetonitrile). The regression equations were linear (r> 0.9990) over the range of 50 nM to 4.00 microM. The precision and accuracy of intra- and inter-day analysis were less than 5.3 and 0.93% for relative standard deviation and relative error, respectively. The limit of detection for plasma was 10 nM for GL (S/N = 3, 10 microL injection). This newly developed method was applied for monitoring blood levels with one healthy volunteer dosing with a GL tablet.

Adult↗

Micellar electrokinetic chromatography for simultaneous determination of six corticosteroids in commercial pharmaceuticals.

We have developed a micellar electrokinetic chromatography (MEKC) method using bile salts for the simultaneous determination of six corticosteroids, including betamethasone, cortisone, prednisolone, 6alpha-methylprednisolone, triamcinolone, and prednisone. The separation was performed using borate buffer containing sodium cholate and sodium deoxycholate. Several parameters were studied, including bile salt concentrations, concentrations and pH of borate buffer, and analytical voltages. In method validation, calibration curves were linear over a range of 10-100 microM for each corticosteroid. The RSD (relative standard deviation) and RE (relative error) were all less than 5% for intra- and interday assays. The limit of detection of each analyte was 5 microM. The recoveries were greater than 95%. Application of this method for quality control of commercial tablets also proved to be feasible. All analytical values fall within the labeled amount of 90-110% for betamethasone and prednisolone, and of the labeled amount of 92.5-107.5% for 6alpha-methylprednisolone, as required by the United State Pharmacopeia 25 (USP 25).

Adrenal Cortex Hormones↗

Simultaneous determination of methotrexate and its eight metabolites in human whole blood by capillary zone electrophoresis.

A simple and selective capillary electrophoretic method was established for the simultaneous determination of methotrexate (MTX), 7-hydroxymethotrexate (7-OHMTX), 2,4-diamino-N10-methylpteroic acid (DAMPA), and polyglutamate derivatives [MTX-(Glu)n, n=2-7] in whole blood. After extraction, those analytes were separated by fused-silica capillary and a running buffer containing glycine (1.2 M, pH 9.3). The quantitative ranges were 10-50 microM for each analyte. The intra- and inter-day RSD and RE values were all less than 6 and 11%, respectively. The limits of detection (S/N= 3, injection 5 s) were found to be 1 microM for MTX, 7-OHMTX, MTX-(Glu)2, MTX-(Glu)3, and MTX-(Glu)4; 3 microM for MTX-(Glu)5 and MTX-(Glu)6; 5 microM for MTX-(Glu)7, and 8 microM for DAMPA. All recoveries were greater than 94%. This method was applied to blood MTX monitoring in a patient with acute lymphoblastic leukemia.

Antimetabolites, Antineoplastic↗