Development of a new insertion technique and a new device for the placement of bifurcated airway stents.

We developed a new insertion technique and designed a forceps device for the placement of bifurcated airway stents; 131 of 142 endoscopically placed tracheobronchial Y-stents were inserted with a forceps and a laryngoscope. For the last 52 stent implantations we used the new stent forceps. It was determined to be a simple and safe method without major complications. In 11 cases alternative techniques had to be used. Technique and device are described in detail.

Bronchoscopy

Three years experience with a new balloon catheter for the management of haemoptysis.

For the management of severe haemoptysis we have developed a double-lumen, bronchus-blocking catheter that can be introduced through the working channel of a standard fibrebronchoscope. We wondered whether this catheter would be suitable to control pulmonary haemorrhage in clinical practice. Over a period of 36 months, 30 of these catheters were used in 27 patients with moderate and massive pulmonary bleeding from various lesions. Underlying diseases were: malignancies (11), vascular deformities (5), tuberculosis (4), silicosis (2), carcinoids (2), silicosis (2), endometriosis (1), bronchiectasis (1). In 26 cases, the transbronchoscopic balloon tamponade was successful. In one patient, tumour growth close to the carina prevented securing of the balloon and double-lumen tube intubation was required. There were only minor complications attributable to the balloon. With the catheter in place for up to seven days, patients underwent surgery, received radiation, chemotherapy, drug treatment or bronchial arterial embolization. In conclusion, we found this double-lumen, bronchus-blocking device safe and the technique practicable to control pulmonary haemorrhage.

Balloon Occlusion