Mechanical and metallurgical properties of carotid artery clamps.

The mechanical and metallurgical properties of carotid artery clamps were evaluated. The pressure plate retreat propensity, metallurgical composition, surface morphology, magnetic properties, and corrosion resistance of the Crutchfield, Selverstone, Salibi, and Kindt clamps were tested. None of the clamps showed evidence of pressure plate retreat. The clamps differed significantly in their composition, surface cleanliness, magnetic properties, and corrosion resistance. The Crutchfield clamp was the only one manufactured from an ASTM-ANSI-approved implantable stainless steel (AISI 316) and the only clamp in which the surfaces were clean and free of debris. The Selverstone clamp was made principally from AISI 304 stainless steel, as was one Salibi clamp. The pressure plate on another Salibi clamp was made from a 1% chromium and 1% manganese steel. Machining and surface debris consisting principally of aluminum, silicon, and sulfur was abundant on the Selverstone and Salibi clamps. The Kindt clamp was manufactured from AISI 301 stainless steel with a silicate-aluminized outer coating. The Crutchfield and Selverstone clamps were essentially nonferromagnetic, whereas the Salibi and Kindt clamps were sensitive to magnetic flux. In the pitting potential corrosion test, the Crutchfield clamp demonstrated good corrosion resistance with a pitting potential of 310 mV and no surface corrosion or pitting by scanning electron microscopy examination. The Selverstone clamp had lower pitting potentials and showed various degrees of corrosion and surface pitting by scanning electron microscopy. The Salibi pressure plate had a very low pitting potential of -525 mV and showed severe corrosion. By metallurgical criteria, only the Crutchfield clamp is suitable for long term implantation.

Animals↗

Corrosion of aneurysm clips: evaluation and clinical implications. Part 1: Metallurgical foundations of stress corrosion failure.

Several intracranial failures of aneurysm clips have been documented recently. Consistently, the mechanism of failure has been stress corrosion. These failures could have been anticipated if certain metallurgical properties of the clips had been known. In this paper, we review the metallurgical basis of stress corrosion failure and offer serial metallurgical tests as screening procedures to determine a clip's propensity towards failure. The importance of determining a clip's fracture toughness, a key indicator of stress corrosion resistance, is emphasized.

Corrosion↗

Corrosion of aneurysm clips: evaluation and clinical implications. Part II: Individual performance.

The intracranial fracture of aneurysm clips due to stress corrosion failure is an increasingly common long term postoperative complication. To study the susceptibility of different clips to this process, we evaluated seven aneurysm clips (Mayfield, Yasargil, Vari-Angle McFadden, Scoville, Heifetz, and Pivot) in vitro. Microstructures, hardness, chemical composition, and susceptibility to stress corrosion failure were all measured by standard metallurgical techniques. Certain clips such as Heifetz, Pivot, and Vari-Angle showed a much greater propensity to stress corrosion failure than the others. As a result of our tests, we recommend that only the clips that show high stress corrosion resistance in in vitro testing should be considered for long term in vivo implantation.

Corrosion↗

Acceleration of capsule formation around silicone implants by infection in a guinea pig model.

Fibrous capsules surrounding silicone implants were investigated in a new guinea pig model to delineate some of the factors leading to capsular contracture. Both the implant surfaces and tissue capsules were examined by light and scanning transmission electron microscopy (STEM + SEM) with x-ray energy spectroscopy (XES). The capsular tissues were qualitatively similar to those recovered clinically, showing dense parallel collagen deposits, fibroblasts, myofibroblasts, macrophages, and foreign-body giant cells. Silicone was positively identified within intercellular vacuoles and the rough endoplasmic reticulum of macrophages by XES. Tissue recovered from the capsules that surrounded implants that were contaminated with S. aureus presented a qualitatively similar histologic spectrum. The contaminated specimens did show an accelerated response. SEM showed a cellular invasion of the silicone envelopes. We conclude that the model accurately simulates the clinical situation and suggests that immune mechanisms may play a key role in capsular contracture.

Animals↗

Intracranial clips: an examination of the devices used for aneurysm surgery.

A properly functioning aneurysm clip is of paramount importance in the surgical ablation of intracranial aneurysms. An appropriate match between the closing force needed to ablate an aneurysm and the force exerted by the aneurysm clip must be achieved. In this study, the opening and closing forces exerted by several commonly used aneurysm clips were tested. There was a strong correlation between the classification of the clips, according to the Dujovny system, and their mechanical behavior. Minimal variability among different lots of the same clip type and minimal hysteresis were generally exhibited by the alpha mobile fulcrum class clips, such as the Yasargil, Vari-Angle McFadden, Scoville, and Mayfield clips. Significant exceptions included several types of Mayfield and Drake clips. In contrast, pivot class clips generally showed twice as much lot variability among different lots, as well as a significant amount of hysteresis. Clips of this class include the Vari-Angle, Heifetz, and Pivot clips. Because significant variations in clip force exist and because several other factors can compound these differences, it is suggested that aneurysm clips be individually tested for their closing forces before they are used in an operation.

Humans↗

Mechanical and metallurgical properties of vascular clips designed for temporary use.

Six commonly used surgical clips (Kleinert-Kutz, Pivot, Vari-Angle, Yasargil, Biemer, and Acland) that have been designated by their manufacturer as being suitable for temporary occlusion were studied in terms of their mechanical and metallurgical properties. Opening and closing forces were determined for all of the clips. Significant variations in clip force with respect to blade gap were found in all of the pivot-type clips. Metallurgical testing showed that all six clips exhibited properties that contraindicate long-term implantation.

Corrosion↗

Biocompatibility of silicone implants.

Polydimethylsiloxane has been considered immunologically inert, and previous work seems to have established that the production of circulating antibodies does not occur in response to its implantation. We have investigated the possibility of a cellular immune response to implanted silicone. We have observed histologically that the cellular response to polydimethylsiloxane in sensitized guinea pigs is consistent with a cellular immune reaction. Further studies with EM and XES have demonstrated intracellular silicon in the Golgi apparatus, rough endoplasmic reticulum, and at both ends of cytoplasmic bridges between macrophages and lymphocytes. All of these findings fit with the hypothesis that the cells are processing a silicon-containing complex as an antigen. Finally, macrophage migration inhibition studies have shown evidence of a cellular immune phenomenon. Further studies are planned to characterize the nature of the sensitizing complex and to attempt to confirm the migration inhibition studies in vivo.

Animals↗

Analysis of the surface morphology of recovered silicone mammary prostheses.

The surfaces of 11 silicone mammary prostheses were examined by scanning electron microscopy (SEM) and energy dispersive x-ray spectrography (XES). Three of the prostheses were recovered from clinical cases and eight from guinea pig models. The surfaces of the prostheses were covered with an investing film and by 2- to 5-microns spheres aggregated in 40- to 200-microns-wide patches. The embeddedded spheres were blebbed and surrounded by pits. X-ray energy spectrography showed high levels of sodium, chlorine, and potassium. It is concluded from this evidence that phagocytic cells embed themselves within the envelope, perhaps by an immune-mediated phagocytic process.

Animals↗

Angiographic demonstration of postoperative cortical artery stenosis induced by Biemer temporary clips.

Five patients underwent cerebral revascularization operations during which Biemer clips were used to occlude temporarily a branch of the middle cerebral artery. Postoperative angiograms revealed stenoses at the sites where the clips were applied. Excessive clip closing forces may have caused significant endothelial injury, which then produced the stenoses. Clips that exert a minimal occlusive force would avoid this response and are recommended for temporary occlusion of cortical arteries.

Adult↗

Failure of a Heifetz aneurysm clip.

A 16-year-old girl died from an acute subarachnoid hemorrhage following the fracture of a blade of a Heifetz aneurysm clip. The clip was manufactured from 17-7PH steel, which on metallurgical testing was found to be highly sensitive to intergranular corrosion. The fracture mechanism was stress corrosion, brought on by the combination of a stress load, an electrolytic environment, and a susceptible steel.

Adolescent↗

Reduced vascular trauma after temporary occlusion with modified Biemer and Yasargil clips.

The Biemer and Yasargil microvascular clips that are currently available commercially can traumatize the vascular endothelium. In an effort to resolve this problem, both clip designs were structurally modified by changing the length, width, and surface of the clip blades in various ways. Also, the occlusion force of the clips was set at either 15 or 30 g. Studies on the carotid arteries of 140 rats showed a correlation between the structural and mechanical alterations of the clips and the degree of endothelial trauma produced by temporary occlusion for 1 hour. The most favorable results were obtained with a Biemer clip that had long, wide, smooth blades and an occluding force of 15 g.

Animals↗

A variable-force microvascular clip.

An easily adjustable, variable-force microvascular clip is described. This clip is superior to other clips currently available because it is essentially three clips in one. Its advantages are that it can reduce the amount of table area needed to store clips during surgery and it offers the surgeon greater flexibility in the selection of clips with different closing forces. This clip was developed by modifying the segments of a basic alpha-type clip, designing a computer program to analyze the force contributions made by each segment of the clip, constructing several clips with different blade lengths, and testing them in a microvascular surgery research program. The clip we have developed is easy to use, reduces clutter on the instrument table, aand offers the surgeon a wide choice of closing forces with a minimal number of clips.

Humans↗

Metallurgical evaluation of the compatibility of surgical clips with their appliers.

Five aneurysm clips and their respective appliers (Heifetz, Vari-Angle, McFadden, Scoville, and Yasargil) were tested for the production of small metal shards that could provoke a foreign body reaction or increase the risk of a stress-corrosion failure. Pivot and Vari-Angle-McFadden clips produced numerous large shards, the Scoville clip produced a few fine shards, and the Yasargil and Heifetz clips produced none. Metal shard production due to cold metal transfer is attributed to the abrasive mechanical interaction between clips and appliers made from metals with different degrees of hardness.

Aneurysm↗

Scanning electron microscopic and X-ray energy spectrographic analysis of the inner surface of human atherosclerotic carotid plaques.

Forty specimens of carotid plaque from 30 patients who had suffered cerebrovascular accidents and 10 asymptomatic control patients were examined by light and scanning electron microscopy (SEM). The SEM revealed the finely detailed surface characteristics of the plaque including cell deposits, endothelial damage, and exposed subendothelium. X-ray energy spectrography scans revealed that 7 specimens had calcium deposits. Presurgical treatment with dipyridamole and aspirin inhibited or substantially reduced thrombus formation.

Aspirin↗

"Nautral" canine model of infantile hydrocephalus.

A review of the current knowledge about a variety of aspects of infantile hydrocephalus suggests that one cause of this problem is subarachnoid blood from birth trauma interacting with an anatomically immature cerebrospinal fluid-absorbing system. To test this hypothesis and to create a model that could be used to devise various methods of prophylaxis and therapy, we injected autologous blood into the cisterna magna of puppies, followed the size of their ventricles with serial computerized tomographic (CT) scans, and evaluated the scanning electron microscopic (SEM) appearance of the ventricular walls and subarachnoid space. By injecting blood until the dogs were just at the point of apnea, we obtained a 46% incidence of hydrocephalus with a 13% mortality rate. The amount of blood accepted was variable. The size of the ventricles was followed with CT scans; the time course of hydrocephalus was variable, and in some cases it resolved. SEM revealed significant regional, free cell aggregation and calcification, especially over the base of the brain, which was most marked in the animals that developed hydrocephalus. This suggests a more marked reaction or a lesser ability to clear the subarachnoid blood in those animals, although the reason for this is not clear. Possible uses for this model are suggested.

Absorption↗

Aneurysm clip testing meter.

The force that a vascular clip exerts when it is applied to a vessel is a critical factor in its safe and proper use. Permanent occlusion effected in the obliteration of aneurysms requires clip closure forces great enough to resist sliping, yet low enough not to damage the neck of the aneurysm. The authors consider the need for a device that can readily measure clip closing forces just prior to pre-surgical sterilization (immediately prior to implantation), and an instrument that meets that need is presented.