PubMed Health⌕ Search

Biomedical subjects

M Tsuneoka

Publications and source records attributed to M Tsuneoka.

25 records · Page 2Linked to original sources

Determination of left ventricular volumes with use of DSA density values.

A computer program was developed to obtain left ventricular volumes from the density values of DSA. The highest density value along the horizontal pixel line over the left ventricle in 30 degrees RAO was considered to correspond to the greatest width on the same level of 60 degrees LAO. The left ventricular volume was obtained by converting the density values into the width and integrating these values. This method estimated experimental balloon volumes with considerable accuracy. Clinical evaluation revealed good correlation between DSA and conventional cine left ventriculography.

Angiography↗

Digital TV tomography: description and physical assessment.

A digital TV tomography system, capable of retrospective reconstruction of multiple digital tomographic images, has been developed and its basic physical characteristics have been evaluated. The multiple tomographic images were formed through retrospective reconstruction of digital data acquired on a linear tomographic x-ray unit and an image intensifier-television system. Digital data were obtained with a series of 30 pulsed exposures during linear motion of the system. A distortion correction algorithm for the convex surface of the image intensifier was developed in order to reduce image distortion. A phantom study showed that the square-wave response at the fulcrum plane was slightly inferior to that in conventional tomography. There was also a slight decrease in the square-wave response away from the fulcrum plane and upon application of a correction algorithm, as compared with the response of the original reconstructed image at the fulcrum plane. The exposure dose for a single image was approximately half that in conventional tomography. Because of many advantages, including low exposure, short examination time, digital image manipulation, and applicability to picture archiving and communication systems, this is likely to become an important method in radiology when further technical refinements have been made.

Angiography, Digital Subtraction↗

Clinical evaluation of digital TV tomography.

A unit for digital TV tomography (DTT) was constructed on the basis of retrospective reconstruction of digital data acquired during a linear tomographic motion of an X-ray tube and an image intensifier with a television camera (I.I.-TV) assembly. Following descriptions of the basic principle and system, clinical results obtained with this unit are presented. Comparison with conventional tomography of the chest, skull, abdomen, and bones and joints showed equal diagnostic information in more than 50% of cases. Application to intra-arterial and intravenous DSA provided excellent images, but no new information was obtained. DTT has the advantages of technical simplicity, short examination time, low exposure dose, and compatibility with a PACS system. This technique will become an important part of a modern radiology department.

Aged↗