PubMed Health⌕ Search

Biomedical subjects

F G Baggioni

Publications and source records attributed to F G Baggioni.

3 recordsLinked to original sources

A simple new method for atrial triggered pacemaker. Preliminary clinical trials.

Inadequacy of cardiac output in the ventricular paced complete heart block is generally related to lack of rate adaptation to varying atrial contribution. The status of the art in leads technology had not yet freed the clinicians from the necessity to use a double lead system (atrial and ventricular) in order to maintain the atrial synchronization. Authors present a new simple catheter with double electrode (atrial receptor and ventricular stimulator) which, connecting to a new conceived VAT generator, let to obtain atrial synchronization with a single catheter-electrode implantation. External units were tried in 22 patients who needed temporary treatment. The results appear to be positive in all cases and encourage the Authors to continue with the task undertaken until the system is made implantable.

Clinical Trials as Topic↗

Intracardiac electrogram parameters, electrode surface area and pacer input impedance: their correlations.

Small surface area electrodes are accused of sensing defects which were related to alterations that they induce in the endocardiac electrograms. Since several factors affect the cardiac signal coming from electrode to sensing circuit, i.e. electrode surface area, electrode-tissue interface, pacemaker input impedance and sensing amplifier pass-band, Authors present their studies performed on 252 implanted electrodes of various type. Study was carried out by connecting in parallel to the recorder a variable resistor in order to simulate different pacer input impedances. The results showed a significant reduction in RS amplitude when recorder was paralleled with resistor values lower than 40 K. Slew rates showed a similar behaviours since RS steep tract did not change his duration with load, while total QRS duration is reduced. High speed analysis has shown that the RS segment is not linear in about 40% of cases: the main tract is used for calculations. The most significant attenuations and distortions of endocardial electrogram were observed with smallest electrodes and lowest resistances parallel connected: in these cases the sensing impedance at the electrode-tissue interface appears to be between 3 to 5 K ohms. The results suggest that the most of the alledged sensing faults experienced in the past were probably due to small tip electrodes connected to low input impedance generators or to impending failure situations. The AA. conclude that the main question does not concerne a true electrode inefficiency but a wrongly chosen pacemaker-electrode combination, i.e. small tip electrode connected with old generator models. To avoid the evaluation error, it would be instrumental that the pacemaker manufacturers would specify input characteristics of their generators. So, the implanting clinician becomes able to exactly evaluate the true signal arriving to the sensing circuit by connecting in parallel with the recorder input a resistor whose value approximates the input resistance of the generator to be implanted.

Cardiac Pacing, Artificial↗

Endocardial electrogram amplitudes and stimulation thresholds in monopolar electrodes: possible correlations.

The possible correlations between unipolar endocardial electrogram amplitudes and voltage threshold stimulation were tested with electrodes having metallic tips of two different areas. The threshold measurements were always studied with the same pulse characteristics. The data were analyzed through five mathematical functions (linear, exponential, potential, hyperbola and translate hyperbola). The results show a significant correlation, mainly by hyperbolic and exponential functions. Consequently, we could deduce a parameter value since we already knew the other one. This was possible not only during the first implantation, but also later on.

Adult↗