PubMed HealthSearch

Biomedical subjects

E K Hillard

Publications and source records attributed to E K Hillard.

12 recordsLinked to original sources

A new approach to artificial expansion and ventilation of the lung in the severely asphyxiated neonate.

In severe asphyxia, expansion of the newborn lung must precede ventilation by intermittent positive pressure. Inadequate expansion may lead to hypoxia and excessive expansion to rupture of the lung. The only way in which a predetermined pressure can be achieved, and not exceeded in the alveoli, is by applying that pressure at the mouth and waiting until all flow ceases. Applying pressure in this way and increasing it according to a "staircase" pattern, with appropriate steps at suitable intervals, should ensure that the maximum pressure produced in the alveoli is no greater than the minimum necessary for resuscitation in each individual infant. On the basis of published work, ranges of optimum values for the increments and time intervals are suggested. Clinical judgement is still necessary to select from within these ranges, but this "pressure staircase" method should provide a systematic approach to the problem. A suitable apparatus is described.

Asphyxia Neonatorum

Heat losses from a breathing system with a heated-water humidifier.

Air was "breathed" in the laboratory through a heated-water humidifier and a breathing tube. Several different humidifiers and tubes were used. The temperature rise of the air on passing through the humidifier and the temperature drop on passing through the tube were measured. Both were dependent on ventilation. Insulating the tube and humidifier together with the insertion of baffles in the latter reduced the rise and fall and their dependence on ventilation. With suitable design the dependence on ventilation and the need to use high water temperatures could be greatly reduced. In addition, a thermostat with a reduced dead zone is needed.

Air

The flomasta. A new anaesthetic ventilator.

A new ventilator for use during anaesthesia is described. It operates as a minute-volume divider and derives its power from the energy stored in a distended reservoir bag. The unit is autoclavable, suitable for use with explosive anaesthetics and has a rational control system which avoids many of the hazards associated with other ventilators when the mode of ventilation is changed. The nature of the inspired gas mixture is known accurately because the internal volume of the ventilator is small and the system is non-rebreathing. The design of the exhaust system facilitates ventilation monitoring, the use of positive end-expiratory pressure and connection to ducted pollution-control systems.

Airway Resistance