Three-point Dixon method with a MISSTEC sequence.

We propose an adaptation of the MISSTEC sequence (simultaneous acquisition of a spin echo and several stimulated echoes) for performing fat suppression with a three-point Dixon method. In vivo measurements were performed on healthy volunteers using a sequence of three echoes (one spin echo and two stimulated echoes) within a third of the acquisition time taken by the regular three-point method, with the same spatial resolution and with half the signal-to-noise ratio (SNR). Homogeneity of fat suppression was greatly increased in comparison with the two-point method.

Algorithms

Diffusion imaging with a multi-echo MISSTEC sequence.

An imaging method is presented to measure the water-diffusion coefficient. The sequence (MISSTEC) uses the simultaneous acquisition of a spin echo and several stimulated echoes with the same intensity except for diffusion weighting. The optimal number of stimulated echoes was calculated to minimize the diffusion coefficient error (D). D values obtained in vitro and in vivo were in good agreement with those from the spin-echo sequence (IntraVoxel Incoherent Motion [IVIM] method). The total acquisition time is half that of the classic IVIM method.