Dose variations with varying calculation grid size in head and neck IMRT.

Ever since the advent and development of treatment planning systems, the uncertainty associated with calculation grid size has been an issue. Even to this day, with highly sophisticated 3D conformal and intensity-modulated radiation therapy (IMRT) treatment planning systems (TPS), dose uncertainty due to grid size is still a concern. A phantom simulating head and neck treatment was prepared from two semi-cylindrical solid water slabs and a radiochromic film was inserted between the two slabs for measurement. Plans were generated for a 5,400 cGy prescribed dose using Philips Pinnacle(3) TPS for two targets, one shallow ( approximately 0.5 cm depth) and one deep ( approximately 6 cm depth). Calculation grid sizes of 1.5, 2, 3 and 4 mm were considered. Three clinical cases were also evaluated. The dose differences for the varying grid sizes (2 mm, 3 mm and 4 mm from 1.5 mm) in the phantom study were 126 cGy (2.3% of the 5,400 cGy dose prescription), 248.2 cGy (4.6% of the 5,400 cGy dose prescription) and 301.8 cGy (5.6% of the 5,400 cGy dose prescription), respectively for the shallow target case. It was found that the dose could be varied to about 100 cGy (1.9% of the 5,400 cGy dose prescription), 148.9 cGy (2.8% of the 5,400 cGy dose prescription) and 202.9 cGy (3.8% of the 5,400 cGy dose prescription) for 2 mm, 3 mm and 4 mm grid sizes, respectively, simply by shifting the calculation grid origin. Dose difference with a different range of the relative dose gradient was evaluated and we found that the relative dose difference increased with an increase in the range of the relative dose gradient. When comparing varying calculation grid sizes and measurements, the variation of the dose difference histogram was insignificant, but a local effect was observed in the dose difference map. Similar results were observed in the case of the deep target and the three clinical cases also showed results comparable to those from the phantom study.

Algorithms↗

The effect of intravenous contrast on intensity-modulated radiation therapy dose calculations for head and neck cancer.

OBJECTIVE: The objective of this study was to determine if the use of intravenous contrast results in clinically important errors in intensity-modulated radiation therapy (IMRT) dose calculations for head and neck radiotherapy treatment planning. MATERIALS AND METHODS: Nonionic, iodinated intravenous contrast (Iohexol) was administered during the treatment planning computed tomography (CT) scan of 5 patients with head and neck cancer of varying disease sites. The potential effect of intravenous contrast was studied by changing the density of the contrast-enhanced vessels. An inverse IMRT plan was generated from an unmanipulated "normal contrast" planning scan. We then applied the same planning parameters to a "no contrast" planning scan. The effect of intravenous contrast was quantified by calculating the percent change of dose in a variety of target and normal structures. To evaluate a worst-case scenario situation, this comparison was repeated by assigning the vessels the maximum density in our planning system ("maximum contrast" density plan). RESULTS: Dose differences between a planning set of images using intravenous contrast and a set of images without contrast were less than 0.2% for all relevant target volumes and critical structures. A worst-case scenario in which normal contrast was overridden with "maximum contrast" led to small dose differences, generally less than 0.5%. CONCLUSIONS: Planning head and neck IMRT from CT scans that contain intravenous contrast does not result in clinically important errors in dose delivery.

Contrast Media↗

A novel dose uncertainty model and its application for dose verification.

Based on statistical approach, a novel dose uncertainty model was introduced considering both nonspatial and spatial dose deviations. Non-space-oriented uncertainty is mainly caused by dosimetric uncertainties, and space-oriented dose uncertainty is the uncertainty caused by all spatial displacements. Assuming these two parts are independent, dose difference between measurement and calculation is a linear combination of nonspatial and spatial dose uncertainties. Two assumptions were made: (1) the relative standard deviation of nonspatial dose uncertainty is inversely proportional to the dose standard deviation sigma, and (2) the spatial dose uncertainty is proportional to the gradient of dose. The total dose uncertainty is a quadratic sum of the nonspatial and spatial uncertainties. The uncertainty model provides the tolerance dose bound for comparison between calculation and measurement. In the statistical uncertainty model based on a Gaussian distribution, a confidence level of 3sigma theoretically confines 99.74% of measurements within the bound. By setting the confidence limit, the tolerance bound for dose comparison can be made analogous to that of existing dose comparison methods (e.g., a composite distribution analysis, a gamma test, a chi evaluation, and a normalized agreement test method). However, the model considers the inherent dose uncertainty characteristics of the test points by taking into account the space-specific history of dose accumulation, while the previous methods apply a single tolerance criterion to the points, although dose uncertainty at each point is significantly different from others. Three types of one-dimensional test dose distributions (a single large field, a composite flat field made by two identical beams, and three-beam intensity-modulated fields) were made to verify the robustness of the model. For each test distribution, the dose bound predicted by the uncertainty model was compared with simulated measurements. The simulated measurements were within the tolerance bound as expected by a statistical prediction of the model. Using the dose uncertainty distributions, an uncertainty length (uncertainty area and uncertainty volume for two-dimensional and three-dimensional, respectively) histogram (a plot of the dose uncertainty of 1sigma received by a length of field) was made. The histogram provides additional information on superiority of a treatment plan in terms of uncertainty. In summary, the uncertainty model provides the dose comparison tool as well as the evaluation tool of a treatment planning system.

Algorithms↗

Evaluation of surface and build-up region dose for intensity-modulated radiation therapy in head and neck cancer.

Despite much development, there remains dosimetric uncertainty in the surface and build-up regions in intensity-modulated radiation therapy treatment plans for head and neck cancers. Experiments were performed to determine the dosimetric discrepancies in the surface and build-up region between the treatment planning system (TPS) prediction and experimental measurement using radiochromic film. A head and neck compression film phantom was constructed from two semicylindrical solid water slabs. Treatment plans were generated using two commercial TPSs (PINNACLE3 and CORVUS) for two cases, one with a shallow (approximately 0.5 cm depth) target and another with a deep (approximately 6 cm depth) target. The plans were evaluated for a 54 Gy prescribed dose. For each case, two pieces of radiochromic film were used for dose measurement. A small piece of film strip was placed on the surface and another was inserted within the phantom. Overall, both TPSs showed good agreement with the measurement. For the shallow target case, the dose differences were within +/- 300 cGy (5.6% with respect to the prescribed dose) for PINNACLE3 and +/- 240 cGy (4.4%) for CORVUS in 90% of the region of interest. For the deep target case, the dose differences were +/- 350 (6.5%) for PINNACLE3 and +/- 260 cGy (4.8%) for CORVUS in 90% of the region of interest. However, it was found that there were significant discrepancies from the surface to about 0.2 cm in depth for both the shallow and deep target cases. It was concluded that both TPSs overestimated the surface dose for both shallow and deep target cases. The amount of overestimation ranges from 400 to 1000 cGy (approximately 7.4% to 18.5% with respect to the prescribed dose, 5400 cGy).

Algorithms↗

A comparison of step-and-shoot leaf sequencing algorithms that eliminate tongue-and-groove effects.

The performances of three recently published leaf sequencing algorithms for step-and-shoot intensity-modulated radiation therapy delivery that eliminates tongue-and-groove underdosage are evaluated. Proofs are given to show that the algorithm of Que et al (2004 Phys. Med. Biol. 49 399-405) generates leaf sequences free of tongue-and-groove underdosage and interdigitation. However, the total beam-on times could be up to n times those of the sequences generated by the algorithms of Kamath et al (2004 Phys. Med. Biol. 49 N7-N19), which are optimal in beam-on time for unidirectional leaf movement under the same constraints, where n is the total number of involved leaf pairs. Using 19 clinical fluence matrices and 100000 randomly generated 15 x 15 matrices, the average monitor units and number of segments of the leaf sequences generated using the algorithm of Que et al are about two to four times those generated by the algorithm of Kamath et al.

Algorithms↗

Optimal leaf sequencing with elimination of tongue-and-groove underdosage.

The individual leaves of a multileaf collimator (MLC) have a tongue-and-groove or stepped-edge design to minimize leakage radiation between adjacent leaves. This design element has a drawback in that it creates areas of underdosages in intensity-modulated photon beams unless a leaf trajectory is specifically designed such that for any two adjacent leaf pairs, the direct exposure under the tongue-and-groove is equal to the lower of the direct exposures of the leaf pairs. In this work, we present a systematic study of the optimization of a leaf sequencing algorithm for segmental multileaf collimator beam delivery that completely eliminates areas of underdosages due to tongue-and-groove or stepped-edge design of the MLC. Simultaneous elimination of tongue-and-groove effect and leaf interdigitation is also studied. This is an extension of our previous work (Kamath et al 2003a Phys. Med. Biol. 48 307) in which we described a leaf sequencing algorithm that is optimal for monitor unit (MU) efficiency under most common leaf movement constraints that include minimum leaf separation. Compared to our previously published algorithm (without constraints), the new algorithms increase the number of sub-fields by approximately 21% and 25%, respectively, but are optimal in MU efficiency for unidirectional schedules.

Algorithms↗

Algorithms for optimal sequencing of dynamic multileaf collimators.

Dynamic multileaf collimator (DMLC) intensity modulated radiation therapy (IMRT) is used to deliver intensity modulated beams using a multileaf collimator (MLC), with the leaves in motion. DMLC-IMRT requires the conversion of a radiation intensity map into a leaf sequence file that controls the movement of the MLC while the beam is on. It is imperative that the intensity map delivered using the leaf sequence file be as close as possible to the intensity map generated by the dose optimization algorithm, while satisfying hardware constraints of the delivery system. Optimization of the leaf-sequencing algorithm has been the subject of several recent investigations. In this work, we present a systematic study of the optimization of leaf-sequencing algorithms for dynamic multileaf collimator beam delivery and provide rigorous mathematical proofs of optimized leaf sequence settings in terms of monitor unit (MU) efficiency under the most common leaf movement constraints that include leaf interdigitation constraint. Our analytical analysis shows that leaf sequencing based on unidirectional movement of the MLC leaves is as MU efficient as bi-directional movement of the MLC leaves.

Algorithms↗

Optimal field splitting for large intensity-modulated fields.

The multileaf travel range limitations on some linear accelerators require the splitting of a large intensity-modulated field into two or more adjacent abutting intensity-modulated subfields. The abutting subfields are then delivered as separate treatment fields. This workaround not only increases the treatment delivery time but it also increases the total monitor units (MU) delivered to the patient for a given prescribed dose. It is imperative that the cumulative intensity map of the subfields is exactly the same as the intensity map of the large field generated by the dose optimization algorithm, while satisfying hardware constraints of the delivery system. In this work, we describe field splitting algorithms that split a large intensity-modulated field into two or more intensity-modulated subfields with and without feathering, with optimal MU efficiency while satisfying the hardware constraints. Compared to a field splitting technique (without feathering) used in a commercial planning system, our field splitting algorithm (without feathering) shows a decrease in total MU of up to 26% on clinical cases and up to 63% on synthetic cases.

Algorithms↗

Leaf sequencing algorithms for segmented multileaf collimation.

The delivery of intensity-modulated radiation therapy (IMRT) with a multileaf collimator (MLC) requires the conversion of a radiation fluence map into a leaf sequence file that controls the movement of the MLC during radiation delivery. It is imperative that the fluence map delivered using the leaf sequence file is as close as possible to the fluence map generated by the dose optimization algorithm, while satisfying hardware constraints of the delivery system. Optimization of the leaf sequencing algorithm has been the subject of several recent investigations. In this work, we present a systematic study of the optimization of leaf sequencing algorithms for segmental multileaf collimator beam delivery and provide rigorous mathematical proofs of optimized leaf sequence settings in terms of monitor unit (MU) efficiency under most common leaf movement constraints that include minimum leaf separation constraint and leaf interdigitation constraint. Our analytical analysis shows that leaf sequencing based on unidirectional movement of the MLC leaves is as MU efficient as bidirectional movement of the MLC leaves.

Motion↗