PubMed Health⌕ Search

Biomedical subjects

J Hosaka

Publications and source records attributed to J Hosaka.

25 records · Page 2Linked to original sources

[A hanger for thoracic surgery].

A new hanger for thoracic operation is introduced. This hanger for draping cloth is made for the purpose 1) of isolating the operative field from the anesthetist, 2) of keeping the broad hygienic area around the operator, 3) of keeping the area to handle the endotracheal tube for the anesthetist, 4) of keeping the area on the cranial side of the patient where the second assistant is able to stand, 5) of keeping the area for the watcher to stand.

Humans↗

[In vitro antifungal activity of itraconazole, a new triazole antifungal agent, against clinical isolates from patients with dermatomycoses].

In vitro antifungal activities of itraconazole (ITZ), a triazole antifungal agent, against clinical isolates obtained from patients with superficial and subcutaneous mycoses were examined using the agar dilution method on casitone agar. The clinical isolates tested were 7 species and 263 isolates including Trichophyton mentagrophytes (104 isolates), Trichophyton rubrum (103 isolates), Microsporum canis (3 isolates), Epidermophyton floccosum (2 isolates), Candida albicans (32 isolates), Malassezia furfur (7 isolates) and Sporothrix schenckii (12 isolates). The results are summarized as follows: 1. MIC values of ITZ for the isolates of dermatophytes and M. furfur distributed in a range of less than 0.0012-5 micrograms/ml indicating that ITZ had greater in vitro activities. These in vitro activities of ITZ were greater than those of clotrimazole or bifonazole. 2. C. albicans isolates were divided into 2 groups in terms of ITZ-susceptibilities, a high susceptibility group and low-susceptibility group with MIC values of 0.02-0.08 micrograms/ml and greater than 10 micrograms/ml, respectively. 3. The in vitro activities of ITZ against S. schenckii isolates with a geometric mean MIC of 0.119 micrograms/ml were greater than those of ketoconazole, miconazole or amphotericin B used as reference drugs.

Amphotericin B↗

[A case of multiple peripheral hepatic artery aneurysms after chemoembolization].

A case of 57-year-old man with multiple peripheral hepatic artery aneurysms is reported. The patient was admitted with liver dysfunction and detected hepatoma in the right lobe. Chemoembolization was performed for the treatment of hepatoma. After 35 days, reangiography revealed multiple peripheral hepatic artery aneurysms in the region of chemoembolization. The etiology of aneurysms were presumably drug induced arteritis.

Aneurysm↗

[Rupture of right gastric artery aneurysm--a case report and the usefulness of TAE].

A case of a 39-year-old male with ruptured right gastric artery aneurysm was reported. An emergency angiography was carried out followed by transarterial embolization. The TAE was of great value for the control of massive bleeding and for getting out of the urgent surgical operation. In the present report the clinical value of TAE as well as frequency and some diagnostic imagings on this disease was discussed.

Adult↗

In vitro evaluation of a new temporary venous filter: the Spring filter.

PURPOSE: To evaluate in vitro the function efficacy of a new variable-sized, temporary venous filter, the Spring filter (SF). METHODS: The SF was tested in a flow phantom, using flexible, thin-walled polyethylene tubes 12, 14, and 16 mm in diameter to simulate veins. Clots of three sizes were used: 6 x 10 mm, 6 x 20 mm, 9 x 20 mm. Filter deployment was performed, ensuring that its functional diameter after placement was a predetermined multiple of the radius 'r' of the tube: 3.0r, 2.8r, 2.6r, 2.4r. The terminal coil was visually examined for changes in configuration. Clot retention by the filter for a period of 3 min was recorded as a success. The baseline intraluminal pressure and the maximum pressure reached after clot introduction were recorded. The trapping ability was calculated from the results for 10 clots. The effect of the following parameters on filter function were studied: size of embolus, caliber of simulated vein, and functional diameter of filter. RESULTS: The filter configuration was stable. Not a single instance of axial tilting occurred on deployment. Functional diameter was the major determinant of filter function. One hundred percent of clots were trapped irrespective of clot size and tube caliber when the functional diameter was 3.0r. Only 73% (95% confidence interval (CI): 57% - 88%) and 77% (95% CI: 61% - 92%) of small emboli were held up by the filter when the functional diameters were 2.6r and 2.4r, respectively (p = 0.0001). A significant reduction in clot trapping was encountered even with medium clots when the functional diameter was 2.4r (p = 0.02). CONCLUSION: Against the background of available data on retrievable vena caval filters, the current model of SF warrants further investigation.

Equipment Design↗

Multiple emboli and filter function: an in vitro comparison of three vena cava filters.

PURPOSE: To establish the influence of number of emboli on the trapping ability of vena cava filters in vitro. MATERIALS AND METHODS: Three filters, the titanium Greenfield, Vena Tech-LGM, and Günther Tulip retrievable, were studied with use of 20- or 26-mm-diameter tubes to simulate the inferior vena cava. In the first protocol, five small (4 x 20-mm) or medium (6 x 10-mm) emboli were delivered in sequence, and the fate of each was recorded. In the second protocol, medium or large (6 x 30-mm) clots were sequentially introduced until filter occlusion occurred or 50 clots had been delivered. RESULTS: For the first protocol, 82% of first small clots and 60% of second clots were trapped in 20-mm tubes (P = .001) and 63% and 45%, respectively, were trapped in 26-mm tubes (P - .02). With medium clots, the proportion trapped also dropped significantly with ascending clot rank. In the second protocol the proportion of clots captured was invariably higher for the first 10 clots (P < .001 for all combinations of covariables). CONCLUSION: Filter function deteriorates with number of emboli delivered, irrespective of embolus size and simulated vein caliber.

Confidence Intervals↗

Induced thrombosis in the pig inferior vena cava: a model of deep venous thrombosis.

PURPOSE: To establish a new animal model of deep venous thrombosis. MATERIALS AND METHODS: Fifteen young pigs underwent temporary interruption of the inferior vena cava (IVC) below the entry of the right renal vein by means of either a silicone band (surgical technique, n = 6) or an intraluminal balloon catheter (endovascular technique, n = 9), followed by injection of absolute ethanol. Lumbar veins within 3 cm below the obstruction were ligated or occluded interventionally. The iatrogenic caval obstruction was relieved after 2 days. RESULTS: Procedure-related mortality was 33% (n = 2) and 11% (n = 1) for the surgical and endovascular groups, respectively. An adherent, occlusive thrombus was found in all four of the remaining surgically treated animals and in six of eight animals treated percutaneously. The IVC remained patent in two animals in whom balloon migration occurred. Severe caval stenosis invariably occurred after surgical banding. CONCLUSION: IVC thrombi suitable for the study of various recanalization therapies can be reliably created with this pig model.

Angioplasty, Balloon↗