High intensity ultrasound clamp for bloodless partial nephrectomy: In vitro and in vivo experiments.

In some patients at risk of disease recurrence of renal cancers, maximum conservation of the kidney is possible through partial nephrectomy. However, bloodless surgery is difficult to achieve. The article describes an ultrasonic clamp, which optimises energy deposition and monitors lesion development with an echo-based technique. Using this novel apparatus, coagulation necroses have been obtained in vitro on substantial thicknesses (23 to 38 mm) over exposure durations ranging from 10 s to 130 s, and with acoustic intensities of less than 15 W/cm(2) per transducer. When used for coagulation purposes, two transducers situated on opposite arms of the clamp are driven, while for monitoring, only one is used. Lesions are monitored in real time by analysing the echo signal returned by the opposite arm of the clamp. The presence of a lesion is evaluated on the basis of energy changes and echo phase as a function of time. Both kidneys of two pigs (30 to 36 mm thick) were treated in vivo with the clamp, and the partial nephrectomies performed proved to be bloodless.

Animals↗

Numerical field intensity factor calculations for 1-3 piezocomposite structures.

The 1-3 piezocomposite transducers used in specific medical applications--as lithotripsy--must be excited by intense electrical impulses. By inference, the composite material is subjected to high-stress levels. To explain possible failure, a numerical tool for analyzing singular two- and three-dimensional stress fields in piezocomposite structures is proposed. Following a finite-element iterative method, singularity parameters values and intensity factors for the mechanical stresses and the electrical field are computed in several bimaterial configurations. The finite-element analysis uses commercial software.

Ceramics↗

Intraluminal high intensity ultrasound treatment in the esophagus under fast MR temperature mapping: in vivo studies.

New curative and palliative treatments are needed to respond to the poor prognosis of esophageal cancer. The purpose of this study was to determine whether magnetic resonance imaging (MRI) and MR thermometry can be used to monitor the thermal ablation induced by an intraluminal high-intensity ultrasound applicator positioned in the esophagus. Experiments were performed in vivo in 2 pig esophagi (25 thermal lesions per pig). Respiratory gated or cardiac gated MR thermometry was performed with segmented echo-planar imaging gradient echo sequences. All MR acquisitions were performed without susceptibility artifacts or radiofrequency interference with the ultrasound device. The experimental procedure proposed for accurate measurement of temperature in the esophagus was found to achieve an SD of +/- 1.5 degrees C for respiratory gating and +/- 3.1 degrees C for cardiac gating. Gd-enhanced T(1)-weighted images were used to depict coagulation necrosis. Autopsy was performed immediately after the treatment. Ultrasound effects were inspected visually, and the dimensions of the lesions in the liver neighboring the esophagus were compared with those determined on the MRI images. The visually assessed thermal lesions showed good correlation with the MRI data (10% mean volume difference). The feasibility of esophageal thermal ablation using intraluminal high-intensity ultrasound and of on-line MR temperature monitoring was demonstrated.

Animals↗

Monitoring the formation of thermal lesions with heat-induced echo-strain imaging: a feasibility study.

We investigated the feasibility of using echo-strain images to visualize the extent of high-intensity ultrasound (US)-induced thermal lesions during their formation. Echo-strain, defined as the relative deformation of the backscattered ultrasonic signal, is due to tissue expansion and to changes in the speed of sound during heating. First, a theoretical framework was developed to predict the influence of these effects on the echo signal. Then, a simulation tool was developed to create simulated echo-strain images in thermal lesions. Finally, experimental echo-strain images were acquired in 10 porcine liver samples in vitro for various exposure durations and ultrasonic intensities (resulting in lesions that extended 3 to 8 mm deep from the surface). For this purpose, radiofrequency (RF) frames were acquired at 8 frames per s while heating. For each consecutive pair of RF frames, an echo-strain image was calculated using standard elastographic processing. The echo-strain images were cumulated and displayed. The experimental echo-strain images were compared with gross pathology. The (isoechoic) lesions were visible both in simulated and in experimental cumulated echo-strain images as apparent expansion areas (tensile echo-strain), whereas surrounding tissues exhibited apparent compression. The tensile echo-strain area underestimated the lesion in simulations, but was representative of the lesion in experiments. High correspondence was found between the lesion depth measured from experimental cumulative echo-strain images (y) and from gross pathology (x) (Pearson's correlation = 0.90, linear regression y = x-0.1 mm, residual error = 0 +/- 0.9 mm). We hypothesized that significant tissue expansion made the thermal lesions highly visible in the experimental echo-strain images. In two cases, the ultrasonic intensity was too low to induce a lesion, and the corresponding experimental echo-strain images showed no visible lesion. We conclude that cumulative echo-strain images have the potential to monitor the formation of high-intensity US-induced thermal lesions.

Algorithms↗

[Haemostatic efficacy of a high intensity focused ultrasound applicator in lower pole partial nephrectomy in the pig].

OBJECTIVE: To evaluate the haemostatic property of a high intensity focused ultrasound (HIFU) applicator in lower pole partial nephrectomy in a porcine model. MATERIALS AND METHODS: An applicator with a flat HIFU transducer operating at a frequency of 3.78 MHz was designed. Nine female pigs weighing between 35 and 39 kg were used. The kidneys of 5 pigs were used to characterize basic lesions, study haemostatic efficacy and measure the temperature rise in treated tissues. The following 4 pigs were treated by HIFU prior to bilateral lower pole partial nephrectomy. Treatment consisted of circumferential juxtaposition of basic lesions in the lower pole with vascular clamping. Haemostatic efficacy was evaluated immediately and 30 minutes after lower pole resection. RESULTS: A 50-second ultrasound pulse with an intensity of 26 W/cm2 induced a full thickness lesion of the kidney, between 22 and 36 mm. Lesions were reproducible. The temperature rise observed in treated tissues was 62 +/- 12, 59 +/- 9 and 58 +/- 11 degrees C at distances of 15, 20 and 25 mm from the transducer, respectively. All lower pole partial nephrectomies were exsanguinated, except for one case (12.5%), in which a central renal artery was not controlled. CONCLUSIONS: Our HIFU applicator is an excellent haemostatic tool ensuring effective coagulation of the renal parenchyma, allowing dry lower pole partial nephrectomy. Improvement of the ergonomics of this applicator and adaptation to laparoscopy are currently under investigation.

Animals↗

Comparison of two methods of treatment for intraluminal thermal ablation using an ultrasound cylindrical phased array.

Intraluminal (within the alimentary tract) thermal surgery has been shown to be a useful therapeutic option when extracorporeal focused ultrasound applicators cannot be used since their beam may not reach the target site. If plane transducers are used for the treatment of alimentary tract tumours, the applicator must be rotated in order to generate a cylindrical volume of necrosis. However, rotating these applicators and controlling their shooting direction presents technical difficulties. If tubular transducers are used it is difficult to treat arbitrary angles with a large therapeutic length. To solve these difficulties, the feasibility of an ultrasound phased array applicator has been evaluated using a cylindrical prototype (outer diameter 10.6 mm), which is composed of 16 elementary transducers working at 4.55 MHz and arranged on a quarter of the cylinder. Using this applicator it is possible to generate plane or cylindrical waves. Plane waves were generated by exciting eight successive elements of the array with appropriate delay times. The exposure direction was changed by exciting a different set of eight elements. In this way, the ultrasound beam was electronically rotated through the tissues. Cylindrical waves were generated by exciting several transducers without delay times. Imaging was provided using a miniature echographic probe. Ex vivo experiments were carried out in pig liver to compare two approaches of treatment. The first consisted of generating successive plane waves separated from each other by a 6 degrees angle. The second one consisted of exciting all the 16 elements without delay times. In the two cases, the lesions were well-defined and occupied a quarter of cylinder. In both sets of experiments, the sonication time and the intensity were 20 s and 17 W/cm(2), respectively. In the first case, the depth was up to 17 mm compared to 6 mm in the second case.

Animals↗

Combination of thermal and cavitation effects to generate deep lesions with an endocavitary applicator using a plane transducer: ex vivo studies.

In the high-intensity focused ultrasound (US), or HIFU, field, it is well-known that the cavitation effect can be used to induce lesions of larger volume. The principle is based on the increase in the equivalent attenuation coefficient of the tissue in the presence of the bubbles created by cavitation. The elementary lesions produced by combination of cavitation and thermal effects, using focused transducers, were spherical and developed upstream of the focal point. This paper presents a method that combines cavitation with a thermal effect to obtain deeper lesions using a plane transducer, rather than a focused one. The cavitation effect was produced by delivering intensities of 60 W/cm2 at the face of the transducer for 0.5 s. The applicator was then rotated through 90 degrees at a constant speed of between 0.5 and 1.5 degrees /s. During this rotation, ex vivo tissues were exposed continuously to an acoustic intensity of 14 W/cm2 to combine the cavitation effect with a thermal effect. The necroses were, on average, twice as deep when the cavitation effect was used as those obtained with a thermal effect alone. Observed macroscopically, the lesions have a very well-delimited geometry. Temperature measurements made at different angles of treatment have shown that they were coagulation necroses.

Acoustics↗

Feasibility of a transurethral ultrasound applicator for coagulation in prostate.

Transurethral resection of the prostate (TURP) is the surgical method routinely used in clinics to treat benign prostate hyperplasia (BPH). The purpose of this work is to demonstrate the feasibility of a transurethral ultrasound (US) applicator based on a miniature US flat transducer to coagulate prostatic tissues. Rabbit liver was found to comply well with human prostate. A significant fall in Doppler signal amplitude immediately after treatment demonstrated the applicator's ability to achieve haemostasis. The therapeutic depth extended from 6 to 10 mm, depending on conditions of exposure, and the coagulation rate ranged between 51% and 99%. The coagulated zone pinpointed on histological examination could be easily correlated to a permanent hypoechoic zone observed on B-scans of treated zones. This observation is most likely due to temperature-related changes in the acoustic attenuation of liver and, unfortunately, may not be visible in the prostate.

Animals↗

Determination of the acoustic pressure at and the reflection coefficient of a target through measurements of the absorbed power and the emitter voltage.

A harmonic acoustic wave, fed back by a reflecting target, modifies the electric impedance of the emitter. This effect is studied using a tightly focused beam and various flat targets with known reflection coefficients, which are placed at pressure maxima or minima of the standing wave in the focal zone. It is possible to establish relationships that allow one to determine, for flat targets with unknown acoustic properties, the acoustic impedance and reflection coefficient of the target as well as the acoustic pressure present at the target, only from measurements of the absorbed power and the emitter voltage.

Acoustics↗

Generation of higher pressure pulses at the surface of piezo-composite materials using electrical pre-strain.

New clinical concepts in lithotripsy demand smaller shock heads. Reducing the size of piezoelectric shock heads requires increasing the surface pressure of each transducer so that the total pressure at the focus remains the same. A new method allowing generation of large surface pressures is described. The hypothesis is that piezoelectric plots in piezo-composite material are more fragile in extension than in compression. For this reason, actuators are mechanically prestressed between two flasks. This method cannot be used for transducers working at high frequencies, such as 0.5 MHz. So we tried to electrically prestrain compressively the piezoelectric material by applying a high-electrical field in the opposite direction of polarization. Three protocols were tested and compared to classically driven transducers. In the first protocol, prestrain is permanently applied, in the second protocol prestrain is applied for 100 micros before the compressive impulse, and in the third protocol prestrain is applied for 100 micros and followed by a bipolar field that allows the material to be repoled between two successive pulses. With the two first protocols, rapid depoling and repoling in the opposite direction was noticed. Only with the last protocol was it possible to increase the maximum surface pressure. This increase was approximately the same whether the material was hard or soft. Using this protocol, aging tests were conducted on three samples of each kind of material, and a pressure of 4 MPa was obtained over 10(6) shocks. This value seems to be enough to develop a piezoelectric shock-wave generator with a diameter of approximately 20 cm instead of the 45-cm commercially available.

Journal Article↗

Transoesophageal ultrasound applicator for sector-based thermal ablation: first in vivo experiments.

New curative and palliative treatments must be proposed to respond to the bad long-term prognosis of oesophageal cancers. It has been demonstrated that high intensity ultrasound (US) can induce rapid, complete and well-defined coagulation necrosis. For the treatment of this cancer, we designed an applicator that uses an intraductal approach. The active part is an air-backed plane transducer. It has an external water-cooling system and operates at 10 MHz. Ex vivo experiments conducted on pig liver demonstrated the ability of this applicator to generate, by rotating the transducer, circular or sector-based coagulation necroses at predetermined depths up to 13 mm, with an excellent angular precision. The treatment of sector-based oesophageal tumours may be critical, where both malignant and healthy tissues are covered by the US beam. Thus, in vivo trials were conducted on five healthy pig oesophaguses to determine the maximal thermal dose that will not induce a perforation of the oesophagus or surrounding tissues. From the results of previous studies, this dose is high enough to treat pathological tissues. These promising results indicate that this US system represents a safe and effective tool for the clinical treatment of oesophageal tumours.

Animals↗

Long-lasting stable cavitation.

Stable cavitation is produced on the surface of a special target, i.e., an abrasive foil with a grain size of about 15 microm, insonified by a 473 kHz focused beam. Cavitation bubbles are first created by a pressure of about 2 MPa. Progressive reduction of the pressure to about 100 kPa leads to a state of stable cavitation characterized by strong stable emission of the half-order subharmonic. This state can be maintained for five hours if the defining parameters (position and pressure) are optimized and constant to within a few percent. There is strong evidence for the presence of "latent" cavitation bubbles, which can persist for a few minutes without being excited.

Journal Article↗

Ultrasound cylindrical phased array for transoesophageal thermal therapy: initial studies.

This work was undertaken to investigate the feasibility of constructing a cylindrical phased array composed of 64 elements spread around the periphery (OD 10.6 mm) for transoesophageal ultrasound thermotherapy. The underlying operating principle of this applicator is to rotate a plane ultrasound beam electronically. For this purpose, eight adjacent transducers were successively excited with appropriate delay times so as to generate a plane wave. The exposure direction was changed by exciting a different set of eight elements. For these feasibility studies, we used a cylindrical prototype (OD 10.6 mm) composed of 16 elementary transducers distributed over a quarter of the cylinder, all operating at 4.55 MHz. The active part was mechanically reinforced by a rigid damper structure behind the transducers. It was shown that an ultrasound field similar to that emitted by a plane transducer could be generated. Ex vivo experiments on pig's liver demonstrated that the ultrasound beam could be accurately rotated to generate sector-based lesions to a suitable depth (up to 19 mm). Throughout these experiments, exposures lasting 20 s were delivered at an acoustic intensity of 17 W cm(-2). By varying the power from exposure to exposure, the depth of the lesion at different angles could be controlled.

Esophageal Neoplasms↗

Endoscopic treatment of cholangiocarcinoma and carcinoma of the duodenal papilla by intraductal high-intensity US: Results of a pilot study.

BACKGROUND: Local and infiltrative extension make some biliary carcinomas accessible to nonoperative intraductal destruction. This study assessed the clinical feasibility and short-term results of local tumor destruction with an intraductal high-intensity US probe during ERCP. METHODS: The probe is a flexible catheter with an 8 x 2.8 mm US transducer and a lumen for a guidewire. Ten patients (6 women, 4 men; mean age 74.8 years) were treated with this device. Lesions treated included carcinoma of the papilla (3), bile duct cholangiocarcinoma (2), Bismuth grade I and II hilar cholangiocarcinomas (4), and intrahepatic cystadenocarcinoma (1). Two patients underwent US therapy before surgery. Treatment was performed during standard ERCP: the probe was inserted through the malignant stricture and US therapy was applied over 360 degrees under fluoroscopic control. RESULTS: No serious adverse effects were observed; right upper abdominal pain developed in one patient for 12 hours. In one patient, histopathologic assessment of the resected tumor revealed extensive coagulation necrosis with inflammation up to 10 mm in depth surrounding the bile duct lumen. In the other operated patient, biopsy specimens from the treated portion of the bile duct were negative for malignancy. There was complete regression of cholangiocarcinoma of the bile duct in our patient, allowing for permanent stent removal (follow-up 30 months). A partial response was noted in 4 other patients and no response in 3 patients. CONCLUSIONS: This new method of intraductal tumor destruction by high-intensity US during ERCP is feasible and can induce objectively measurable tumor necrosis. Long-term follow-up will determine whether this method is curative in some cases and if it can reduce the need for biliary stent placement.

Aged↗

Numerical simulation of the electro-acoustical response of a transducer excited by a time-varying electrical circuit.

Existing methods for the modeling of piezoelectric transducer response are generally frequency domain-based. The major disadvantage of this type of model is that they cannot take into account the electrical elements present in the emitting or receiving circuit whose values vary with respect to time. The need for a method that accounts for time-varying elements arises, for example, when the circuit comprises active electrical elements, such as diodes, or when the transducer is excited by capacitive discharge via a switch. Indeed, in this last example, it is known that the output impedance of the generator depends on the state of the switch: if it is off, its value is high; if it is on, its value is low. A time domain-based method is presented to compute the electro-acoustical response of a piezoelectric transducer and its electrical circuit, taking into account the presence of time-varying elements. An application to a current example makes it possible to show the influence of these elements on waveforms and the capacity of our model to account for them.

Journal Article↗

Damping, amplitude, aging tests of stacked transducers for shock wave generation.

New clinical concepts in lithotripsy demand small shock heads. Reducing the size of piezoelectric shock heads will be possible only if the pressure generated at the surface of each transducer can be increased so that the total pressure at the focus remains the same. To solve this problem, different solutions were proposed. For example, it has been demonstrated that piezocomposite material, as opposed to piezoceramic material, allows the generation of a higher surface pressure before breaking, mainly because radial modes are dramatically reduced. In addition, in a previous paper, we showed the feasibility of generating high-pressure pulse waves without increasing the transducer voltage by using sandwiched transducers, which are a stack of two or more transducers. Some discrepancies appeared, however, between the pressure measured at the surface of the front transducer and the arithmetic sum of the pressures generated by each transducer constituting the stack. In fact, development of such stacked transducers capable of generating surface pressures in the range of 2 to 5 MPa is very complex, which may explain why no aging tests have been reported in the literature thus far. In the first part of this paper, we theoretically determine the importance of the electroacoustical coupling between the two transducers on the generated surface pressure. We show that pressure losses due to these electroacoustical couplings are less than 5%. Experimental measurements done on a stacked transducer assembled and tightened in a castor oil-filled tank are in excellent accordance with the theoretical measurements. Using this assembly technique, it was possible to obtain, on average, out of four elements, a pressure of 7.5 MPa for the duration of 4 million shocks, which would allow the treatment of approximately 1000 patients.

Computer Simulation↗

Optimizing the shape of ultrasound transducers for interstitial thermal ablation.

Heat deposition by interstitial routes, especially with ultrasound-based instruments, is becoming a valuable therapeutic option for the treatments of sites, which are difficult to access from outside of the body. The active part of most interstitial ultrasound applicators described in the literature is logically tubular to induce cylindrical volumes of coagulation necrosis. Because the pressure generated by such tubular transducers falls off rapidly with radial distance, we previously proposed using a rotating plane transducer. For a plane wave, the pressure fall-off is only due to attenuation, which makes deeper lesions and shorter treatment times possible. This work represents an advance in the development of ultrasound applicators designed for interstitial applications. This new applicator used a rotating slightly focused transducer. A brief theoretical analysis resulted in the choice of a long focal distance of 22 mm to obtain a nearly constant pressure all along the therapeutic depth. To experimentally validate this focal distance, pressure measurements were made in a tissue mimicking liquid phantom and the results were compared with those obtained with a plane transducer. In vitro experiments showed that necrosis could be induced at a depth of 15 mm. In the same conditions, the greatest depth attained with a plane transducer was only 10 mm. Because each individual lesion is narrower, more lesions and more time are required to necrose a cylindrical volume. The main advantage of this new type of applicator is that it can be used to induce necrosis at a greater depth without varying either the frequency, the intensity or the transducer cooling efficiency.

Acoustics↗