PubMed Health⌕ Search

Biomedical subjects

Gero Strauss

Publications and source records attributed to Gero Strauss.

6 recordsLinked to original sources

Evaluation of a navigation system for ENT with surgical efficiency criteria.

The aim of this study is the evaluation of a navigation system (NaviBase) for ENT surgery. For this purpose, a new methodology for the evaluation of surgical and ergonomic system properties has been developed. The practicability of the evaluation instruments will be examined using the example of the overall assessment of the system in comparison with the current surgical standard and with other systems using clinical efficiency criteria. The evaluation is based on 102 ENT surgical applications; of these, 89 were functional endoscopic sinus surgeries (FESS). The evaluation of surgical and ergonomic performance factors was performed by seven ENT surgeons. To evaluate surgical system properties, the Level of Quality (LOQ) in 89 cases of the FESS was determined. It compares the existing information of the surgeon with that of the navigation system on a scale of 0 to 100 and with a mean value of 50 and places it in a relationship to the clinical impact. The intraoperative change of the planned surgical strategy (Change of Surgical Strategy) was documented. The ergonomic factors of the system with the categories of Overall Confidence (Trust), awareness of the situation (Situation Awareness), influence on the operating team, requirements for specific skills (Skill Set Requirement), and cognitive load (Workload Shift) were recorded for all surgical procedures as Level of Reliance (LOR). In the evaluation of the surgical system properties, an average evaluation of the quality of the information, as an LOQ of 63.59, resulted. Every second application of the navigation system (47.9%), on average, led to a change in the surgical strategy. An extension/enhancement of the indication of the endonasal approach through the use of the navigation system was shown in 7 of 102 (6.8%) cases. The completion of the resection in the FESS was rated by 74% of group I and 11% of group II as better in comparison with the standard approach. Total confidence shows a positive evaluation of 3.35 in the LOR. To supplement the evaluation of the navigation system, the technical parameters were included. The maximum deviation, Amax, of the displayed position of the reference value amounted to 1.93 mm. The average deviation was at 1.29 mm with an SD above all values, sd, of 0.29. The subsequent economic evaluation resulted in an effective average extra expenditure of time of 1.35 minutes per case. The overall evaluation of the system imparts application-relevant information beyond the technical details and permits comparability between different assistance systems.

Endoscopy↗

FESS control: realization and evaluation of navigated control for functional endoscopic sinus surgery.

In this paper, a new system for navigated control in functional endoscopic sinus surgery (FESS) is presented. The system allows the safe and convenient use of a shaver that can be enabled by the surgeon only within a specified working space. Preoperatively, the surgeon defines this working space in the axial slices of the CT scan. During the surgery, the positions of the shaver and patient are tracked by an optical navigation system, which calculates whether the shaver is within the working space. The navigated control electronics receives a signal from the navigation system and disables the shaver if it is outside the working space. If the shaver is inside the working space, the surgeon can set its speed freely with a foot pedal. Experimental evaluation shows that the system allows convenient and intuitive safe removal of inflamed tissue while protecting sensitive structures. The clinical applicability was proven in a clinical trial with 10 patients.

Adult↗

Surgical PACS for the digital operating room. Systems engineering and specification of user requirements.

For better integration of surgical assist systems into the operating room, a common communication and processing plattform that is based on the users needs is needed. The development of such a system, a Surgical Picture Aquisition and Communication System (S-PACS), according the systems engineering cycle is oulined in this paper. The first two steps (concept and specification) for the engineering of the S-PACS are discussed.A method for the systematic integration of the users needs', the Quality Function Deployment (QFD), is presented. The properties of QFD for the underlying problem and first results are discussed. Finally, this leads to a first definition of an S-PACS system.

Computer Simulation↗

NEUROGATE: a new MR-compatible device for realizing minimally invasive treatment of intracerebral tumors.

The authors report on the handling and the practicability of a newly developed MR-compatible device, the NEUROGATE (Daum GmbH, Germany), which allows precise planning, simulation and control of stereotactic biopsy in patients with suspect intracranial lesions, and which allows minimally invasive maneuvers to be performed in a comfortable way. Twenty-eight patients were examined stereotactically in the Signa SP interventional 0.5 Tesla MRI (General Electric Medical Systems, USA), including 15 patients with malignant intracerebral tumors and poor general medical conditions (8 gliomas, 7 metastases) who were treated by laser-induced interstitial thermotherapy (LITT) after definite intraoperative neuropathological diagnosis. As a special stereotactic holding device, the NEUROGATE was favored as a reliable tool for stereotaxy and minimally invasive procedures.

Adult↗

Investigation of time-dependency of intracranial brain shift and its relation to the extent of tumor removal using intra-operative MRI.

The object of the paper is to investigate intra-operative brainshift and its relation to the extent of tumor removal. Repeated T1w 3D datasets were acquired at different time points intra-operatively (T0; T1; T2...Tx) using a vertical open 0.5T MR scanner in six patients with intracranial tumor. An offline analysis with initial linear registration, intensity adjustment and finally nonlinear registration of the first versus subsequent time points (T0/T1; T0/T2...To/Tx) was performed, yielding a 3D displacement vector field that describes the brainshift. Brainshift was analysed qualitatively and quantitatively. A semi-automatic segmentation technique was used for calculation of the tumor size and the size of tumor remnants. Semi-automatic segmentation was reliable in all but two cases. Segmentation was difficult and unreliable in astrocytomas grade II. The shift basically followed gravity. The major shift reached levels up to 25 mm. Significant shift was observed at the first time point (T0). Intra-operative brainshift can be analysed qualitatively and also captured quantitatively. Neuronavigation that is based on pre-operatively acquired datasets is associated with a significant risk of surgical morbidity at a very early time point. Parallelisation on a workstation cluster may reduce computation time so that information about the displacement can facilitate updated navigation.

Adult↗

Iterative neuronavigation using 3D ultrasound. A feasibility study.

Intra-operative ultrasound (iUS) can generate 2D images in real-time as well as near real-time 3D datasets of the current situation during an intervention. Tracked ultrasound can locate the images in 3D space and relate them to patient, devices, andpre-operative planning data. Therefore, tracked US is an efficient means for controlling the validity of pre-operative planning, recognition of changes (brain shift) during the intervention, replanning of the operational path due to situational changes (iterative navigation), and finally, controlling the results (residual tumor). This paper describes a neuronavigation system exploiting this potential of interventional tracked US for permanent control of intervention progress and iterative adaptation of the planned procedure to the current situation.

Brain Mapping↗