PubMed Health⌕ Search

Biomedical subjects

F Cumhur Oner

Publications and source records attributed to F Cumhur Oner.

18 recordsLinked to original sources

Analysis of ectopic and orthotopic bone formation in cell-based tissue-engineered constructs in goats.

Despite decades of extensive research, the application of cell-based bone tissue engineering in clinically relevant models remains challenging. To improve effectiveness, a better understanding of how the technique should work is crucial. In the current study, we investigated the onset time, rate, location and direction of bone formation in ectopically and orthotopically implanted clinically sized tissue-engineered constructs to gain insight the mechanism behind it. Bone marrow stromal cells (BMSCs) were obtained from 10 goats, culture expanded and cryopreserved. Porous biphasic calcium phosphate (BCP) disks of 17mmx6mm were per-operatively seeded with BMSCs or left empty. Both conditions were implanted intramuscularly and in bilateral critical-sized iliac wing defects. Fluorochromes were administered at 3, 5 and 7 weeks and samples were retrieved after 9 weeks. Histology showed abundant and homogeneous bone formation throughout the intramuscular BMSC samples and little bone in the controls. Histomorphometry and measurements of the fluorochrome labels of the ectopical BMSC samples indicated that osteogenesis started at the periphery and subsequent osteoconduction filled the whole scaffold within 7 weeks. In the orthotopically implanted disks, there was good integration with the surrounding bone, but minimal bone in the center of the implants, in both conditions. Bone was only derived from the interface with the surrounding bone, there was no early bone at the surfaces in contact to soft tissue as was seen in the ectopical samples. Apparently cell survival was minimal and insufficient for relevant additional bone formation. However, the speed of integration with surrounding bone and subsequent bone apposition on the BMSC-seeded orthotopic scaffolds were found to be significantly enhanced, which may be relevant especially in challenging environments.

Animals↗

Timing of thoracic and lumbar fracture fixation in spinal injuries: a systematic review of neurological and clinical outcome.

A systematic review of all available evidence on the timing of surgical fixation for thoracic and lumbar fractures with respect to clinical and neurological outcome was designed. The purpose of this review is to clarify some of the controversy about the timing of surgical fracture fixation in spinal trauma. Better neurological outcome, shorter hospital stay and fewer complications have been reported after early fracture fixation. But there are also studies showing no difference in neurological outcome when compared to late treatment. Mortality is another controversial point since a recent report of higher mortality in early treated patients. A systematic review of the literature was preformed. Ten articles were included. Early fracture fixation is associated with less complications, shorter hospital and ICU stay. The effect of early treatment on the neurological outcome remains unclear due to the contradictory results of the included studies. Early thoracic and lumbar fracture fixation results in improvement of clinical outcome, but the effect on neurological outcome remains controversial.

Fracture Fixation, Internal↗

Analysis of preexistent vertebral rotation in the normal quadruped spine.

STUDY DESIGN: In this CT study, vertebral rotation was analyzed in the transverse plane of the normal, nonscoliotic canine spine with a computer-based measurement method. OBJECTIVES: To determine if a rotational pattern exists in the normal, nonscoliotic quadruped spine, similar to what is seen in humans. SUMMARY OF BACKGROUND DATA: Idiopathic scoliosis does not occur in quadrupeds. In humans, the normal, nonscoliotic spine shows a preexistent pattern of vertebral rotation, which corresponds to the most prevalent curve types of idiopathic scoliosis. Since this rotational tendency has only been demonstrated in humans, it is not clear if it can be considered as a part of the pathogenesis of idiopathic scoliosis or as a normal anatomic feature. METHODS: CT scans of the thorax of 42 dogs without clinical or radiologic evidence of scoliosis were used to measure axial vertebral rotation from T1-T13 with a previously developed computer-based CT measurement method. RESULTS: The results of this study demonstrated a predominant rotation to the right of the upper, mid, and lower thoracic vertebrae of the normal canine spine. The mean vertebral rotation angles differed significantly from zero degrees rotation at level T1, from level T4-T7, and from T11-T13. CONCLUSIONS: The normal spine of quadrupeds shows rotation of the thoracic vertebrae with a preferred direction to the right, similar to what is seen in humans. Since idiopathic scoliosis does not exist in quadrupeds, this preexistent rotation seems to be a physiologic process in normal spinal development, independent of the pathogenesis of scoliosis.

Animals↗

Cement augmentation techniques in traumatic thoracolumbar spine fractures.

STUDY DESIGN: Review of human cadaveric and in vivo animal studies and clinical trial. OBJECTIVE: To develop less invasive surgical techniques for reconstruction of the anterior column in thoracolumbar fractures. SUMMARY OF BACKGROUND DATA: Persistent central endplate depression can cause anterior column insufficiency after posterior surgery for traumatic thoracolumbar fractures. Reduction of the central endplate followed by intravertebral cement augmentation could restore weight-bearing capacity. MATERIALS AND METHODS: In human cadaveric burst fracture models, balloon-assisted endplate reduction (BAER) and vertebroplasty techniques have been investigated in terms of their safety and biomechanical properties. The histologic properties of different cement polymers were studied in an animal vertebral body and endplate defect model. In addition, the clinical outcome of percutaneous cement augmentation in the setting of a burst fracture examining the BAER technique and vertebroplasty with adjunctive posterior pedicle screw fixation is reviewed. RESULTS: These techniques have proven to be safe and effective, although cement leakage outside the confines of the vertebral body may occur. Calcium phosphate cements are preferable over methylmethacrylate because of their in vivo histologic properties. Using the BAER technique and posterior pedicular fixation, anterior vertebral height restoration is possible. Following balloon removal, some loss of fracture height restoration is observed. Further loss of vertebral height reduction was not observed following cement curing clinically. CONCLUSIONS: These studies show that less invasive anterior vertebral reconstruction using percutaneous cement augmentation techniques is feasible following traumatic vertebral fractures.

Animals↗

A new in vivo screening model for posterior spinal bone formation: comparison of ten calcium phosphate ceramic material treatments.

This study presents a new screening model for evaluating the influence of multiple conditions on the initial process of bone formation in the posterior lumbar spine of a large animal. This model uses cages designed for placement on the decorticated transverse process of the goat lumbar spine. Five conduction channels per cage, each be defined by a different material treatment, are open to both the underlying bone and overlying soft tissue. The model was validated in ten adult Dutch milk goats, with each animal implanted with two cages containing a total of ten calcium phosphate material treatments according to a randomized complete block design. The ten calcium phosphate ceramic materials were created through a combination of material chemistry (BCP, TCP, HA), sintering temperature (low, medium, high), calcination and surface roughness treatments. To monitor the bone formation over time, fluorochrome markers were administered at 3, 5 and 7 weeks and the animals were sacrificed at 9 weeks after implantation. Bone formation in the conduction channels was investigated by histology and histomorphometry of non-decalcified sections using traditional light and epifluorescent microscopy. According to both observed and measured bone formation parameters, materials were ranked in order of increasing magnitude as follows: low sintering temperature BCP (rough and smooth) approximately medium sintering temperature BCP approximately = TCP > calcined low sintering temperature HA > non-calcined low sintering temperature HA > high sintering temperature BCP (rough and smooth) > high sintering temperature HA (calcined and non-calcined). These results agree closely with those obtained in previous studies of osteoconduction and bioactivity of ceramics thereby validating the screening model presented in this study.

Animals↗

Anterior spinal column augmentation with injectable bone cements.

A vertebral fracture, whether originating from osteoporosis or trauma, can be the cause of pain, disability, deformation and neurological deficit. The treatment of vertebral compression fractures has, for many years until the advent of vertebroplasty, consisted of bedrest and analgesics. Vertebroplasty is a percutaneous technique during which bone cement is injected in a vertebral body to provide immediate pain relief by stabilization. Inflatable bone tamps can, prior to the injection of cement, be used to create a void in the vertebral body, in which case the technique is known as balloon vertebroplasty (or kyphoplasty). The chance of extracorporal cement leakage is smaller for balloon vertebroplasty than for vertebroplasty. Some authors also claim to have gained some correction in vertebral body height or angulation. Both interventions can be used for several indications, including osteoporotic compression fractures and osteolytic lesions of the vertebral body such as myeloma, hemangioma or metastasis, and also for traumatic burst fractures in combination with pedicle screw instrumentation. Polymethyl methacrylate cement is the bone void filler that is used most frequently, although the application of calcium phosphate cements has been studied widely in vitro, in vivo and also in small-scale clinical series. The clinical results of (balloon-) vertebroplasty are favorable with 85-95% of all patients experiencing immediate and long-lasting relief of pain. Serious complications are relatively rare but include neurological deficit and pulmonary embolism. In this paper, both vertebroplasty and balloon vertebroplasty and their respective indications, techniques and results are described in relation with the application and limitations of permanent and resorbable injectable bone cements.

Absorbable Implants↗

The reduction of endplate fractures during balloon vertebroplasty: a detailed radiological analysis of the treatment of burst fractures using pedicle screws, balloon vertebroplasty, and calcium phosphate cement.

STUDY DESIGN: In a human cadaveric burst fracture model, the amount of endplate fracture reduction after posterior instrumentation and balloon vertebroplasty was investigated quantitatively. OBJECTIVES: To assess, in a burst fracture model, the vertebral body and adjacent disc heights, in parallel sagittal planes with 3-dimensional (3D) rotational x-ray imaging, at various phases during pedicle screw fixation and subsequent balloon vertebroplasty. SUMMARY OF BACKGROUND DATA: In recent human cadaveric thoracolumbar fracture studies, it was found that vertebral body height could be restored significantly with inflatable bone tamps. However, limited quantitative data exist on the amount of fracture reduction that can be achieved and how much of the reduction will be lost after deflation and removal of the bone tamps before the cement is injected. METHODS: Twenty burst fractures were created and balloon vertebroplasty with calcium phosphate cement was performed after pedicle screw instrumentation. A 3D dataset was obtained during the following phases: intact, fractured, after reduction and stabilization with pedicle screws, after inflation of the balloons, after deflation and removal of the balloons, after injection of the cement. The fractured vertebral body and adjacent disc heights were measured from five reconstructed sagittal images and compared for the six phases of the procedure. Furthermore, the difference between the vertebral body height centrally and peripherally was calculated. RESULTS: The mean vertebral body height at the thoracic level was Tintact = 19.5 +/- 2.2 mm, Tfractured = 14.6 +/- 3.8 mm, Treduction = 17.3 +/- 2.2 mm, Tinflation = 20.1 +/- 2.0 mm, Tdeflation = 18.0 +/- 2.0 mm, and Tcement = 17.8 +/- 1.8 mm. The overall change in vertebral body height between these phases was significant (P < 0.001). At the lumbar level the mean vertebral body height was Tintact = 23.2 +/- 3.8 mm, Tfractured = 14.7 +/- 3.0 mm, Treduction = 18.4 +/- 2.5 mm, Tinflation = 23.2 +/- 3.5 mm, Tdeflation = 19.3 +/- 2.3 mm, and Tcement = 20.2 +/- 2.8 mm. The overall change in MCVBH between these phases was also significant (P < 0.001). The increase in vertebral body height resulted in a decrease of the adjacent disc height. No difference was found for the amount of endplate reduction in the center or at the periphery. No leakage of cement was detected in the spinal canal. CONCLUSIONS: Reduction of endplate fractures, both in the center and at the periphery, seems feasible and safe with combined fracture reduction and balloon vertebroplasty. The endplate fracture reduction that was gained by inflation of the bone tamps could not be maintained after deflation.

Aged↗

Bone displacement and the role of longitudinal ligaments during balloon vertebroplasty in traumatic thoracolumbar fractures.

STUDY DESIGN: In a human cadaveric burst fracture model with and without longitudinal ligament damage, the amount of anterior and posterior bone displacement (ABD, PBD) during balloon vertebroplasty after pedicle-screw instrumentation was investigated quantitatively. OBJECTIVES: To investigate, in a burst fracture model with and without longitudinal ligament damage, the amount of ABD, PBD, and cement leakage at various phases during balloon vertebroplasty in combination with pedicle-screw instrumentation. SUMMARY OF BACKGROUND DATA: The role of intact longitudinal ligaments in traumatic spine fractures, for prevention of bone retropulsion and subsequent reduction, has been discussed in several studies but is still up for debate. In a recent human cadaveric burst fracture study, inflatable bone tamps and calcium phosphate cement were used for the augmentation of the anterior column after pedicle-screw instrumentation. The additional balloon vertebroplasty procedure was found to be feasible and safe, but no data pertaining to unwarranted bone displacement or cement leakage during the procedure are available for burst fractures with damaged longitudinal ligaments. METHODS: Ten thoracic and 10 lumbar burst fractures, with rotation or flexion components, were created, and balloon vertebroplasty with calcium phosphate cement was performed after pedicle-screw instrumentation. Volumetric datasets (using the 3-dimensional (3D) rotational x-ray imaging technique) of the fractures were obtained during the following phases: intact, fractured, after reduction and stabilization with pedicle-screws, after inflation of the balloons, after deflation and removal of the balloons, and after injection of the cement. The amount of ABD and PBD was measured on reconstructed sagittal images and recorded together with the presence of extracorporal cement leakage. The continuity of the longitudinal ligaments was assessed after anatomic dissection. RESULTS: During the balloon vertebroplasty procedure, a significant (P < 0.05) increase of ABD (at both thoracic and lumbar level) and PBD (thoracic level) occurred after inflation of the balloons. After deflation and subsequent injection of the cement, however, the ABD and PBD returned to the preinflation levels. The absolute amount of ABD and PBD (<1 mm) during inflation was considered of little clinical importance. No differences in ABD or PBD were observed for specimens with or without continuity of the corresponding longitudinal ligament, irrespective of the level, at any of the phases during the experiment (P > 0.5 in all cases). A small amount of cement leakage was observed in the psoas compartment of one specimen with intact longitudinal ligaments. CONCLUSIONS: It is suggested that balloon vertebroplasty after pedicle-screw instrumentation may safely be used, in terms of bone displacement and cement leakage, in fracture types where damage to longitudinal ligaments is to be expected.

Aged↗

Redislocation in a halo vest of an atlanto-occipital dislocation in a child: recommendations for treatment.

STUDY DESIGN: A case of a child with a traumatic atlanto-occipital dislocation (AOD), with unexpected secondary redislocation in a halo vest is presented. An extensive literature review was performed. OBJECTIVES: To show the difficulties in diagnosing AOD, to emphasize its intrinsic instability, and to make recommendations for treatment of this condition. SUMMARY OF BACKGROUND DATA: AOD is a rare condition that is difficult to diagnose. Because of the high neurologic morbidity, this injury often results in death. Different diagnostic approaches have been evaluated throughout the literature. Also, there are different opinions on how to treat AOD. There is no large trial that compares conservative treatment with surgery. METHODS: A 4-year-old child with a longitudinal distraction of the head is described. Because of his tenuous medical condition, initial treatment was in a halo vest. However, redislocation occurred inside the vest, so surgical stabilization was performed as soon as the patient's condition allowed. RESULTS: To diagnose AOD, different measuring techniques can be applied on a lateral cervical spine radiograph. Using a combination of these tools, the sensitivity of the examination can be increased. There are 3 types of dislocation. Type II, which is a longitudinal distraction, is most unstable. Once confirmed, AOD should be immobilized. This can temporarily be performed with a halo-frame, but our case shows that this does not necessarily provide a stable situation, and early surgical reconstruction should be considered. CONCLUSIONS: AOD can be an extremely unstable lesion and is difficult to diagnose. Although initial treatment in a halo vest is recommended, redislocation can occur even in the vest. We recommend urgent surgical stabilization.

Atlanto-Occipital Joint↗

Three-dimensional rotational X-ray imaging for spine surgery: a quantitative validation study comparing reconstructed images with corresponding anatomical sections.

STUDY DESIGN: A validation study was done in which reconstructed three-dimensional rotational x-ray images were quantitatively compared with corresponding anatomic sections. OBJECTIVES: To assess the accuracy of reconstructed images acquired on a three-dimensional rotational x-ray imaging device. SUMMARY OF BACKGROUND DATA: Minimally invasive procedures have proven quite successful as alternatives for a multitude of open treatments. An unfavorable property of this type of procedure is a lack of direct vision of the operating area. Three-dimensional rotational x-ray imaging may be able to merge the advantages of computed tomography and fluoroscopy: real-time two-dimensional projections for fast visual feedback and three-dimensional reconstructions for detailed volumetric imaging of complex anatomy. METHODS: Twenty traumatic thoracolumbar burst fractures were created and underwent pedicle screw instrumentation and balloon vertebroplasty. Subsequently, a three-dimensional dataset was obtained, and the midsagittal image was reconstructed. The specimens were sliced, and photographs were obtained. Multiple parameters on the reconstructed images and photographs were measured two times by two observers using a graphical method. The differences and standard deviations were calculated for the corresponding parameters and for the intraobserver data. RESULTS: The mean difference between the corresponding values ranged between -1.1 and 2.1 mm for all parameters. The standard deviation for the differences per parameter ranged between 1.2 and 3.2 mm. The intraobserver differences ranged from -0.8 to 1.4 mm, and the standard deviation varied between 0.4 and 2.4 mm. CONCLUSIONS: The reconstructed midsagittal images were accurate in all measured parameters. The three-dimensional rotational x-ray technique may prove to be valuable for less invasive spine surgery.

Aged↗

Balloon vertebroplasty in combination with pedicle screw instrumentation: a novel technique to treat thoracic and lumbar burst fractures.

STUDY DESIGN: Clinical trial (phase II). OBJECTIVES: To assess the feasibility and safety of balloon vertebroplasty after posterior short-segment reduction and fixation for the treatment of traumatic burst fractures. SUMMARY OF BACKGROUND DATA: Hardware failure and loss of reduction after posterior short-segment instrumentation are complications caused by insufficiency of anterior column support. This is due to migration of disc tissue through the endplate into the fractured vertebral body that cannot be restored with posterior instrumentation. METHODS: Patients with traumatic thoracolumbar burst fractures without neurologic deficits were included. After posterior reduction and fixation, bilateral transpedicular balloon reduction of the endplate was performed, and calcium phosphate cement was injected. Preoperative and postoperative Cobb angle and central and anterior height were assessed with radiographs and MRI. RESULTS: Twenty patients underwent surgery without technical difficulties, and a substantial reduction of the endplates could be achieved with the technique. All patients recovered uneventfully, and the neurologic examination revealed no deficits. The postoperative radiographs and magnetic resonance images demonstrated a good fracture reduction and filling of the bone defect without unwarranted bone displacement. The central and anterior height of the vertebral body could be restored to 78 and 91% of the estimated intact height, respectively. Complications were cement leakage in five cases without clinical implications and one wound hematoma. CONCLUSIONS: Transpedicular balloon vertebroplasty for the direct restoration of burst fractures seems feasible in combination with posterior instrumentation. Cement leakage occurred but had no clinical consequences.

Adolescent↗

Less invasive anterior column reconstruction in thoracolumbar fractures.

Posterior short-segment pedicle screw constructs are commonly used for reduction and fixation of traumatic thoracolumbar spine fractures. Although this technique is usually simple and effective, complications such as loss of fixation or recurrence of deformity are common because of the insufficiency of the damaged anterior column. Anterior approaches to address this deficiency are associated with high morbidity and complications. We have developed a technique to reduce and support the fractured anterior column through a transpedicular approach. Balloon-assisted-endplate-reduction (BAER) followed by vertebroplasty (VTP) with calcium phosphate cement in combination with short segment pedicle screw construct seem to be a safe and effective technique to reconstruct the anterior column in a less invasive manner. In this article, the rationale behind this technique, experimental studies, and the first clinical results are discussed.

Biocompatible Materials↗

The role of 3-D rotational x-ray imaging in spinal trauma.

The most widely used imaging devices in trauma spine surgery are fluoroscopy and computed tomography. Both techniques have their specific pros and cons and expose both patient and operating staff to considerable radiation during image acquisition. Three-dimensional-rotational x-ray imaging (3-DRX) is a relatively new technique in which a C-arm is moved around the patient to allow for an "acquisition run" during which multiple fluoroscopy images are obtained. The images can subsequently be processed on a workstation into a 3-D volume, which can then be manipulated in a similar way to 3-D computed tomography data. The 3-DRX technique combines the advantages of both fluoroscopy and computed tomography: fast visual feedback and high resolution multiplanar reformatted images, respectively, and could be used in an intraoperative setting. In this paper some technical aspects of 3-DRX imaging and its potential role in spinal trauma treatment are presented.

Humans↗

Optimization of bone tissue engineering in goats: a peroperative seeding method using cryopreserved cells and localized bone formation in calcium phosphate scaffolds.

BACKGROUND: Bone tissue engineering by combining cultured bone marrow stromal cells with a porous scaffold is a promising alternative for the autologous bone graft. Drawbacks of the technique include the delay necessary for cell culture and the complicated logistics. We investigated methods to bypass these drawbacks. Furthermore, we investigated the localization of bone formation inside the scaffold. METHODS: Bone marrow stromal cells from seven goats were culture expanded and cryopreserved. One week before surgery, some of the cells were thawed, cultured, and seeded on porous calcium phosphate scaffolds. The constructs were cultured for another week until implantation. The remaining cryopreserved cells were thawed just before implantation and peroperatively resuspended in plasma before combining with the scaffold. Scaffolds impregnated with fresh bone marrow, devitalized cultured constructs, and empty scaffolds served as controls. All samples were implanted in the back muscles of the goats for 9 weeks. RESULTS: Histologic examination showed minimal (<1%) bone in the empty and devitalized scaffolds, 4.2 +/- 5.1 bone area percent in the bone marrow samples, and significantly more bone in both the cultured and peroperatively seeded constructs (11.7 +/- 2.5 and 14.0 +/- 2.0%). The peripheral 350 microm of the implants contained significantly less bone. CONCLUSION: Peroperative preparation of osteogenic constructs with cryopreserved cells is feasible. These constructs yield substantially more bone than the scaffolds alone or scaffolds impregnated with fresh bone marrow. Bone deposition is much less on the scaffold periphery.

Animals↗

Noninvasive magnetic resonance to three-dimensional rotational x-ray registration of vertebral bodies for image-guided spine surgery.

STUDY DESIGN: Magnetic resonance (MR) and three-dimensional rotational x-ray (3DRX) images of cadaveric spinal segments were registered using a conventional point-based technique and a noninvasive technique called maximization of mutual information (MMI). OBJECTIVES: To assess the feasibility and accuracy of MMI-based registration in comparison with point-based registration as a new noninvasive image-to-patient registration technique for use in minimally invasive image-guided spine surgery. SUMMARY OF BACKGROUND DATA: In image-guided orthopedic surgery, correspondence between before surgery acquired images and the patient is required. Currently, this necessitates an invasive registration step, in which anatomic landmarks on the bone surface have to be physically touched by a pointer. To overcome this invasive procedure, we propose using a calibrated 3DRX system, a modality that can visualize high contrast objects intraoperatively and that provides a direct correspondence between the image data and the patient. A noninvasive voxel-based technique is used to register the intraoperative 3DRX image to a before surgery acquired MR image. METHODS: Cadaveric thoracic and lumbar spine segments were implanted with markers, which were used as landmarks. The accuracy of the landmark-based registration was compared with MMI-based registration using the residual errors on the marker positions. RESULTS: The errors made using the point-based registration technique were compared with the errors made with MMI. The results show a statistically significant lower error (P < 0.01) for the proposed MMI method. CONCLUSIONS: Noninvasive MMI registration of intraoperative 3DRX images to preoperative MR images is more accurate than point-based registration in cadaveric spine segments. It is therefore a promising technique for replacing the invasive landmark-based registration that is currently used in image-guided spine surgery.

Humans↗

Temperature elevation after vertebroplasty with polymethyl-methacrylate in the goat spine.

Although the general complication rate for vertebroplasty is low, neural compression and thermal damage have been related to the use of polymethyl-methacrylate cement. This study focuses on the risk of thermal tissue damage after vertebroplasty. In this study, cavities were created by a transpedicular approach in the vertebral bodies (L3, L4, and L5) of four milk goats, and polymethyl-methacrylate cement was injected. In three locations (the bone-cement interface, the epidural space, and the disc space) the temperature was measured in regular intervals after injection of the cement. The mean injected volume was 0.8 ml, which accounted for a 22% volume fraction. The mean peak temperature of the bone-cement interface was 44.6 degrees C, while the maximum temperature at the epidural space and disc space was 37.0 and 37.5 degrees C, respectively. The local temperature measured after in vivo vertebroplasty did not reach values that are known to cause tissue necrosis.

Animals↗

Complications of artificial disc replacement: a report of 27 patients with the SB Charité disc.

Disc prosthesis surgery is rapidly becoming an option in treating patients with symptomatic degenerative disc disease. Only short-term and midterm results are described in the literature. Most operated patients belong to the age group of 30-50 years. In these active patients, complications can be expected to increase with longer follow-up, similar to total joint replacements in the extremities. Reported here is a series of 27 patients from another institution, who presented with unsatisfactory results or complications after SB Charité disc replacement. The objective of this work was to describe the possible short- and long-term unsatisfactory results of disc prosthesis surgery. Twenty-seven patients were seen in a tertiary university referral center with persisting back and leg complaints after having received a Charité disc prosthesis. All patients were operated on in a neighboring hospital. Most patients were operated on at the L4-L5 and /or the L5-S1 vertebral levels. The patients were evaluated with plain radiography, some with flexion-extension x-rays, and most of them with computed tomography scans. The group consisted of 15 women and 12 men. Their mean age was 40 years (range 30-67 years) at the time of operation. The patients presented to us a mean of 53 months (range 11-127 months) following disc replacement surgery. In two patients, an early removal of a prosthesis was required and in two patients a late removal. In 11 patients, a second spinal reconstructive salvage procedure was performed. Mean follow-up for 26 patients with mid- and long-term evaluation was 91 months (range 15-157 months). Early complications were the following: In one patient, an anterior luxation of the prosthesis after 1 week necessitated removal and cage insertion, which failed to unite. In another patient with prostheses at L4-L5 and L5-S1, the prosthesis at L5-S1 dislocated anteriorly after 3 months and was removed after 12 months. Abdominal wall hematoma occurred in four cases. Retrograde ejaculation with loss of libido was seen in one case and erection weakness in another case. A temporary benefit was experienced by 12 patients, while 14 patients reported no benefit at all. Main causes of persistent complaints were degeneration at another level in 14, subsidence of the prosthesis in 16, and facet joint arthrosis in 11. A combination of pathologies was often present. Slow anterior migration was present in two cases, with compression on the iliac vessels in one case. Polyethylene wear was obvious in one patient 12 years after operation. In eight cases, posterior fusion with pedicle screws was required. In two cases, the prosthesis was removed and the segment was circumferentially fused. These procedures resulted in suboptimal long-term results. In this relatively small group of patients operated on with a Charité disc prosthesis, most problems arose from degeneration of other lumbar discs, facet joint arthrosis at the same or other levels, and subsidence of the prosthesis. It is to be expected that many more patients will be seen with late problems some years after this operation as the survivorship will decrease with time.

Adult↗