Polycentric total knee arthroplasty.
Polycentric total knee arthroplasty in 56 patients provided significant relief of pain in 90.7% and a stable knee in 92.7%. Postoperative increase in range of motion averaged 13.7 degrees for all patients (18.5 degrees in rheumatoid arthritis and 8.7 degrees in degenerative arthritis). This increase in excursion is attributed to the design of the prosthesis, a vigorous postoperative rehabilitation program, and careful surgical technic. A failure rate of 12.7% was due to infection, mechanical instability and unexplained pain. Contraindications to surgery and technical recommendations are noted.
Arthritis, Rheumatoid↗