Fast HPLC method for the determination of glimepiride, glibenclamide, and related substances using monolithic column and flow program.

This work presents a fast method for the simultaneous separation and determination of glimepiride, glibenclamide, and two related substances by RP LC. The separation was performed on a Chromolith Performance (RP-18e, 100 mm x 4.6 mm) column. As mobile phase, a mixture of phosphate buffer pH 3, 7.4 mM, and ACN (55:45 v/v) was used. Column oven temperature was set to 30 degrees C. The total chromatographic run time was 80 s. This was achieved using a flow program from 5 to 9.9 mL/min. Precisions of the interday and the intraday assay for both retention times and peak areas for the four analyzed compounds were less than 1.2%. The method showed good linearity and recovery. The short analysis time makes the method very valuable for quality control and stability testing of drugs and their pharmaceutical preparations.

Chromatography, High Pressure Liquid↗

Precision from drug stability studies. Investigation of reliable repeatability and intermediate precision of HPLC assay procedures.

A multi-company investigation is presented to obtain and compare precision results for LC assay procedures. Forty-four drug substances and drug products of various types subjected to 156 stability studies, with 2915 assay values in total, were included. This provides an excellent source of real long-term precision estimates, as the same analytical procedure was applied during the whole stability study, extending from 12 to 60 months. Intermediate precision was calculated either using the residual standard deviation of the regression line or applying an analysis of variances, depending on whether there was a significant degradation of the analyte or not. The results show impressively the large intervals where the individually calculated parameters scatter. Distribution ranges and averages for repeatability, intermediate precision, and the ratio between the two precision levels are mainly dependent on the type of drug product. Repeatabilities were found up to 0.8% for solutions, 1.6% for drug substances, 1.9% for tablets, 2.3% for creams, and 3.4% for a bath. For intermediate precision, which includes additional variability factors due to the reference standard, operator, equipment, reagents, etc., a similar dependency was obtained with a slightly changed order: up to 1.1% for drug substances, 2.2% for solutions, 2.3% for tablets, 3.1% for creams, and 3.2% for a bath. The ratio between the precision levels is up to 2.5 and similar for all investigated drug product types, apart from solutions with up to 5.3. These differences for the types of drug product may be explained by the influence of the sample and/or the sample preparation: the more complex, the higher the variability contribution. For the investigated examples, the impact of the analyte and of the concentration (dosage) seems to be of less importance. Therefore, a classification of drug product types for orientation on acceptable precision (ranges) for LC assay seems to be possible.

Algorithms↗

Protein adsorption in fused-silica and polyacrylamide-coated capillaries.

The model proteins cytochrome c, myoglobin, ovalbumin, and beta-lactoglobulin were investigated with regard to their adsorption properties on capillaries for electrophoresis. The model compounds were selected to cover a wide range of properties. Cytochrome c is a basic protein (isoelectric point (pI): 9.6; M(r): 11.7 kDa), beta-lactoglobulin is rather acidic (pI: 5.4, M(r): 18.4 kDa), myoglobin was chosen as a neutral reference protein (pI: 6.8-7.4, M(r): 17.8 kDa), and ovalbumin (pI: 5.1, M(r): 45.0 kDa) was selected as a relatively larger analyte. First, the pH dependence of adsorption was investigated for the bare fused silica. A clear correlation to the respective pIs was noted. For myoglobin and ovalbumin, none or negligible adsorption was found above the pI, whereas strong adsorption was noted just below this parameter. Cytochrome c and beta-lactoglobulin already showed distinct adsorption above their pIs. However, none of the proteins showed any significant adsorption more than one pH unit above the pIs. For linear polyacrylamide-coated capillaries, a decreased but not a complete lack of adsorption was observed. Here, pH-dependent adsorption was noted as well. Regeneration of the capillaries by rinsing with buffers containing 200 mM SDS was also investigated. This method was completely successful for myoglobin, but that too for only freshly-adsorbed protein. After a storage time of 24 h and due to the aging of the adsorbate, a sufficient regeneration was no longer possible.

Adsorption↗

Quantification in capillary electrophoresis-mass spectrometry: long- and short-term variance components and their compensation using internal standards.

Different approaches were chosen to examine ionization reproducibility of analytes after separation by capillary electrophoresis-mass spectrometry (CE-MS) in a commercially available sheath-flow electrospray interface. For this task three different standard samples were examined. Sample 1 contained neostigmine bromide (cationic), paracetamol (PCM) (neutral) and nicotinic acid (anionic component). Results were evaluated using internal standard (IS) calculations. Sample 2 represented an isotopically labelled IS of the quantified substance (PCM/D4-PCM), while sample 3 (neostigmine bromide/scopolamine hydrobromide) provided an IS closely migrating to the tested substance. Furthermore, short-time variations inside the interface were examined by multiple injections of the same substance. For sample 1, the relative standard deviations (RSD%s) were between 8 and 25% (n at least 58) for the peak area ratios. Multiple injected samples gave 5.5-19.4% (n = 25) for peak area RSD%. Using a closely migrating IS, sample 3, RSD%s between 6.5 and 10% (n at least 63) were achieved. With isotopically labelled IS, sample 2, an RSD% of 3-4% was achieved for peak area ratios over long periods (n = 25), for shorter periods (n = 9) even 1-2% RSD% was obtained. Keeping the instrument settings constant, the influence on the ionization efficiency and reproducibility was tested, varying the buffer pH, the organic buffer modifier and the sample concentration. Repeatabilities of migration time and peak area were measured and compared. Two 10 mM ammonium acetate buffers with pH 4.0 and 8.5 were investigated. No influence of buffer pH on peak area reproducibility was found. Isopropanol as organic buffer modifier significantly improved the ionisation leading to larger peak areas, but reduced reproducibility. The basic buffer produced slightly better RSD%s for migration times (2.5-4.0%) (n = 180) and faster analysis for the different test analytes of sample 1, while with the acetic buffer, RSD%s from 3.9 to 6.0% were obtained (n at least 163). The positioning of the capillary turned out to be the crucial parameter to ensure reproducible results. Thus, a procedure was established to ensure a defined ion-intensity level after capillary changes. The investigation of the different sample concentrations gave negligible differences in RSD%, showing that the signal-to-noise ratio was not the crucial parameter for reproducibility here, in contrast to CE-UV detection.

Acetaminophen↗

Precision and variance components in quantitative gel electrophoresis.

The error in quantitative gel electrophoresis/Western blotting was investigated considering the purity testing of erythropoietin. The overall error was over 35% relative standard deviation. However, an analysis of variance elucidated that the interoperator variability was the dominant error source, which already explained almost 80% of the total variance. Careful compilation and investigation of the possible error sources strongly indicates that the immunoreaction after blotting and the subsequent color reaction are the major error sources in this case.

Analysis of Variance↗

Quantitation in capillary electrophoresis-mass spectrometry.

CE-MS has evolved into a strong alternative to LC-MS. Most of CE-MS applications deal with characterization and identification. However, quantitative aspects have gained importance in, e.g., pharmaceutical and biotechnological applications. Here we summarize and evaluate various methodological aspects in order to achieve sensitive and reproducible results. Similar to LC-MS, aspects of matrix influence on the electrospray process need to be carefully addressed when quantitative results are intended by CE-MS. Due to a more complicated coupling special emphasis needs to be put on the CE-MS interface. Generally linearity over more than three orders of magnitude can be achieved by CE-ESI-MS. Furthermore, a literature survey has been performed in order to give an overview over quantitative measurements performed by CE-MS. The precision can be doubled when changing from a structural related to an isotopically labeled internal standard. Thus a level of precision better than 5% RSD can be achieved.

Biotechnology↗

Capillary isoelectric focusing--reproducibility and protein adsorption.

Capillary isoelectric focusing (CIEF) is an important tool for the quality assurance of biotechnologically maintained drugs and for proteome analysis. The critical performance parameters of this technique are the precisions of isoelectric point (pI) values and peak areas. Compared to capillary zone electrophoresis (CZE), where precise results can be obtained (e.g., 0.5% relative standard deviation (RSD) for peak areas, n = 60), only few data are available for CIEF experiments. So far, reproducible data of pI values (RSD = 0.5%) have been acquired, but peak areas show inferior results (about 3-15% RSD). Nonstable capillary coatings and protein adsorption have been discussed as possible reasons. Recent work of Righetti et al. [25, 27] has proven that the use of coated capillaries can reduce the adsorption of proteins by 50% but cannot prevent it. In our CIEF experiments irregular and poorly reproducible peak patterns have been observed. In a long-time experiment of 106 repeated runs, an overall RSD of 10% was obtained for peak areas, RSD of 2% only in series of about 10 consecutive replicates. Especially at higher concentrations the reproducibility deteriorates. This seems to be the result of a self-amplifying process, induced by adsorbed protein molecules, leading to further agglomerations. CZE control experiments in linear polyacrylamide (LPA)-coated capillaries proved a strong pH dependency of these effects within a small range. Compared to bare fused-silica surfaces, adsorption effects are reduced but not inhibited. An enhancement of reproducibility in CIEF experiments can be achieved only by controlling the interactions of proteins and capillary walls.

Adsorption↗

Capillary electrophoresis-a high performance analytical separation technique.

Capillary electrophoresis (CE) is often one of the preferred techniques in pharmaceutical quality control and in clinical chemistry, particularly considering the high selectivity and lower costs compared to HPLC. The precision of CE is as good as in liquid chromatography (LC). The sample-throughput is high due to short analysis times. Efforts for sample pre-treatment are usually minor in CE. Urine and even blood plasma can be directly injected without further pre-treatment. After summarising the basic principles of CE, general strategies for method development are described to achieve selective, efficient, precise, fast, sensitive, and validated methods. Sample pre-treatment requirements are discussed. Standard buffer recipes, surfactants used in micellar electrokinetic chromatography (MEKC), chiral selectors, useful buffer additives, actions to deal with complex matrices, and aspects of validation have been collected. Other techniques that can be performed with CE instruments, such as capillary isoelectric focusing (CIEF), capillary isotachophoresis (CITP), and capillary electrochromatography (CEC), are briefly discussed.

Acetylcysteine↗

Electroosmotic flow variations caused by the volatility of buffer components: diagnosis and therapy.

In order to separate a polar amine pharmaceutical and its potential impurities, a micellar electrokinetic chromatography method was developed. The main compound and 11 other substances were completely separated using a 20 mM Tris buffer, pH 8.0, containing 50 mM sodium dodecylsulfate (SDS) and 24% (v/v) acetonitrile. However, a strong, continuous reduction in the EOF occurred and quantification was not possible. The EOF reproducibility could not be improved by suitable rinsing procedures. Surface effects or interactions did not cause the EOF changes, but the evaporation of acetonitrile was identified as the major source for EOF instability. However, a high concentration of acetonitrile was decisive for selectivity. Thus a reliable protection against the evaporation of this buffer constituent had to be found. Paraffin and various silicon oils were tested as covering film. In order to quickly test buffer systems if the evaporation of electrophoresis solutions is acceptable, an alternative experimental design without doing CE experiments had to be found. Electrical conductivity was chosen as parameter, because it can be determined simply and fast. The buffers under investigation were placed in a 50-ml beaker with a magnetic stirring rod, placed on a magnetic stirrer. The buffer solution was kept in motion at 120 rev./min. The beaker was covered around the measuring head with laboratory film and in addition to this with paraffin or silicon oil. An acetonitrile content up to 10% (v/v) was acceptable if a coverage was used. The various cover liquids had a similar effect. A content of 15% (v/v) already increased the evaporation effect significantly. Higher acetonitrile contents are not acceptable. A buffer similar to the originally transferred method, 20 mM Tris (pH 8.0), 50 mM SDS containing 10% (v/v) acetonitrile as well as 10% (v/v) isopropanol showed an acceptably low evaporation in the conductivity experiments. With this buffer, a stable EOF was also obtained. Conductivity measurements are generally applicable to quickly test buffers that contain organic solvents.

Buffers↗