In vitro performance test system for pulse oximeters.
An in vitro system was developed capable of testing the accuracy and reproducibility of pulse oximeter readings. The pulse oximeter probe receives signals through a pulsating blood cuvette. The development of the design of the cuvette is described. Using the final design (or 'model finger'), a comparison is made between readings from a Datex Satlite pulse oximeter (SpO2) and saturation values obtained by use of a multi-wavelength bench oximeter (SaO2). Linear regression analysis of the data gives SpO2 = 0.88 SaO2 + 11.2 (r = 0.979, p < 0.001).