Half-life determination of 88Y and 89Sr.

Half-life measurements have been carried out at LNHB for 88Y using a 4pi gamma-ionisation chamber and for 89Sr using a proportional counter. The determined half-life values and associated standard uncertainties are 106.63 +/- 0.05 d for 88Y and 50.65 +/- 0.05 d for 89Sr, being consistent with relevant values reported in literature. Based on the present results and relevant literature values revised recommended half-life values and associated standard uncertainties are proposed, viz. 106.626 +/- 0.021 d for 88Y and 50.57 +/- 0.03 d for 89Sr.

France↗

Towards absolute activity measurements by ionisation chambers using the PENELOPE Monte-Carlo code.

The Monte-Carlo code PENELOPE for Ionisation Chamber Simulation Method has been applied for the calculation of ionisation-chamber (IC) calibration factors. Measuring only a few radionuclides well selected within the relevant energy range, and determining an adjustable parameter, commonly used radionuclides can be measured without any specific calibration. The simulation revealed a discontinuity in the IC response as a function of photon energy and its dependence on the chemical composition of the radioactive solution.

Calibration↗

Study of XK and gamma photon emission following decay of 154Eu.

A joint project has been established between VNIIM (D.I. Mendeleyev Institute for Metrology) and LNHB (Laboratoire National Henri Becquerel) to determine as accurately as possible the X- and gamma-ray emission probabilities of 154Eu. Point sources were prepared by VNIIM, and absolute measurements of activity per unit mass were undertaken by both laboratories using coincidence, anti-coincidence and 4pi-gamma counting methods. Other point sources and one aliquot were also prepared for precise gamma-ray spectrometry measurements. Absolute photon emission probabilities were determined with a maximum uncertainty of 0.5% for the most intense lines, supporting the development of this nuclide as a multigamma standard.

Algorithms↗

Calculation of 18F, 99mTc, 111In and 123I calibration factor using the penelope ionization chamber simulation method.

A new method using the Monte Carlo code PENELOPE for ionization chamber simulation has already been successfully used for calculating calibration factors needed for the measurements of radionuclides with photon emission (2003, Appl. Radiat. Isot., to be published). This work has been continued at (Laboratoire National Henri Becquerel) in order to calculate the calibration factors for radionuclides with short half-lives used in medical services. Activity measurements of 18F, 99mTc,111In and 123I using the calculated calibration factors were obtained with standard uncertainties equal to 0.6% for 18F, 99mTc and 1.5% for 111In and 123I.