[Tools for metrological support in measurements of the characteristics of devices for radionuclide diagnosis].

The application of the sets of GSO IGIM-OT, GSO IGIM allows one to introduce the unified methods for metrological surveillance in the development, production and use medical radiodiagnostic measuring instruments. Using the GSO IGIM, GSO IGIM-OT, as state reference samples of radionuclide activity in the solid radiometric source of ionizing radiation creates the common basis for metrological control of the measurement qualities of the characteristics of medical equipment used in radionuclide diagnosis.

Humans↗

[A solid-state reference sample of iodine radionuclide activity for checking radiometers and a method for its production].

This development relates to nuclear engineering, concerns the sources of ionizing radiation of iodine radionuclides and the ways of their manufacture. The iodine radionuclides are used for calibration of radiometers that measure the specific activity of iodine radionuclide in biological specimens, in particular for their use to identify noncancer diseases. The purpose of the paper is to design a routine activity sample for iodine radionuclide in the solid-state source for transfer of an iodine radionuclide activity unit from the reference iodine radionuclide activity unit to an operating meter with a simultaneous decrease in testing errors and an increase, in terms of iodine radionuclide half-life, in the life of the reference sample by stabilizing the activity unit of iodine radionuclide (in terms of its half-life) by eliminating the process of sublimation of iodine radionuclide from the working surface of the source. The objective is achieved by the fact that the radionuclide is replaced on the lower internal build-up part of the container, fixed with the upper internal build-up part, then kept for at least 10-12 hours at least at 18-20 degrees C, packed into a hollow container and sealed with epoxide resin. According to the proposed technical design, reference activity samples of iodine radionuclide in the solid-state radiometric source with a preset activity value have been prepared. The testing error for iodine radionuclide activity is minimal and equal to 1.0-3.0% at a 0.95 confidence.

Calibration↗

[Government reference samples of the activity of long-life radionuclides used as medical emission test objects for radionuclide diagnostic apparatus].

The paper describes the procedure for producing the solid radiometric sources with their large plane linear capillary and volumetric working surface having a high uniformity of the external ionizing irradiation of a radionuclide and gives the basic parameters of the state reference samples which have been made on the basis of the developed technology for the activity of long-living radionuclides used as medical emission test objects to test radionuclide diagnostic apparatuses.