Comparison of the effects of bilateral posterior dynamic and rigid fixation devices on the loads in the lumbar spine: a finite element analysis.

A bilateral dynamic stabilization device is assumed to alter favorable the movement and load transmission of a spinal segment without the intention of fusion of that segment. Little is known about the effect of a posterior dynamic fixation device on the mechanical behavior of the lumbar spine. Muscle forces were disregarded in the few biomechanical studies published. The aim of this study was to determine how the spinal loads are affected by a bilateral posterior dynamic implant compared to a rigid fixator which does not claim to maintain mobility. A paired monosegmental posterior dynamic implant was inserted at level L3/L4 in a validated finite element model of the lumbar spine. Both a healthy and a slightly degenerated disc were assumed at implant level. Distraction of the bridged segment was also simulated. For comparison, a monosegmental rigid fixation device as well as the effect of implant stiffness on intersegmental rotation were studied. The model was loaded with the upper body weight and muscle forces to simulate the four loading cases standing, 30 degrees flexion, 20 degrees extension, and 10 degrees axial rotation. Intersegmental rotations, intradiscal pressure and facet joint forces were calculated at implant level and at the adjacent level above the implant. Implant forces were also determined. Compared to an intact spine, a dynamic implant reduces intersegmental rotation at implant level, decreases intradiscal pressure in a healthy disc for extension and standing, and decreases facet joint forces at implant level. With a rigid implant, these effects are more pronounced. With a slightly degenerated disc intersegmental rotation at implant level is mildly increased for extension and axial rotation and intradiscal pressure is strongly reduced for extension. After distraction, intradiscal pressure values are markedly reduced only for the rigid implant. At the adjacent level L2/L3, a posterior implant has only a minor effect on intradiscal pressure. However, it increases facet joint forces at this level for axial rotation and extension. Posterior implants are mostly loaded in compression. Forces in the implant are generally higher in a rigid fixator than in a dynamic implant. Distraction strongly increases both axial and shear forces in the implant. A stiffness of the implant greater than 1,000 N/mm has only a minor effect on intersegmental rotation. The mechanical effects of a dynamic implant are similar to those of a rigid fixation device, except after distraction, when intradiscal pressure is considerably lower for rigid than for dynamic implants. Thus, the results of this study demonstrate that a dynamic implant does not necessarily reduce axial spinal loads compared to an un-instrumented spine.

Biomechanical Phenomena↗

An instrumented implant for vertebral body replacement that measures loads in the anterior spinal column.

Realistic loads on a spinal implant are required among others for optimization of implant design and preclinical testing. In addition, such data may help to choose the optimal physiotherapy program for patients with such an implant and to evaluate the efficacy of aids like braces or crutches. Presently, no implant is available that can measure loads in the anterior spinal column during activities of daily life. Therefore, an implant instrumented for in vivo load measurement was developed for vertebral body replacement. The aim of this paper is to describe in detail a telemeterized implant that measures forces and moments acting on it. Six load sensors, a nine-channel telemetry unit and a coil for inductive power supply of the electronic circuits were integrated into a modified vertebral body replacement (Synex). The instrumented part of the implant is hermetically sealed. Patients are videotaped during measurements, and implant loads are displayed on and off line. The average accuracy of load measurement is better than 2% for force and 5% for moment components with reference to the maximum value of 3000 N and 20 Nm, respectively. The measuring implant described here will provide additional information on spinal loads.

Equipment Design↗

Effect of a posterior dynamic implant adjacent to a rigid spinal fixator.

BACKGROUND: A slightly degenerated disc adjacent to a segment that has to be fused is sometimes instrumented with a dynamic fixator. The dynamic implant is assumed to reduce disc loads at that level and to preserve disc function, thus inhibiting the progression of degeneration. METHODS: A three-dimensional finite element model of the lumbar spine was used to study the effect of a dynamic implant on the mechanical behavior at the corresponding level. After studying a healthy lumbar spine for comparison, a rigid fixator and a bone graft were inserted at L2/L3. Healthy and degenerated discs were assumed at the adjacent level, i.e. L3/L4. An additional paired dynamic posterior fixator was then implemented at level L3/L4. Finally, the segment with the dynamic fixator was distracted to the height of a healthy disc. The loading cases of walking, extension, flexion and axial rotation were simulated. FINDINGS: A dynamic implant reduces intersegmental rotation for walking, extension and flexion as well as facet joint forces for axial rotation at its insertion level. Intradiscal pressure is not markedly reduced by a dynamic implant. Moreover, there are no substantial differences between the mechanical behavior of rigid and dynamic fixators. INTERPRETATION: Our model does not predict major differences in the mechanical effects between rigid and dynamic fixators despite the extreme assumption that a dynamic implant does not transfer moments. The results do not support the assumption that disc loads are significantly reduced by a dynamic implant. For axial rotation, however, dynamic fixation devices do reduce the force in the facet joint.

Back↗

Effects of fusion-bone stiffness on the mechanical behavior of the lumbar spine after vertebral body replacement.

BACKGROUND: Implants for vertebral body replacement are often inserted together with an additional stabilizing implant, e.g. an internal fixation device. During implantation bone grafts or milled bone is normally added to the anterior implant. Little is known about the stiffening effect of this fusion-bone mass on the mechanical behavior of the corresponding bone region, including the load distribution between the different parts. METHODS: A three-dimensional finite element model of the lumbar spine was created with a vertebral body replacement at L3, a paired internal fixation device between L2 and L4, and left anterolateral fusion bone. The elastic modulus of fusion bone was varied in discrete steps between 0 MPa and 10,000 MPa. The model was loaded to simulate standing, 20 degrees flexion, 15 degrees extension and 6 degrees axial rotation in the lumbar spine. FINDINGS: The elastic modulus of fusion bone has a considerable effect on the compressive force on vertebral body replacement and fusion bone for all loading cases studied. For extension, it also affects intersegmental rotation, the force in the erector spinae muscle, the compressive force on the internal fixator and intradiscal pressure in the adjacent discs. The elastic modulus most strongly affects the different parameters at values between 0 MPa and 500 MPa. INTERPRETATION: Adding bone mass during vertebral body replacement reduces the loads on the ventral implant for all loading cases studied but extension when the fusion-bone stiffens. This protects the implant from fatigue. The load on the fusion bone increases with increasing elastic modulus. Thus bone grafts should be used whenever possible.

Biomechanical Phenomena↗

Spinal loads after osteoporotic vertebral fractures treated by vertebroplasty or kyphoplasty.

Vertebroplasty and kyphoplasty are routine treatments for compression fractures of vertebral bodies. A wedge-shaped compression fracture shifts the centre of gravity of the upper body anteriorly and generally, this shift can be compensated in the spine and in the hips. However, it is still unclear how a wedge-shaped compression fracture of a vertebra increases forces in the trunk muscle and the intradiscal pressure in the adjacent discs. A nonlinear finite element model of the lumbar spine was used to estimate the force in the trunk muscle, the intradiscal pressure and the stresses in the endplates in the intact spine, and after vertebroplasty and kyphoplasty treatment. In this study, kyphoplasty represents a treatment with nearly full fracture reduction and vertebroplasty one without restoration of kyphotic angle although in reality kyphoplasty does not guarantee fracture reduction. If no compensation of upper body shift is assumed, the force in the erector spine increases by about 200% for the vertebroplasty but by only 55% for the kyphoplasty compared to the intact spine. Intradiscal pressure increases by about 60 and 20% for the vertebroplasty and kyphoplasty, respectively. In contrast, with shift compensation of the upper body, the increase in muscle force is much lower and increase in intradiscal pressure is only about 20 and 7.5% for the vertebroplasty and kyphoplasty, respectively. Augmentation of the vertebral body with bone cement has a much smaller effect on intradiscal pressure. The increase in that case is only about 2.4% for the intact as well as for the fractured vertebra. Moreover, the effect of upper body shift after a wedge-shaped vertebral body fracture on intradiscal pressure and thus on spinal load is much more pronounced than that of stiffness increase due to cement infiltration. Maximum von Mises stress in the endplates of all lumbar vertebrae is also higher after kyphoplasty and vertebroplasty. Cement augmentation has only a minor effect on endplate stresses in the unfractured vertebrae. The advantages of kyphoplasty found in this study will be apparent only if nearly full fracture reduction is achieved. Otherwise, differences between kyphoplasty and vertebroplasty become small or vanish. Our results suggest that vertebral body fractures in the adjacent vertebrae after vertebroplasty or kyphoplasty are not induced by the elevated stiffness of the treated vertebra, but instead the anterior shift of the upper body is the dominating factor.

Arthroplasty↗

Analysis of the influence of disc degeneration on the mechanical behaviour of a lumbar motion segment using the finite element method.

Compared to a healthy intervertebral disc, the geometry and the material properties of the involved tissues are altered in a degenerated disc. It is not completely understood how this affects the mechanical behaviour of a motion segment. In order to study the influence of disc degeneration on motion segment mechanics a three-dimensional, nonlinear finite element model of the L3/L4 functional unit was used. Different grades of disc degeneration were simulated by varying disc height and bulk modulus of the nucleus pulposus. The model was loaded with pure moments of 10Nm in the three main anatomic planes. The finite element model predicted the same trends for intersegmental rotation and intradiscal pressure as described in the literature for in vitro studies. A comparison between calculated intersegmental rotation and experimental data showed a mean difference of 1.9 degrees while the mean standard deviation was 2.5 degrees . A mildly degenerated disc increases intersegmental rotation for all loading cases. With further increasing disc degeneration intersegmental rotation is decreased. For axial rotation the decrease takes place in the final stage. Intradiscal pressure is lower while facet joint force and maximum von Mises stress in the annulus are higher in a degenerated compared to a healthy disc.

Biomechanical Phenomena↗

Effect of different surgical strategies on screw forces after correction of scoliosis with a VDS implant.

Pullout of the cranial end-vertebra screw following the correction of a scoliosis with the VDS implant is a common complication. Very little is known about the forces acting on the screws during ventral derotation spondylodesis (VDS) in ventral scoliosis surgery. These forces determine the risk of screw-loosening. The purpose of this study was to identify implant properties and to determine surgical correction strategies that reduce the risk of cranial end-vertebra screw pullout. For this aim, a three-dimensional nonlinear finite element model of a scoliotic thoracic spine was created with a Cobb angle of 61 degrees and 32 degrees rotation. The VDS implant was inserted between T5 and T9. The longitudinal rod diameter, the implant material and seven surgical correction strategies were examined to determine their influence on the Cobb angle as well as on derotation and on axial and transverse forces in the screws. A stiffer implant achieves a better correction but causes higher axial and transverse screw forces. Axial tensile forces act on the screws fixed to the cranial end vertebra and the middle vertebra, while axial compressive forces act on the other screws. A strong correction at the cranial segment leads to high axial and transverse screw forces in the farthest cranial screw and thus to a high risk of screw pullout. The resultant transverse force is often much higher than the axial force component. Simulation of local trunk muscle forces has only a minor effect on the results. The axial tensile forces and thus the risk of screw pullout are highest at the cranial end vertebra. A strategy in which surgical correction is strong in the middle segments and moderate in the outer ones leads to a good reduction of the Cobb angle, a wide derotation angle, and relatively low axial tensile forces at the cranial end vertebra screw.

Bone Screws↗

Determination of trunk muscle forces for flexion and extension by using a validated finite element model of the lumbar spine and measured in vivo data.

Muscle forces stabilize the spine and have a great influence on spinal loads. But little is known about their magnitude. In a former in vitro experiment, a good agreement with intradiscal pressure and fixator loads measured in vivo could be achieved for standing and extension of the lumbar spine. However, for flexion the agreement between in vitro and in vivo measurements was insufficient. In order to improve the determination of trunk muscle forces, a three-dimensional nonlinear finite element model of the lumbar spine with an internal fixation device was created and the same loads were applied as in a previous in vitro experiment. An extensive adaptation process of the model was performed for flexion and extension angles up to 20 degrees and -15 degrees, respectively. With this validated computer model intra-abdominal pressure, preload in the fixators, and a combination of hip- and lumbar flexion angle were varied until a good agreement between analytical and in vivo results was reached for both, intradiscal pressure and bending moments in the fixators. Finally, the fixators were removed and the muscle forces for the intact lumbar spine calculated. A good agreement with the in vivo results could only be achieved at a combination of hip- and lumbar flexion. For the intact spine, forces of 170, 100 and 600 N are predicted in the m. erector spinae for standing, 5 degrees extension and 30 degrees flexion, respectively. The force in the m. rectus abdominus for these body positions is less than 25 N. For more than 10 degrees extension the m. erector spinae is unloaded. The finite element method together with in vivo data allows the estimation of trunk muscle forces for different upper body positions in the sagittal plane. In our patients, flexion of the upper body was most likely a combination of hip- and lumbar spine bending.

Adolescent↗

Effect of total disc replacement with ProDisc on intersegmental rotation of the lumbar spine.

STUDY DESIGN: The mechanical behavior of the lumbar spine after insertion of a ProDisc prosthesis was studied using 3-dimensional nonlinear finite element models. OBJECTIVE: To determine how the mechanical behavior of the lumbar spine is affected by the implant position and height, as well as by removing different portions of the natural disc and resuturing the anterior longitudinal ligament (ALL). SUMMARY OF BACKGROUND DATA: Little is known about how the affected and adjacent levels of the spine are influenced by the implant height and position or by the surgical procedure. METHODS: The artificial disc ProDisc was integrated in a validated 3-dimensional, nonlinear, finite element model of the lumbar spine. The model was loaded with the upper body weight and muscle forces to simulate standing, 30 degrees flexion, 15 degrees extension, and 6 degrees axial rotation. The disc position was varied by up to 2 mm in both an anterior and posterior direction. Three different disc heights were investigated, as well as the influence of removing different portions of the natural disc and resuturing the ALL. RESULTS: Implant position strongly influences intersegmental rotation for the loading cases of standing and flexion. A disc height 2 mm in excess of the normal disc space increases intersegmental rotation at implant level during standing and extension. The values for intersegmental rotation are closer to those for the intact spine when lateral portions of the anulus are not removed. Resuturing the ALL has a strong effect on the loading cases of extension and standing. CONCLUSIONS: When implanting an artificial disc, great care should be taken in choosing the optimal height and correct position for the implant. Lateral portions of the anulus should be preserved whenever possible. A perfect reconstruction of the ALL would help to restore the biomechanics to normal.

Arthroplasty, Replacement↗

Influence of ligament stiffness on the mechanical behavior of a functional spinal unit.

Data on the stiffnesses of spinal ligaments are required for analytical studies on the mechanical behavior of spinal segments. Values obtained experimentally vary widely in the literature. A finite element model of an L3/L4 functional spinal unit was used to determine the influence of ligament stiffness on intersegmental rotation and forces in the ligaments. The lowest values for ligament stiffness selected from the literature were used in one set of calculations, and the highest values were simulated in a second set. The nonlinear model was loaded with pure moments of 7.5 and 15 Nm in the three main anatomical planes. The mechanical behavior of the functional spinal unit was strongly influenced by ligament stiffness. In some cases, a ligament with low stiffness does not carry any load, while the same ligament with high stiffness has to carry a high load. This indicates that finite element models of spinal segments have to be validated and that a realistic quantitative prediction of ligament forces is extremely difficult.

Biomechanical Phenomena↗

Tibio-femoral loading during human gait and stair climbing.

Surgical intervention of the knee joint routinely endeavors to recreate a physiologically normal joint loading environment. The loading conditions resulting from osteotomies, fracture treatment, ligament replacements, and arthroplasties of the knee are considered to have an impact on the long term clinical outcome; however, knowledge regarding in vivo loading conditions is limited. Using a previously validated musculoskeletal lower limb model, we predicted the tibio-femoral joint contact forces that occur in the human knee during the common daily activities of walking and stair climbing. The average resultant peak force during walking was 3.1 times body weight (BW) across four total hip arthroplasty patients. Inter-individual variations proved larger than the variation of forces for each patient repeating the same task. Forces through the knee were considerably larger during stair climbing than during walking: the average resultant peak force during stair climbing was 5.4 BW although peaks of up to 6.2 BW were calculated for one particular patient. Average anteroposterior peak shear components of 0.6 BW were determined during walking and 1.3 BW during stair climbing. These results confirm both the joint contact forces reported in the literature and the importance of muscular activity in creating high forces across the joint. The magnitudes of these forces, specifically in shear, have implications for all forms of surgical intervention in the knee. The data demonstrate that high contact and shear forces are generated during weight bearing combined with knee flexion angles greater than approximately 15 degrees. Clinically, the conditions that produce these larger contact forces should be avoided during post-operative rehabilitation.

Biomechanical Phenomena↗

Effect of bone graft characteristics on the mechanical behavior of the lumbar spine.

There is little information about the influence of bone graft size, position and elasticity on the mechanical behavior of the lumbar spine. Intersegmental motion, intradiscal pressure and stresses in the lumbar spine were calculated using a three-dimensional, nonlinear finite element model which included an internal spinal fixation device and a bone graft. Cross-sectional area, position, and elastic modulus of the graft were varied in this study. Bone grafts, especially very stiff ones, increase stresses on adjacent endplates. Though larger grafts lead to less contact pressure, it is difficult to judge the quality of different bone graft positions. In general, ventral flexion results in lower maximum contact pressure than lateral bending. There is always little intersegmental rotation in the bridged region compared with that of an intact spine.A larger graft with low stiffness should be favored from a mechanical point of view. Patients should avoid lateral bending of the upper body shortly after surgery.

Biomechanical Phenomena↗

Comparison of the mechanical behavior of the lumbar spine following mono- and bisegmental stabilization.

OBJECTIVE: To determine whether the mechanical behavior of the entire lumbar spine differs following mono- and bisegmental stabilization. DESIGN: The mechanical behavior of the lumbar spine was studied using the finite element method. BACKGROUND: Nonunion is somewhat more frequent after bi- than after monosegmental stabilization of the spine. Little is known about differences between the mechanical behavior associated with these procedures. METHODS: A three-dimensional nonlinear finite element model of the lumbar spine with internal spinal fixators and bone grafts was used to study mechanical behavior after mono- and bisegmental fixation with and without stabilization of the bridged vertebra. Finite element analyses were performed to determine the influence of four different graft positions, five loading conditions, and six different pretensions in the longitudinal fixator rod. The following parameters were considered: the maximum contact pressure at the interface between the bone graft and vertebral body, the force transmitted by the bone graft, and the size of the contact area between the graft and the vertebral body. RESULTS: Our model shows no clear differences between mono- and bisegmental fixation. Additional stabilization of the bridged vertebra exerts a partly adverse influence on the parameters studied. Pretension in the bridged region has a strong effect on the mechanical behavior. CONCLUSIONS: The mechanical behavior of the lumbar spine after mono- and bisegmental stabilization is similar. Thus biological factors and the surgical procedure are probably decisive in determining the fusion rate. RELEVANCE: Knowledge of the mechanical behavior after stabilization of the spine may help to improve the fusion rate. Our results suggest that the mechanical factors studied have only a minor influence on fusion rate and that other factors, such as incomplete resection of cartilage plate and poor local blood supply, are more decisive.

Biomechanical Phenomena↗

Loads on an internal spinal fixation device during physical therapy.

BACKGROUND AND PURPOSE: Modified internal spinal fixation devices allow the measurement of the forces and moments acting on the implants. The aim of this study was to measure the loads on internal fixation devices for selected body positions and movements during physical therapy. SUBJECTS AND METHODS: Loads on an internal spinal fixation device were measured in 10 patients with degenerative instability or compression fractures using a telemeterized implant. RESULTS: Relatively low implant loads were found in the recumbent body positions. Most exercises performed in a lying position caused implant loads less than that measured for standing and are therefore not likely to increase the risk of screw breakage. Fixation device loads were lower for sitting relaxed than for standing. The highest implant loads (128% of the value for standing) were measured during walking. Standing up, sitting down, and lateral bending and axial rotation of the upper body while standing led to fixation device loads between 111% and 120% related to the value for standing. Even higher fixation device loads were measured for ventral flexion and extension of the upper body while standing. Kneeling on hands and knees, and flexing and extending the back in this position, caused implant loads that were lower than for standing. DISCUSSION AND CONCLUSION: Standing up, sitting down, and lateral bending and axial rotation of the upper body while standing may slightly increase the risk of pedicle screw breakage, whereas ventral flexion and extension of the upper body while standing may increase this risk considerably if the region bridged by the implant is distracted (the distance between upper and lower screws was increased) during surgery. However, walking is the exercise that plays the major role concerning pedicle screw breakage because it causes the highest bending moments of all exercises studied and it loads the fixation devices most frequently.

Adult↗