Absorbed radiation dose in mammography.

Radiation dose from mammographic techniques was determined as a function of surface exposure, beam quality, and depth. Relative exposure vs. depth was measured in tissue-substitute materials by thermoluminescent dosimetry. The f-factors were calculated from elemental compositions of mastectomy specimens. Dose at depth depends on beam quality as well as exposure and tissue composition. Analysis of data from the ACS/NCI Screening Centers shows current average midbreast doses to be 25 times lower (film/screen) and 3 times lower (Xerox) than the 2 rads previously estimated. Quantitative risk indicators other than midbreast dose are also discussed.

Adipose Tissue

Pretransfusion blood irradiation: clinical rationale and dosimetric considerations.

The irradiation of blood before transfusion into immunosuppressed patients is an increasingly common technique used to prevent graft-versus-host disease. A technical procedure is described for the calibration of blood irradiators, including the determination of absolute dose rate and relative dose distribution over the blood volume. Results of dose rate measurements on commercially available irradiators indicate differences of +5% to -13% with manufacturer-supplied calibrations and variations in the relative dose rate over the irradiation volume from 70% to 180%. The clinical implications of these findings and the need for accurate dosimetry are discussed.