A comparison of subdermal wire electrodes with collodion-applied disk electrodes in long-term EEG recordings in ICU.
OBJECTIVE: To compare long-term electroencephalographic (EEG) recordings of standard collodion-applied scalp disk electrodes (SDEs) with newly developed subdermal wire electrodes (SWEs) in comatose intensive care unit (ICU) patients. METHODS: Ten comatose ICU patients had simultaneous recordings from 8 active SDEs and 8 active SWE for >24 h. The timing and number of 60 Hz and other electrode artifacts were compared for each set of electrodes by an EEGer who read the recordings in a blinded manner. RESULTS: Sixty Hertz artifact was seen in 16 of 80 SDE and one of 80 SWEs within the first 6 h (P=0.0002). Large, persistent artifacts occurred in 30/80 SDE and 8 of 80 SWE (P=0.0001). Motion artifact with chest physiotherapy was more common in SWEs. CONCLUSIONS: SWE are less susceptible to artifacts and are more suitable for the long-term EEG monitoring in ICU. SIGNIFICANCE: This is the first controlled study that demonstrates the superiority of SWEs compared to SDEs in an ICU population.