PubMed Health⌕ Search

Biomedical subjects

Miguel A Mayosky

Publications and source records attributed to Miguel A Mayosky.

2 recordsLinked to original sources

A practical approach to electrode-skin impedance unbalance measurement.

Unbalance between electrode-skin impedances is a major problem in biopotential recordings, leading to increased power-line interference. This paper proposes a simple, direct method to measure that unbalance at power-line frequency (50-60 Hz), thus allowing the determination of actual recording conditions for biopotential amplifiers. The method is useful in research, amplifier testing, electrode design and teaching purposes. It has been experimentally validated by using both phantom impedances and real electrode-skin impedances.

Amplifiers, Electronic↗

Two-electrode biopotential measurements: power line interference analysis.

In this paper, an analysis of power line interference in two-electrode biopotential measurement amplifiers is presented. A model of the amplifier that includes its input stage and takes into account the effects of the common mode input impedance Z(C) is proposed. This approach is valid for high Z(C) values, and also for some recently proposed low-Z(C) strategies. It is shown that power line interference rejection becomes minimal for extreme Z(C) values (null or infinite), depending on the electrode-skin impedance's unbalance deltaZ(E). For low deltaZ(E) values, minimal interference is achieved by a low Z(C) strategy (Z(C) = 0), while for high deltaZ(E) values a very high Z(C) is required. A critical deltaZ(E) is defined to select the best choice, as a function of the amplifier's Common Mode Rejection Ratio (CMRR) and stray coupling capacitances. Conclusions are verified experimentally using a biopotential amplifier specially designed for this test.

Amplifiers, Electronic↗