Reliability of an automatic sensing algorithm.

Automatic adaptation of the atrial sensitivity was evaluated in 18 patients with dual chamber pacemakers (Intermedics, Inc., Relay) in the unipolar mode. After atrial sensitivity was stabilized in the upright position, patients underwent a 1.0 W/kg body weight exercise for 5 minutes. A 24-hour Holter ECG was recorded, and the maximum and minimum atrial sensitivity values reached were stored in the memory of the pulse generator. In a second series of 12 patients, Holter ECGs were recorded twice, starting with the same sensitivity but with automatic adaptation alternately switched "on" or "off." Results of the exercise test: mean atrial sensitivity declined from 2.30 +/- 0.77 mV to 2.03 +/- 0.68 mV. There was no change in five patients, a slight increase in two patients, and lowering of the atrial sensitivity was observed in 11 patients, the difference ranging from 0.2 to 1.0 mV. A total of two P waves in two patients were missed by the atrial amplifier. The minimum and maximum sensitivity reached during Holter monitoring averaged 2.31 +/- 0.67 mV versus 1.72 +/- 0.71 mV (difference 0-1.7 mV). Normal pacemaker function was found in six patients, including one patient without any intrinsic atrial activity. Malsensing of less than five P waves occurred in four patients. More than 50 sensing defects resulted from ectopic atrial beats (four patients). We observed atrial oversensing in three cases; one patient showed atrial over- and undersensing. The comparison between fixed and variable sensitivity did not reveal any superiority of automatic adaptation.(ABSTRACT TRUNCATED AT 250 WORDS)

Aged

Threshold and polarization properties of modern active fixation atrial leads.

The purpose of the study was to compare the stimulation characteristics of two modern active fixation leads (Ela 583F, vitreous carbon tip [ELA] and Intermedics 82-0008-1601, iridium oxide tip [IROX]) with a standard lead (Osypka KY 67 VC, carbon-covered elgiloy tip [OSY]). In three groups of ten patients each, minimum charge threshold delta Qmin and polarization properties were determined via charge telemetry of the pacemaker (Intermedics Cosmos II and Relay) 0, 2, 5, 10, 28, 90, and 180 days after implant (dai). The polarization parameters global capacitance Cg, global resistance Rg, polarization voltage U(p), and a time constant t* (t* = Cg.Rg) were obtained by nonlinear regression. U(p) was always significantly (sig) lower in ELA and IROX (0.04-0.10 V) compared to OSY (0.54-0.76 V). Rg was sig lower in ELA (330-437 omega) compared to OSY and IROX (414-588 omega) from 0 to 28 dai. From 2 to 10 dai, Cg was sig higher in ELA and IROX (3.8-4.2 microF) compared to OSY (3.3-3.4 microF). In the three groups, delta Qmin reached a comparable maximum (1-1.2 microC) at 5 dai. Therefore, vitreous carbon and iridium oxide atrial fixation leads exhibit low chronic polarization effects compared to a standard elgiloy lead, but do not show a sig reduction in charge threshold.

Aged

Loss of atrial signal detection with increasing sensitivity settings: a software problem and its practical solution.

A logic error in the software of Cosmos II results in apparent malsensing of atrial signals at high rather than low sensitivity settings. The paradoxical function was found in 54 out of 87 units tested (= 62%) if the atrial sensitivity was increased and in 54 out of 87 units tested (= 62%) if the atrial sensitivity was increased and the ventricular refractory period was lengthened beyond critical limits that could be defined individually. The problem inherent in both unipolar (66%) and bipolar systems (54%) is attributed to a summation of a physiological signal and internal noise, which has to be detected at the atrial level within a specific timing window to disable atrial sensing for the next pacing cycle. The malfunction virtually disappears if the ventricular refractory period does not exceed 210 msec. It is recommended, therefore, that Cosmos II pacemakers be left at their nominal ventricular refractory value (200 msec).

Adult

Modern strategies for treatment of acute myocardial infarction: significance of haemostaseological and rheological findings.

In spite of equivalent clinical efficacy of various thrombolytic agents for the treatment of acute myocardial infarction there is evidence of different drug-depending influences on the haemostatic and fibrinolytic system. In the present study 40 patients with acute myocardial infarction have been investigated. 20 patients received 750,000 and 1.5 mio U streptokinase (SK), respectively, 10 patients 30 mg anisoylated plasminogen streptokinase activator complex (BRL 26921) and 10 patients a combination of 200,000 U urokinase (UK) and 4.5 mio U pro-urokinase (PUK) as a short-term intravenous treatment. Reperfusion of coronary arteries has been achieved in 70 to 100 percent. Major not fatal bleedings occurred in 2 patients. One patient died within 72 hours after beginning of the myocardial infarction. The longest duration of fibrinolytic activity was observed in the BRL 26921 group (half-disappearance time close to 2 h). It was significantly shorter in the SK groups showing a dose-dependency. Plasma concentration of fibrinogen dropped beyond normal levels following SK and BRL 26921, but not under UK/PUK. Plasma viscosity correlated with fibrinogen decrease and displayed a dose-depending relationship with the presence of SK. Haemorheological effects are suggested to be important for the clinical efficacy of thrombolytic therapy in myocardial infarction.

Anistreplase

Optimised function for determining time to peak creatine kinase and creatine kinase-MB as non-invasive reperfusion indicators after thrombolytic therapy in acute myocardial infarction.

STUDY OBJECTIVE: The aim of the study was to investigate the use of an optimised function to approximate and interpolate the time course of serum creatine kinase and creatine kinase-MB values after thrombolytic therapy in acute myocardial infarction. DESIGN: A three parameter interpolating function was developed which approximates the time course of serum enzyme levels. In the proposed function, time to peak creatine kinase and maximum of creatine kinase determined from raw data were used as starting parameters of the non-linear interpolation routine, thus providing ideal starting conditions for the iteration. The efficacy of the function was compared with that of three other functions cited in published reports (log-normal distribution function, modified gamma density function, three compartment function). SUBJECTS: Serum enzyme data from 20 patients with acute myocardial infarction were used in the comparisons. The patients have all been treated with anisoylated plasminogen streptokinase activator complex. RESULTS: In comparison with the other models, deviations of the experimental model function from the raw data were minimal. The fit remained stable for time intervals between blood samples of up to 6 h. CONCLUSIONS - Due to its numerical stability, the function outlined in this study is suitable for large clinical reperfusion trials. In the case of uncomplicated infarctions without thrombolytic therapy, the area under the creatine kinase activity curve could be directly calculated in terms of maximum activity and time to peak.

Anistreplase

[Third replacement of the tricuspid valve in combined aortic-tricuspid defect].

The implantation of valvular prostheses in tricuspid position is problematic due to the slow blood flow in the low-pressure system with the risk of valvular thromboses. Today there are prostheses with supportable low transvalvular gradients; nevertheless, the risk of valvular thromboses is, in spite of anticoagulation, increased by using mechanical prostheses in tricuspid position. We report on a 51-year-old woman in whom a dysfunction of the Starr Edwards ball prosthesis 15 years after its implantation in the tricuspid position was diagnosed. The prosthesis, which was unable to function due to connective tissue proliferation, was replaced by a St. Jude Medical prosthetic valve, which also presented a disturbed function a few days after the implantation. Because of a suspected valve thrombosis the patient was treated with rt-PA-lysis which proved to be ineffective. Thus it was inevitable to implant a third tricuspid valve; this time a bioprosthesis was used.

Aortic Valve

Atrial signal variations and pacemaker malsensing during exercise: a study in the time and frequency domain.

To give some explanation for atrial malsensing in dual chamber pacing that occurs only during exercise, atrial electrograms from 33 patients were telemetrically recorded and analyzed in both the time and frequency domains. During exercise, an overall decrease from 6.4 +/- 1.9 to 5.6 +/- 1.9 mV (-11%) in the atrial signal amplitude was noted. Despite considerable variability among patients, marked changes occurred in 15 patients whose signals diminished by 11 to 49%. Slew rates showed a similar decrease from 1.35 +/- 0.45 to 1.18 +/- 0.45 V/s (-10.8%), with individual changes of as much as -51%. Signal attenuation in the time domain correlated well with frequency data, exhibiting a highly significant reduction of signal energy between 25 and 105 Hz. However, spectral distribution changed from rest to exercise, with a relative increase of signal energy in the range between 5 and 25 Hz and a decrease at higher frequencies. Individual changes differed widely when low (15 to 65 Hz) and high (65 to 115 Hz) frequencies were compared, but in a group of 11 patients signal attenuation in the high frequency band was more pronounced (-45%) than in the low frequency band (-23%). The clinical impact of the change in frequency distribution during ergometry was visualized by computer simulation of two different (low and high bandpass) filters. Although in individual patients, both characteristics may be favorable with respect to atrial sensing, it was observed in 11 patients that high pass filtering attenuates signal amplitudes by 10 to 24% in excess of the variation without filtering.(ABSTRACT TRUNCATED AT 250 WORDS)

Adolescent

Atrial sensing performance of AV universal pacemakers during exercise.

To investigate the atrial sensing function of dual chamber pacemakers during exercise, we studied 57 patients aged 12 to 81 years (m = 65). They were paced for sinoatrial disorders (n = 15), second or third degree AV block (n = 37), or binodal disease (n = 11). The examination was performed 3-24 months (m = 11) after pacemaker implantation. Individual sensing thresholds were determined at rest in the supine position, and proper detection of atrial signals at the programmed sensitivity level was verified during a period of 5-9 minutes (m = 5.8). Without a change in the program of the unit, symptom-limited bicycle ergometry was performed at a maximum load of 25-200 Watts (m = 95) and 6-channel chest wall electrocardiograms were continuously recorded throughout the test, including recovery. During exercise, 25/57 patients (44%) exhibited poor atrial sensing of the pulse generator; after termination of exercise in 16 of the 25 patients, proper atrial sensing resumed within 1 to 7 minutes of recovery. In the remaining nine cases, ergometry was continued after lowering the sensitivity threshold to half the initial setting or, depending on the pacemaker model, by a value of 0.4 mV. This resulted in normal function of the pulse generator in all patients but one, who needed a sensitivity adjustment of another 0.4 mV. In a subgroup of 25 patients, telemetric atrial electrogram recordings were monitored during ergometry, 19 of which could be evaluated quantitatively. Besides random atrial signal variations, presumably due to ectopic beats or runs, a systematic decrease of the peak-to-peak amplitude by 0.1 to 1.6 mV (3-30%) was observed in 16/19 patients during exercise. Mean signal reduction amounted to 11.8% and was statistically significant at the 0.1% level. It is concluded from these findings that, besides testing of atrial sensing at rest, the follow-up of dual chamber pacemakers should include an exercise test.

Adolescent

High-dose systemic streptokinase and acylated streptokinase-plasminogen complex (BRL 26921) in acute myocardial infarction: alterations of the fibrinolytic system and clearance of fibrinolytic activity.

We report the results of two consecutive studies using intravenous bolus injections of streptokinase (SK) or acylated plasminogen-SK complex (BRL 26921) in patients with acute myocardial infarction (AMI). In the first study, 20 patients received either 750,000 units (U) SK (group IA, n = 10) or 1,500,000 U SK (group IB, n = 10) within 5-10 min intravenously. In the second study 10 consecutive patients received 750,000 U SK within 15 min (group IIA) intravenously. The following 10 consecutive patients received 30 mg BRL 26921 within 2 min (group IIB). Early reperfusion was found in 16 patients in the first study (8 in each group) and in 18 patients in the second study (9 in each group). The decrease of fibrinolytic activity was biphasic with a half-disappearance time of 112.5 min for BRL 26921 and 31 (IA) and 18 (IB) min for SK. alpha 2-Antiplasmin depletion and a decrease of fibrinogen was observed with no differences after bolus injections of SK and of BRL 26921.

Aged

Measurement of factor VIII procoagulant antigen in normal subjects and in hemophilia A patients by an immunoradiometric assay and by an enzyme-linked immunosorbent assay.

Factor VIII coagulant antigen (FVIII:Ag) and FVIII coagulant activity (FVIII:C) were measured in 102 healthy individuals, in 5 hemophilia A carriers and in 21 hemophilia A patients before and after infusion of heat-treated high-purity FVIII concentrates. Factor VIII:Ag was determined by a solid-phase micro enzyme-linked immunosorbent assay (ELISA) using monoclonal antibodies and by a conventional solid-phase immunoradiometric assay (IRMA). Factor VIII:C was assessed using a one-stage assay. The micro ELISA was decidedly more precise than the IRMA. There was a close correlation between the results obtained by the three assays in the plasma of healthy subjects and in hemophilia A carriers. After transfusion of FVIII concentrates to hemophilia A patients, the FVIII:Ag recoveries were significantly lower than the FVIII:C recoveries and the biological half-life of FVIII:Ag was significantly shorter than for FVIII:C. The calculated half-life of FVIII:C was longer than in any previous study.

Adolescent

Factor VIII: C (FVIII: C) recovery and half-life after infusion of steam-treated high purity factor VIII concentrate in severe hemophilia A--comparison of one-stage assay, two-stage assay and a chromogenic substrate assay.

Factor VIII:C recovery and half-life was measured in 16 hemophilia A patients under comprehensively standardized conditions. Each patient received the same lot of a steam-treated high purity FVIII concentrate at a dose of 19-33 U/kg body weight. A comparison was made between the one-stage assay, the two-stage assay and a chromogenic substrate test for FVIII:C determination using a FXa-sensitive chromogenic substrate. Factor VIII:C potency of the administered FVIII concentrate was measured using calibration curves derived from a concentrate standard and FVIII:C plasma levels were read from calibration curves derived from a plasma standard. The chromogenic assay showed a good reproducibility at FVIII:C levels between 0.015 and 0.50 U/ml. The FVIII:C recoveries calculated from the results of the one-stage assay, the two-stage assay and the chromogenic substrate test were 109 +/- 20, 92 +/- 14 and 81 +/- 11% (mean +/- SD), respectively. The elimination half-lives of FVIII:C were calculated by non-linear least square analysis using a modified computerized Gauss-Newton algorithm. The half-lives calculated from the FVIII:C plasma levels measured by the one-stage assay, the two-stage assay and the chromogenic test were 23.8 +/- 6.4, 22.2 +/- 5.7 and 17.1 +/- 4.8 h (mean +/- SD), respectively. No previous study has reported such long half-life values. Our findings indicate that measurements of recoveries and half-lives by the chromogenic FVIII:C assay and by computerized non-linear least square analysis allow the possibility of individualized FVIII replacement therapy.

Adolescent

[Bolus injection of anisoylated plasminogen-streptokinase activator complex (BRL 26921) as an alternative concept of systemic lysis in acute myocardial infarct].

The thrombolytic properties of anisoylated plasminogen streptokinase activator complex (BRL 26921) and clinical results of the treatment were studied in 10 consecutive patients with acute myocardial infarction. Exclusion criteria were general contraindications against thrombolytic therapy and a time interval of more than 4 h between the onset of symptoms and admission to the hospital. All patients received a 250-mg bolus of prednisolone prior to intravenous injection of 30 mg BRL 26921 within 2 min. A continuous infusion of heparin at a dose of 1,000 USPU/h was started 2 h after the injection. Blood pressure was monitored via an arterial line. Arrhythmias and changes in the ST segments were documented by conventional ECG recording and computer-based ECG monitoring. Coronary arteriography and left ventriculography were carried out within 72 h. Besides routine laboratory tests, serial CK and CK-MB activity measurements were carried out. We determined the following hemostaseological parameters before and 15 min, 30 min, 1 h, 4 h, and 12 h after application of BRL 26921: prothrombin time, activated partial thrombosplastin time, thrombin time, thrombin coagulase time, fibrinogen, streptokinaseplasminogen activator activity, plasminogen and alpha-2-antiplasmin. Our results (reperfusion in all patients angiographically and in 7 to 8 of 10 patients from noninvasive criteria) show that BRL 26921 is a highly effective thrombolytic agent in patients with myocardial infarction, when compared with high-dose systemic fibrinolysis. Applied in dosages required for early reperfusion, it does not appear to be selectively thrombolytic and is not free of hypotensive effects in man. The decrease of fibrinolytic activity is biphasic with a half-disappearance time of 112 min.

Adult

[Thrombolytic therapy with ultra-high doses of streptokinase].

Two pilot studies were carried out in patients suffering from acute myocardial infarction. They received high doses of streptokinase and acylated streptokinase-plasminogen activator complex intravenously as a bolus injection. Furthermore, the possibility of using streptokinase bolus therapy was examined in patients with acute vascular occlusion. Due to the limited number and the heterogenicity of cases the results obtained in patients with acute myocardial infarction are to evaluate with some reservation. In patients with acute vascular occlusion streptokinase bolus therapy at high doses proved to be a promising alternative to the conventional therapy.

Coronary Angiography