PubMed Health⌕ Search

Biomedical subjects

Greg W Burgreen

Publications and source records attributed to Greg W Burgreen.

3 recordsLinked to original sources

Computational fluid dynamics analysis of a maglev centrifugal left ventricular assist device.

The fluid dynamics of the Thoratec HeartMate III (Thoratec Corp., Pleasanton, CA, U.S.A.) left ventricular assist device are analyzed over a range of physiological operating conditions. The HeartMate III is a centrifugal flow pump with a magnetically suspended rotor. The complete pump was analyzed using computational fluid dynamics (CFD) analysis and experimental particle imaging flow visualization (PIFV). A comparison of CFD predictions to experimental imaging shows good agreement. Both CFD and experimental PIFV confirmed well-behaved flow fields in the main components of the HeartMate III pump: inlet, volute, and outlet. The HeartMate III is shown to exhibit clean flow features and good surface washing across its entire operating range.

Assisted Circulation↗

Predicting membrane oxygenator pressure drop using computational fluid dynamics.

Three-dimensional computational fluid dynamic (CFD) simulations of membrane oxygenators should allow prediction of spatially dependent variables and subsequent shape optimization. Fiber bed complexity and current computational limitations require the use of approximate models to predict fiber drag effects in complete device simulations. A membrane oxygenator was modified to allow pressure measurement along the fiber bundle in all cardinal axes. Experimental pressure drop information with water perfusion was used to calculate the permeability of the fiber bundle. A three-dimensional CFD model of a commercial membrane oxygenator was developed to predict pressure drops throughout the device. Darcy's Law was used to account for the viscous drag of the fibers and was incorporated as a momentum loss term in the conservation equations. Close agreement was shown between experimental and simulated pressure drops at lower flow rates, but the simulated pressure drops were lower than experimental results at higher flows. Alternate models of fiber drag effects and flow field visualization are suggested as means to potentially improve the accuracy of the flow simulation. Computational techniques coupled with experimental verification offer insight into model validity and show promise for the development of accurate three-dimensional simulations of membrane oxygenators.

Artificial Organs↗

Effects of turbulent stresses upon mechanical hemolysis: experimental and computational analysis.

Experimental and computational studies were performed to elucidate the role of turbulent stresses in mechanical blood damage (hemolysis). A suspension of bovine red blood cells (RBC) was driven through a closed circulating loop by a centrifugal pump. A small capillary tube (inner diameter 1 mm and length 70 mm) was incorporated into the circulating loop via tapered connectors. The suspension of RBCs was diluted with saline to achieve an asymptotic apparent viscosity of 2.0 +/- 0.1 cP at 23 degrees C to produce turbulent flow at nominal flow rate and pressure. To study laminar flow at the identical wall shear stresses in the same capillary tube, the apparent viscosity of the RBC suspension was increased to 6.3 +/- 0.1 cP (at 23 degrees C) by addition of Dextran-40. Using various combinations of driving pressure and Dextran mediated adjustments in dynamic viscosity Reynolds numbers ranging from 300-5,000 were generated, and rates of hemolysis were measured. Pilot studies were performed to verify that the suspension media did not affect mechanical fragility of the RBCs. The results of these bench studies demonstrated that, at the same wall shear stress in a capillary tube, the level of hemolysis was significantly greater (p < 0.05) for turbulent flow as compared with laminar flow. This confirmed that turbulent stresses contribute strongly to blood mechanical trauma. Numerical predictions of hemolysis obtained by computational fluid dynamic modeling were in good agreement with these experimental data.

Animals↗