PubMed Health⌕ Search

Biomedical subjects

Hamid R Zare

Publications and source records attributed to Hamid R Zare.

4 recordsLinked to original sources

Sulfate-selective electrode based on a complex of copper.

A new sulfate-selective electrode based on the complex N,N'-bis(2-amino-1-oxo-phenelenyl)phenylenediamine copper(II) (CuL) as the membrane carrier was developed. The electrode exhibited a good Nernstian slope of -29.5 +/- 0.5 mV/decade and a linear range of 1.0 x 10(-7) - 1.0 x 10(-1) M for sulfate. The limit of detection was 1.0 x 10(-8) M. It has a fast response time of 10 s and can be used for more than three months. The selective coefficients were determined by the fixed interference method (FIM). The electrode could be used in the pH range 3.5 - 8.0. It was employed as an indicator electrode for direct determination of sulfate in pharmacy and cement samples.

Journal Article↗

Computational electrode potential of a coumestan derivative: theoretical and experimental studies.

Electrode potential of a coumestan derivative, an important biological molecule, in aqueous solution is computed theoretically using Self-Consistent Field (SCF) theory at the level of Hartree--Fock and employing 6-31G(d) basis set and also obtained experimentally by employing electrochemical technique of cyclic voltammetry (CV). Frequency calculations have been carried out and thermal corrections and entropies have been taken into account. Polarizable continuum model is used to describe the solvent. The theoretical and experimental values for the standard electrode potential of the studied coumestan are in excellent agreement with each other and there is only 0.001 V discrepancy between experiment and theory. The agreement mutually verifies the accuracy of experimental method and the validity of mathematical model.

Computational Biology↗

Potentiometric determination of silver(I) by selective membrane electrode based on derivative of porphyrin.

The performance of silver metal complexes with meso-tetraphenylporphyrin ([H2T(4-CH3)]PP) as ionophores for ion-selective electrodes was studied. The electrode exhibited linear response with Nernstian slope of 59.2 +/- 1.0 mV per decade within the concentration range of 1.0 x 10(-7)-1.0 x 10(-1) M silver ions. The limit of detection as determined from the intersection of the extrapolated linear segments of the calibration plot, was 1.0 x 10(-7) M. The response time of the electrode was < 10 s over the entire concentration range. The silver-selective electrode exhibited good selectivity for Ag(I) with respect to alkali, alkaline earth and heavy metal ions. The electrodes could be used at least three months without a considerable divergence in their potential. The electrodes are suitable for use in aqueous solutions in a wide pH range of 3.0-9.0. They were used as indicator electrodes in titration of Ag(I) with sodium iodide solution and were successfully applied to direct determination of silver in real samples.

Journal Article↗

Coated wire silver-ion selective electrode based on a N,N'-bis(2-thienylmethylene)-1,2-diaminobenzene.

A novel membrane coated platinum-wire electrode (MCPWE) based on N,N'-bis(2-thienylmethylene)-1,2-diaminobenzene (BTMD) for highly selective determination of Ag+ ion has been developed. The influences of membrane composition and pH on the potentiometric responses of electrode were investigated. The potentiometric responses are independent of the pH of the test solution in the range of 5.0 - 9.0. The electrode shows a linear response for Ag+ ion over the concentration range of 1.0 x 10(-60 to 1.0 x 10(-1) M with a lower detection limit of 6.0 x 10(-7) M. The electrode possesses a Nernstian slope of 59.7 mV decade(-1) and a fast response time of < or = 17 s and can be used for at least 2 months without any observable deviation. The proposed electrode displayed very good selectivity for Ag+ ion with respect to NH4+ and alkali, alkaline earth and some common transition metal ions. The practical utility of the electrode has been demonstrated by its use as the indicator electrode in the potentiometric titration of an AgNO3 solution with a NaI solution and in determination of the silver content of a developed radiological film.

Journal Article↗