Bench evaluation: three face-shield CPR barrier devices.
UNLABELLED: Due to the fear of disease transmission, the practice of mouth-to-mouth (M-M) rescue breathing is rarely performed; to address this concern, many types of CPR barrier devices have been developed. These include bag-valve-mask devices, mouth-to-mask devices, and face shields (FS). The purpose of this study was to measure the volumes delivered during mouth-to-face shield (M-FS) breathing, to measure the back pressure and calculate the resistance to flow through their 1-way valves, and to test for backward leak of gas through the valves. METHODS: Three FS brands were evaluated: Kiss of Life (KOL), MicroSHIELD (Micro) and Res-Cue Key (RCK). Volume delivered during M-M and M-FS breathing was evaluated by 10 rescuers who used the devices while performing rescue breathing on a CPR mannequin. Back pressure was measured and resistance calculated by directing airflow through the 1-way valves. Backward leak was evaluated by measuring the O2 concentration at the rescuer side of the valve while 100% O2 was directed toward the patient side of the valve. Differences among the brands were evaluated using analysis of variance. RESULTS: The mean (SD) values for volumes in L were: M-M 1.00 (0.25), Micro 0.77 (0.20), RCK 0.64 (0.10), and KOL 0.24 (0.11). Mean values for back pressure in cm H2O at 50 L/min were Micro 16.7 (1.29), KOL 7.22 (0.13), and RCK 2.15 (0.16). Significant backward leak only occurred with RCK. CONCLUSION: Not one of the FSs tested met all of the requirements suggested by the American Heart Association and by the International Standards Organization.