Preclinical hemodynamic assessment of a new trileaflet mechanical valve in the aortic position in a bovine model.

BACKGROUND AND AIM OF THE STUDY: The hemodynamic characteristics of a new trileaflet mechanical heart valve (TTV) (TriFlo Inc., Costa Mesa, CA, USA) in the aortic position were evaluated in a bovine model. The TTV was designed to combine the durability of mechanical heart valves with the central flow characteristics of a bioprosthesis. METHODS: Using nine calves, the native aortic valve was replaced with a 21-mm TTV (n = 6) or a St. Jude Medical (SJM) valve (n = 3). Hemodynamic values were assessed with echocardiography at implantation and with catheterization at explantation. All calves underwent a necropsy, followed by gross pathology and light microscopy studies. RESULTS: The mean implant duration was 159 +/- 55 days for the TTV and 102 +/- 67 days for the SJM valve. Immediately before euthanasia, the peak and mean pressure gradients were respectively 35 +/- 14 and 24 +/- 9 mmHg for the TTV, and 100 +/- 72 and 59 +/- 38 mmHg for the SJM valve (p = 0.03). Two of the SJM-valve animals were electively sacrificed after showing symptoms of aortic stenosis. At necropsy, all nine valves were free from thrombi, pannus, occlusive tissue, or mechanical impairment of leaflet motion. The prematurely sacrificed SJM-valve animals had concentric myocardial hypertrophy consistent with severe functional aortic stenosis. CONCLUSION: The significantly higher gradients and left ventricular hypertrophy in the SJM-valve animals were related to a smaller effective orifice area, which precluded adaptation to increasing transvalvular volumes in the growing animal. This problem was not seen with the TTV, which performed hemodynamically as well as the SJM valve. The high transvalvular gradients seen with the SJM valve at study end may suggest that the hemodynamic characteristics of the TTV may be superior, though additional studies are needed to confirm this.

Animals↗

Transmyocardial laser revascularization as an adjunct to coronary artery bypass grafting: a randomized, multicenter study with 4-year follow-up.

We evaluated transmyocardial laser revascularization (TMLR) with coronary artery bypass grafting (CABG) versus CABG alone for severe coronary artery disease involving 21 myocardial region unsuited for CABG. At 4 centers, 44 consecutive patients were randomized for CABG+TMLR (n = 23) or CABG alone (n = 21). Operative and in-hospital mortality and morbidity rates were monitored. Clinical status was evaluated at hospital discharge, 1 year, and 4 years. Success was characterized by relief of angina and freedom from repeat revascularization and death. Preoperatively, 20 patients (47%) were at high risk. The CABG technique, number of grafts, and target vessels were similar in both groups. Patients undergoing CABG+TMLR received 25 +/- 11 laser channels. Their < or = 30-day mortality was 13% (3/23) compared with 28% (6/21) after CABG alone (P = 0.21). There were no significant intergroup differences in the number of intraoperative or in-hospital adverse events. The follow-up period was 50.3 +/- 17.8 months for CABG alone and 48.1 +/- 16.8 months for CABG+TMLR. Both groups had substantially improved angina and functional status at 1 and 4 years, with no significant differences in cumulative 4-year mortality. The incidence of repeat revascularization was 24% after CABG alone versus none after CABG+TMLR (P < 0.05). The 4-year event-free survival rate was 14% versus 39%, respectively (P < 0.064). In conclusion, CABG+TMLR appears safe and poses no additional threat for high-risk patients. Improved overall success and repeat revascularization rates may be due to better perfusion of ischemic areas not amenable to bypass. Further studies are warranted to determine whether these trends are indeed significant.

Aged↗

Ex vivo resuscitation of adult pig hearts.

One possible way to expand the human heart donor pool is to include non-heart-beating human donors. To begin validating this approach, we developed an ex vivo cardiac perfusion circuit to support large mammalian hearts in Langendorff mode and beating-ejecting mode and to assess and improve their ischemic tolerance. In vivo hemodynamic data and heparinized blood (4.0 +/- 0.5 L) were collected from 6 anesthetized pigs. Hearts were isolated and connected to a recirculating perfusion circuit primed with autologous buffered blood (pH, 7.40). After retrograde aortic perfusion in Langendorff mode, the left atrium was gravity-filled at 10-20 mmHg, and the left ventricle began to eject against a compliance chamber in series with a systemic reservoir set to a hydraulic afterload of 100-120 mmHg. Left ventricular function was restored and maintained in all 6 hearts for 30 min. Cardiac output, myocardial oxygen consumption, stroke work, aortic pressure, left atrial pressure, and heart rate were measured. The mean myocardial oxygen consumption was 4.8 +/- 2.7 mL/min/100 g (95.8% of in vivo value); and mean stroke work, 5.3 +/- 1.1 g x m/100 g (58.95% of in vivo value). One resuscitated heart was exposed to 30 min of normothermic ischemic arrest, then flushed with Celsior and re-resuscitated. The ex vivo perfusion method described herein restored left ventricular ejection function and allowed assessment of ischemic tolerance in large mammalian hearts, potentially a 1st step toward including non-heart-beating human donors in the human donor pool.

Animals↗

Biventricular support with the Jarvik 2000 axial flow pump: a feasibility study.

Patients with congestive heart failure who are supported with a left ventricular assist device (LVAD) may experience right ventricular dysfunction or failure that requires support with a right ventricular assist device (RVAD). To determine the feasibility of using a clinically available axial flow ventricular assist device as an RVAD, we implanted Jarvik 2000 pumps in the left ventricle and right atrium of two Corriente crossbred calves (approximately 100 kg each) by way of a left thoracotomy and then analyzed the hemodynamic effects in the mechanically fibrillated heart at various LVAD and RVAD speeds. Right atrial implantation of the device required no modification of either the device or the surgical technique used for left ventricular implantation. Satisfactory biventricular support was achieved during fibrillation as evidenced by an increase in mean aortic pressure from 34 mm Hg with the pumps off to 78 mm Hg with the pumps generating a flow rate of 4.8 L/min. These results indicate that the Jarvik 2000 pump, which can provide chronic circulatory support and can be powered by external batteries, is a feasible option for right ventricular support after LVAD implantation and is capable of completely supporting the circulation in patients with global heart failure.

Animals↗

Myocardial hemodynamics, physiology, and perfusion with an axial flow left ventricular assist device in the calf.

The Jarvik 2000 axial flow left ventricular assist device (LVAD) is used clinically as a bridge to transplantation or as destination therapy in end-stage heart disease. The effect of the pump's continuous flow output on myocardial and end-organ blood flow has not been studied experimentally. To address this, the Jarvik 2000 pump was implanted in eight calves and then operated at speeds ranging from 8,000 to 12,000 rpm. Micromanometry, echocardiography, and blood oxygenation measurements were used to assess changes in hemodynamics, cardiac dimensions, and myocardial metabolism, respectively, at different speeds as compared with baseline (pump off, 0 rpm) in this experimental model. Microsphere studies were performed to assess the effects on heart, kidney, and brain perfusion at different speeds. The Jarvik 2000 pump unloaded the left ventricle and reduced end-diastolic pressures and left ventricular dimensions, particularly at higher pump speeds. The ratio of myocardial oxygen consumption to coronary blood flow and the ratio of subendocardial to subepicardial blood flow remained constant. Optimal adjustment of pump speed and volume status allowed opening of the aortic valve and contribution of the native left ventricle to cardiac output, even at the maximum pump speed. Neither brain nor kidney microcirculation was adversely affected at any pump speed. We conclude that the Jarvik 2000 pump adequately unloads the left ventricle without compromising myocardial metabolism or end-organ perfusion.

Animals↗

Baseline hemodynamic and echocardiographic indices in anesthetized calves.

The experimental calf model is used to assess mechanical circulatory support devices and prosthetic heart valves. Baseline indices of cardiac function have been established for the normal awake calf but not for the anesthetized calf. Therefore, we gathered hemodynamic and echocardiographic data from 16 healthy anesthetized calves (mean age, 189.0 +/- 87.0 days; mean body weight, 106.9 +/- 32.3 kg) by cardiac catheterization and noninvasive echocardiography, respectively. Baseline hemodynamic data included heart rate (65 +/- 12 beats per minute), mean aortic pressure (113.5 +/- 17.4 mm Hg), left ventricular end-diastolic pressure (16.3 +/- 38.9 mm Hg), and mean pulmonary artery pressure (21.7 +/- 8.3 mm Hg). Baseline two-dimensional echocardiographic data included left ventricular systolic dimension (3.5 +/- 0.7 cm), left ventricular diastolic dimension (5.6 +/- 0.8 cm), end-systolic intraventricular septal thickness (1.7 +/- 0.2 cm), end-diastolic intraventricular septal thickness (1.2 +/- 0.2 cm), ejection fraction (63 +/- 10%), and fractional shortening (37 +/- 10%). Doppler echocardiography revealed a maximum aortic valve velocity of 0.9 +/- 0.5 m/s and a cardiac index of 3.7 +/- 1.1 L/minute/m2. The collected baseline data will be useful in assessing prosthetic heart valves, cardiac assist pumps, new cannulation techniques, and robotics applications in the anesthetized calf model and in developing calf models of various cardiovascular diseases.

Anesthetics, Dissociative↗

Thrombogenicity of mechanical aortic valves in an animal model: site specific testing is crucial.

We evaluated a new trileaflet prosthesis and a control bileaflet prosthesis in the mitral and aortic positions in 27 calves. The prototype trileaflet valve (TV1) functioned satisfactorily in the mitral position (TV1m, n = 7) but later yielded thrombogenic complications in the aortic position (TV1a, n = 4). The valve was redesigned (TV2) and retested in the mitral (TV2m n = 4) and aortic (TV2a, n = 5) positions, along with control valves (Cm, n = 4; Ca, n = 3). At necropsy, the valves were graded on a scale of 0 (no visible thrombi) to 4 (thrombi greater than 5 mm and/or obstructed leaflets). The TV1m, TV2m, and Cm animals, respectively, had implant durations of 215+/-112, 140+/-63, and 159+/-89 days and thrombus grades of 0.71+/-0.76, 0.33+/-0.58, and 1.50+/-0.58. The TV1a, TV2a, and Ca animals had implant durations of 18+/-12, 159+/-61, and 108+/-62 days and thrombus grades of 2.75+/-1.00, 0.50+/-0.58, and 0.67+/-0.58 (p < .005; TV2a vs. TV1a). Three TV1a calves died early of valve related complications. A design irregularity, undetected in the mitral position but revealed in the aortic position, caused a high early mortality in the TV1a animals. Redesigning the prosthesis eliminated valve related mortality and significantly reduced the thrombus grade. Because satisfactory performance in the mitral position does not guarantee safety or efficacy in the aortic position, site specific preclinical testing is crucial for mechanical heart valves.

Animals↗

Evaluation of a new cardiac recovery system in a bovine model of volume overload heart failure.

We evaluated the short-term hemodynamic efficacy of the Cancion cardiac recovery system (CRS) in a bovine model of volume overload heart failure. We created severe mitral regurgitation (MR) by disrupting the mitral chordae tendineae of two calves, which were allowed to survive for 13 and 11 months. Four hours before we killed the calves, we introduced the CRS, which maintained flows of 1.2 L/min. Hemodynamic data were recorded before and after MR creation and CRS initiation. Left ventriculography was performed at those same intervals and 4 hours after CRS initiation. After chordal disruption, the left ventricular end-diastolic pressure (LVEDP) increased from 14 to 23 mm Hg in calf 1 and 15 to 27 mm Hg in calf 2. After 4 hours of CRS support, the LVEDP decreased from 23 to 8 mm Hg (calf 1) and 27 to 14 mm Hg (calf 2); dP/dt increased from 1,373 to 2,900 mm Hg/s (calf 1) and 1,068 to 2,384 mm Hg/s (calf 2). MR decreased from 3+ to 1 in calf 2 but could not be determined in calf 1. In this bovine model of volume overload heart failure, the CRS unloaded the left ventricle and reduced the afterload. Future trials will determine the pump's ability to treat congestive heart failure.

Animals↗

Effect of pump flow mode of novel left ventricular assist device upon end organ perfusion in dogs with doxorubicin induced heart failure.

End organ effects of nonpulsatile (NP) and pulsatile (P) left ventricular assist device (LVAD) flow were compared in a canine model of doxorubicin-induced heart failure. After heart failure induction, a prototype bimodal LVAD was implanted. Hemodynamics, cardiac dimensions, and myocardial metabolism were monitored with the LVAD off (baseline) and on (in NP and P modes at 70% or 100% power). End organ perfusion was assessed by colored microsphere analysis. Seven dogs were used: two died before pump implantation and were excluded from analysis, and the remaining five survived to study termination. At 70% NP, ascending aortic flow and myocardial oxygen consumption (MVO2) decreased significantly. At 100% NP, LV dimensions decreased, aortic systolic, pulse, and LV pressures decreased but not significantly, and ascending aorta flow reversed. At 100% NP, coronary blood flow, MVO2, and LV free wall subepicardial and subendocardial blood flows decreased significantly. However, as NP support increased, the subepicardial/subendocardial blood flow ratio remained near baseline. At 100% NP, right ventricular perfusion decreased but not significantly, cerebral perfusion decreased significantly, and renal perfusion stayed constant. P mode results were similar, except that ascending aorta flow decreased significantly at 100% P instead of reversing as at 100% NP. These results suggest that end organ perfusion is not differentially affected by LVAD flow mode during chronic heart failure.

Animals↗

The effect of intermittent low speed mode upon aortic valve opening in calves supported with a Jarvik 2000 axial flow device.

We assessed the effects of an axial flow left ventricular assist device (LVAD) upon aortic valve opening, pump outflow, and biologic and hematologic parameters when operated in intermittent low speed (ILS) mode. An ILS controller equipped Jarvik 2000 LVAD was implanted in six calves. Pump speed was maintained at 10,000 rpm, and pump outflow was measured throughout the study period (71 +/- 6 days [mean +/- SD]). Hematologic and biochemical parameters were analyzed daily for the first 10 days, weekly for the first month, and biweekly thereafter to monitor for kidney or liver dysfunction, hemolysis, bleeding, or infection. Before study termination, esmolol hydrochloride was infused to induce low cardiac output and totally impair aortic valve opening. Radiopaque cineaortography was performed over 30 second intervals (10 seconds before, 10 seconds during, and 10 seconds immediately after ILS controller activation) to assess the effect of ILS mode upon aortic valve opening. After study termination, major end organs and the major vascular tree were removed and examined macroscopically and histologically for thrombus formation and infarction; the aortic valve was examined for thickening and fusion. All pumps were explanted and examined for thrombus formation. All six calves recovered without surgical or mechanical complications. Hematologic and biochemical parameters did not change significantly between baseline and study termination. The aortic valve successfully opened when ILS mode was activated, even under low cardiac output conditions. No thrombus was detected in the major end organs and vascular tree, except for some small renal infarcts in three calves that did not affect renal function. These results indicate that operating an axial flow LVAD in ILS mode allows aortic valve opening and aortic root washout.

Animals↗

Total heart replacement with dual centrifugal ventricular assist devices.

In an ovine feasibility study, we implanted two HeartMate-III centrifugal ventricular assist devices (VADs) for total heart replacement. With cardiopulmonary bypass support, both ventricles were transected at the atrioventricular groove, preserving a rim of ventricular tissue. The atrioventricular valves were excised, and the aorta and pulmonary artery were transected above the ventriculoarterial valves. An interatrial septal window was created by excising the foramen ovale. The VADs' sewing rings were attached to the left and right ventricular remnants, respectively. Outflow grafts were anastomosed to the aorta and pulmonary artery. The left VAD operated continuously at 4,500 rpm. Right VAD speed increased from 2,000 to 4,500 rpm in 500 rpm increments. Outflow graft flow, pressure, oxygen saturation, and shunt direction were recorded. The pulmonary artery to aortic ratio of flow and pressure increased from 0.26 and 0.15 (at 2,000 rpm) to 1.21 and 0.53, respectively (at 4,500 rpm). The interatrial shunt, which was right to left at lower right VAD speeds, progressed to bidirectional, then to left dominant as right VAD speed increased. Outflow-graft oxygen saturation was reflective of the shunt direction. In this acute experiment, total heart replacement with continuous flow VADs satisfactorily balanced left and right ventricular flows and preserved the physiologic circulatory response.

Animals↗

Total heart replacement using dual intracorporeal continuous-flow pumps in a chronic bovine model: a feasibility study.

Continuous-flow pumps are small, simple, and respond physiologically to input variations, making them potentially ideal for total heart replacement. However, the physiological effects of complete pulseless flow during long-term circulatory support without a cardiac interface or with complete cardiac exclusion have not been well studied. We evaluated the feasibility of dual continuous-flow pumps as a total artificial heart (TAH) in a chronic bovine model. Both ventricles of a 6-month-old Corriente crossbred calf were excised and sewing rings attached to the reinforced atrioventricular junctions. The inlet portions of 2 Jarvik 2000 pumps were positioned through their respective sewing rings at the mid-atrial level and the pulseless atrial reservoir connected end-to-end to the pulmonary artery and aorta. Pulseless systemic and pulmonary circulations were thereby achieved. Volume status was controlled, and systemic and pulmonary resistance were managed pharmacologically to keep mean arterial pressures at 100+/-10 mmHg (systemic) and 20+/-5 mmHg (pulmonary) and both left and right atrial pressures at 15+/-5 mmHg. The left pump speed was maintained at 14,000 rpm and its output autoregulated in response to variations in right pump flow, systemic and pulmonary pressures, fluid status, and activity level. Hemodynamics, end-organ function, and neurohormonal status remained normal. These results suggest the feasibility of using dual continuous-flow pumps as a TAH.

Animals↗