Eradication of posterior capsule opacification: documentation of a marked decrease in Nd:YAG laser posterior capsulotomy rates noted in an analysis of 5416 pseudophakic human eyes obtained postmortem.

OBJECTIVE: (1) To report the neodymium:yttrium-aluminum-garnet (Nd:YAG) laser posterior capsulotomy rate (%) of eight rigid and foldable intraocular lens (IOL) designs in a series of 5416 pseudophakic human eyes obtained postmortem, accessioned in our center between January 1988 and January 2000. (2) To identify factors that are instrumental in reducing the incidence of posterior capsule opacification, (PCO, secondary cataract) and hence the need for Nd:YAG laser posterior capsulotomy. DESIGN: Comparative autopsy tissue analysis. PARTICIPANTS: A total of 5416 globes with posterior chamber intraocular lens (PC-IOLs) obtained postmortem received from Lions Eye Banks between 1988 and 2000. METHODS: Miyake-Apple posterior photographic technique. Special reference was given to the presence or absence of Nd:YAG laser posterior capsulotomy orifice on the posterior capsule of each eye. MAIN OUTCOME MEASURES: The Nd:YAG laser posterior capsulotomy rate (%) as of January 2000 was documented. In addition, the Nd:YAG laser posterior capsulotomy rate for each lens was plotted on a monthly basis for the same period, creating a computerized trend or "timeline" for each IOL style. RESULTS: Relatively high Nd:YAG laser posterior capsulotomy rates ranging from 20.3% to 33.4% were noted with four relatively older designs (high incidence of implantation between 1988 and the early 1990s). Four modern foldable IOLs manufactured from silicone and acrylic materials had significantly lower Nd:YAG laser posterior capsulotomy rates ranging from 0.9% (Alcon Acrysof) to 17.1%. The difference in Nd:YAG rates among the eight IOL designs was found to be significant (P < 0.0001, chi-square test). Comparing foldable versus rigid designs, the foldable IOLs were associated with a much lower Nd:YAG laser posterior capsulotomy rate (14.1% vs. 31.1%). CONCLUSIONS: By use of the six factors regarding surgical technique and IOL choice described in this article, we strongly believe that the overall incidence of PCO and hence the incidence of Nd:YAG laser posterior capsulotomy is now rapidly decreasing from rates as high as 50% in the 1980s to early 1990s. Surgical tools and IOLs are now available to bring these rates down to single digits. Careful application and use of these tools by surgeons can genuinely lead in the direction of virtual eradication of secondary cataract, the second most common cause of visual loss worldwide.

Cataract↗

Anterior capsule opacification: a histopathological study comparing different IOL styles.

OBJECTIVE: To compare the degree of anterior capsule opacification (ACO) in human eyes obtained post-mortem containing various rigid and foldable posterior chamber intraocular lens (PC-IOL) designs. DESIGN: Comparative autopsy tissue study with clinicopathologic correlations. MATERIALS: Four hundred sixty human globes containing the following PC-IOL styles were analyzed: (1) one-piece polymethylmethacrylate (PMMA) optic-PMMA haptic (n = 50), (2) one-piece silicone-plate IOL, large hole (n = 40), (3) one-piece silicone-plate IOL, small hole (n = 67), (4) three-piece PMMA optic-PMMA/Prolene haptic (n = 51), (5) three-piece acrylic optic-PMMA haptic (n = 96), (6) three-piece silicone optic-PMMA haptic (n = 24), (7) three-piece silicone optic-polyimide haptic (n = 40), and (8) three-piece silicone optic-prolene haptic (n = 92). TESTING: The globes were sectioned in the equatorial plane for gross examination and then processed through paraffin; sectioned, and stained with hematoxylin-eosin, periodic acid-Schiff, and Masson's trichrome stains; and examined by light microscopy. MAIN OUTCOME MEASURES: Anterior capsule opacification was scored in each eye by grading the histologic sections from 0 to III, according to the amount (thickness) of proliferative tissue and cells measured in sagittal sections on the inner surface of the anterior capsule at the capsulorhexis margin. RESULTS: The difference among the eight groups was significant (P < 0.0001). Mean ACO scores were highest with the large and small hole one-piece silicone-plate lenses (1.77 +/- 0.86 and 1.28 +/- 0.77, respectively). The lowest mean score was observed in the group of three-piece acrylic optic-PMMA haptics lenses (0.51 +/- 0.52). CONCLUSIONS: Our results confirm previous clinical observations that the rate of ACO is relatively high with plate-haptic silicone IOLs. The lowest rate was noted with the three-piece acrylic optic-PMMA haptic IOL. The IOL design and IOL material are significant factors in the development of ACO.

Cataract↗

Opacification of piggyback IOLs associated with an amorphous material attached to interlenticular surfaces.

PURPOSE: To report the pathological and ultrastructural features and interval surgical management of an atypical case of opacification between piggyback intraocular lenses (IOLs). SETTING: Center for Research on Ocular Therapeutics and Biodevices, Storm Eye Institute, Medical University of South Carolina, Charleston, South Carolina, and Nature Coast EyeCare Institute and Surgery Center, Perry, Florida, USA. METHODS: Opacification between 2 acrylic piggyback lenses was observed 16 months after implantation, with decreased best corrected visual acuity and a hyperopic shift. Elschnig pearls were observed in the peripheral interface between the lenses, and the central interface was occupied by an amorphous material. The pearls were surgically aspirated, but attempts to remove the central material were unsuccessful. The lenses were explanted and sent to the laboratory. Staining with hematoxylin and eosin (H&E), examination under a light microscope, and scanning electron microscopic analysis were performed. RESULTS: The surfaces of the anterior IOL were relatively clear. The amorphous material, mostly attached to the center of the anterior surface of the posterior IOL, was homogeneously stained with H&E. No cell nucleus was observed in this region. Scanning electron microscopy showed that the IOL edge presented a smooth, regular surface relatively free of deposits. The most central region was covered by an irregular layer of an amorphous compact material with some cracks, fissures, or both on its surface. CONCLUSIONS: Although the exact composition of the material between the lenses could not be established, hypotheses were advanced to understand the pathological mechanism associated with this condition. This case is different from those in previous reports of opacification composed of cortex and cells between piggyback IOLs.

Acrylates↗

Surgical prevention of posterior capsule opacification. Part 3: Intraocular lens optic barrier effect as a second line of defense.

PURPOSE: To emphasize an important aspect of preventing posterior capsule opacification (PCO), the barrier effect established by the optic of a posterior chamber intraocular lens (PC IOL), and present a new classification regarding capsular bag status after extra-capsular cataract extraction, including phacoemulsification. SETTING: Center for Research on Ocular Therapeutics and Biodevices, Storm Eye Institute, Medical University of South Carolina, Charleston, South Carolina, USA. METHODS: This analysis included 150 consecutive eyes obtained postmortem with United States-manufactured PC IOLs including (1) poly(methyl methacrylate), (2) silicone, and (3) hydrophobic acrylic designs that were accessioned in the Center from September 1995 to January 1, 1998. Gross photographs from behind (Miyake-Apple views) were taken and serial histologic sections prepared. RESULTS: Microscopic analysis of the 150 eyes showed that the morphologic appearance of the capsular bag could be grouped into 2 categories: (1) those with little or no evidence of retained cortical material and cells, and (2) those with retained cortical material and cells in which a Soemmering's ring formed. With the latter, when a distinct barricade to cellular migration created by the IOL optic was noted, 2 discrete configurations occurred, depending on the different geometries of the optic components. With a classic biconvex optic with a curved and tapered edge, in many instances some ingrowth of cells proceeded posteriorly around the edge of the IOL optic in the direction of the central axis. With a lens optic that had a squared, truncated, and relatively thick edge, there was often abrupt termination of cells at the peripheral edge of the optic. The posterior capsule subtending the entire optic zone was therefore relatively or totally cell free. CONCLUSIONS: The barrier effect of the IOL optic appears to be of critical importance in retarding ingrowth of cells, functioning as a second line of defense when cortical cleanup is incomplete. Analysis of PC IOLs obtained postmortem showed that a square, truncated optic edge seemed to provide the maximum impediment to cell growth behind the IOL optic.

Acrylic Resins↗

Surgical prevention of posterior capsule opacification. Part 1: Progress in eliminating this complication of cataract surgery.

PURPOSE: To evaluate over almost 2 decades the success of a component of cataract surgery that represents a critical step in reducing the incidence of posterior capsule opacification (PCO); namely, the efficacy of cortical cleanup. SETTING: Center for Research on Ocular Therapeutics and Biodevices, Storm Eye Institute, Department of Ophthalmology, Medical University of South Carolina, Charleston, South Carolina, USA. METHODS: Accessioned from the early 1980s to 1997, 3320 eyes obtained postmortem with posterior chamber intraocular lenses were analyzed with respect to formation of a postoperative Soemmering's ring. This anatomic lesion, the precursor of clinical PCO, represents an important and measurable indication of the quality of cortical cleanup. Its formation was documented using Miyake-Apple posterior photographic analysis. RESULTS: The quality and thoroughness of cortical cleanup and overall effectiveness in eliminating retained and/or regenerating cortical cells, as measured by scoring of Soemmering's rings, showed virtually no net change since the early 1980s. The intensity of Soemmering's ring was higher in the most recent specimens than in those in the early 1980s. CONCLUSION: The results indicate that renewed attention to cortical cleanup in cataract surgery is warranted for significant reduction in incidence or the elimination of PCO. More attention to the hydrodissection (cortical cleaving hydrodissection) step of the procedure is likely a practical, immediately implementable, and inexpensive remedy.

Cadaver↗

Surgical prevention of posterior capsule opacification. Part 2: Enhancement of cortical cleanup by focusing on hydrodissection.

PURPOSE: To experimentally analyze the role and efficacy of hydrodissection in achieving maximal cortical cleanup. SETTING: Center for Research on Ocular Therapeutics and Biodevices, Storm Eye Institute, Department of Ophthalmology, Medical University of South Carolina, Charleston, South Carolina, USA. METHODS: Phacoemulsification and irrigation/aspiration were performed in 10 pairs of human eyes (20 eyes) obtained postmortem. Ten eyes had previous hydrodissection and 10 eyes, no hydrodissection. The time (seconds) required for complete lens substance removal in each procedure was measured. In addition, a qualitative evaluation of difficulty of surgery was noted. RESULTS: Phacoemulsification required 28.6% less time in eyes with previous hydrodissection than in those without. Irrigation/aspiration time was reduced by 50.9% when hydrodissection was performed. The total time of each procedure required for complete evacuation of the capsular bag was reduced by an average of 37.7% in eyes with hydrodissection. Furthermore, qualitatively the procedure was far easier, less stressful, and caused less posterior capsule stress or rupture when copious hydrodissection was performed. CONCLUSIONS: Hydrodissection enhances the general safety and efficiency of cortical cleanup, especially at 12 o'clock. Hydrodissection is the best available, practical, immediately implementable, and inexpensive means to help remove equatorial E-cells and thus alleviate the incidence of posterior capsule opacification.

Aged↗

Creating cataracts of varying hardness to practice extracapsular cataract extraction and phacoemulsification(2).

We describe a method to harden the nucleus of crystalline lenses in postmortem human globes by intralenticular injection of a mixture of paraformaldehyde and glutaraldehyde (Karnovsky's solution). Evaluation of the nuclear manipulations, using the Miyake-Apple posterior video/photographic technique, shows that the method provides suitable specimens for practicing extracapsular cataract extraction and phacoemulsification.

Cataract↗

Interlenticular opacification: clinicopathological correlation of a complication of posterior chamber piggyback intraocular lenses.

PURPOSE: To present a clinicopathological correlation of 2 pairs of piggyback posterior chamber intraocular lenses (PC IOLs) explanted because of opacification between the lens optics. SETTING: Gayton Health Center, Eyesight Associates of Middle Georgia, Warner Robins, Georgia, and Center for Research on Ocular Therapeutics and Biodevices, Storm Eye Institute, Medical University of South Carolina, Charleston, South Carolina, USA. METHODS: Two pairs of piggyback AcrySof lenses were explanted from 2 patients with significant visual loss related to opacification between the optics. They were submitted for pathological analysis. Gross and histopathological examinations were performed, and photomicroscopy was used to document the results. RESULTS: Gross examination showed accumulation of a membrane-like white material between the lenses. Histopathological examination revealed that the tissue consisted of retained/proliferative lens epithelial cells (bladder cells or pearls) mixed with lens cortical material. CONCLUSION: Piggyback PC IOLs were explanted in 2 cases because of a newly described complication, interlenticular opacification. Three surgical means may help prevent this complication: meticulous cortical cleanup, especially in the equatorial region; creation of a relatively large continuous curvilinear capsulorhexis to sequester retained cells peripheral to the IOL optic within the equatorial fornix; insertion of the posterior IOL in the capsular bag and the anterior IOL in the ciliary sulcus to isolate retained cells from the interlenticular space.

Acrylic Resins↗

Update on fixation of rigid and foldable posterior chamber intraocular lenses. Part I: Elimination of fixation-induced decentration to achieve precise optical correction and visual rehabilitation.

PURPOSE: Realizing that precise posterior chamber intraocular lens (PC-IOL) centration is needed to consistently achieve good optical results and visual rehabilitation after modern cataract surgery, the authors assessed the status and success rate of lens haptic fixation and its correlation with lens optic centration-decentration in a large series of eyes with PC-IOLs obtained postmortem. DESIGN: Prospective analysis of pseudophakic eyes obtained postmortem. PARTICIPANTS: A total of 3493 human eyes obtained postmortem, the largest database of such specimens available to date. METHODS: Miyake-Apple posterior photographic technique. MAIN OUTCOME MEASURES: The amount of decentration of rigid PC-IOLs and small-incision foldable PC-IOLs was analyzed, and the results were correlated with the type of fixation that had been achieved in each case. RESULTS: Determination of mean values revealed that capsular bag fixation was achieved in 52.05% of eyes, asymmetric bag-sulcus fixation in 34.21 % of eyes, and sulcus-sulcus fixation in 13.74% of eyes. Asymmetrically fixed lenses consistently showed significant decentration. During the past 5 years (1993-98), the overall rate of in-the-bag fixation increased to 59.2%; by 1998 it had increased to 64%. Most significantly, success in achieving bag-bag fixation of foldable IOLs implanted in association with modern capsular surgery with small incisions has surged to 90% over the past 4 years. CONCLUSIONS: The authors found a direct correlation of decentration in eyes with asymmetric fixation, and the results underscore the need for careful in-the-bag haptic placement. Although few surgeons today would dispute the goal to implanting haptics in the capsular bag, these findings show that the overall success rate over the years has, while improving, remained surprisingly low. The overall success rate of about 60% seen with all lens designs is probably as good as can be expected with classic large-incision extracapsular techniques. However, and most important and encouraging, the success rate of haptic fixation in cases with foldable lenses has improved dramatically during the past 4 years (up to the 90% range). This coincides with the present emphasis on modern capsular surgery and small-incision techniques used to insert these lenses.

Acrylic Resins↗

Update on fixation of rigid and foldable posterior chamber intraocular lenses. Part II: Choosing the correct haptic fixation and intraocular lens design to help eradicate posterior capsule opacification.

PURPOSE: Posterior capsule opacification (PCO) is one of the most common complications of cataract surgery, and there is an urgent need to reduce its incidence. Its main treatment, neodymium:YAG (Nd:YAG) laser posterior capsulotomy, may be associated with significant morbidity and is expensive. In this study, the authors ascertain the effect of posterior chamber intraocular lens (PC-IOL) fixation and various IOL designs (rigid polymethyl methacrylate [PMMA] optic and small-incision foldable designs) on the pathogenesis of PCO. DESIGN: Prospective analysis of pseudophakic eyes obtained postmortem. PARTICIPANTS: A total of 3493 eyes with PC-IOLs obtained postmortem and received between 1984 and 1998. METHODS: Miyake-Apple posterior photographic technique. MAIN OUTCOME MEASURES: Peripheral Soemmering's ring, central PCO in the visual axis, and eyes that have had Nd:YAG laser posterior capsulotomy were analyzed and scored, and these findings were correlated with the type of fixation and IOL design. RESULTS: The degree of formation of peripheral Soemmering's ring was not influenced by IOL fixation. The acrylic design revealed relatively low Soemmering's ring scores. In contrast, central PCO and Nd:YAG laser capsulotomy scores were consistently influenced by fixation. The scores were significantly lower in eyes with in-the-bag fixation. The bag-bag fixated acrylic-PMMA design and the three modern silicone IOL designs analyzed in this study had less central PCO and lower posterior capsulotomy scores than did the PMMA IOLs. CONCLUSIONS: The formation of peripheral PCO (the Soemmering's ring), the precursor of clinically significant, vision-threatening PCO, is not significantly influenced by the haptic fixation pattern. It is much more dependent on the quality and thoroughness of surgical cortical cleanup. Reduction of Soemmering's ring is an important goal because the retained regenerative cortical cells within this lesion are the cells of origin of PCO. The acrylic IOL design was associated with a lessor amount of peripheral Soemmering's ring as compared with all other designs. In sharp contrast to peripheral PCO, fixation of the IOL was a highly significant factor affecting the formation and quantity of central PCO-the clinically significant opacity behind the IOL optic, measured in this series either by scoring an intact retro-optical membrane or by documenting the presence of a Nd:YAG laser posterior capsulotomy orifice. The quantity of central PCO was consistently much lower in eyes with in-the-bag fixated IOLs compared with lenses with one or both haptics out of the bag. This is best explained by the fact that secure in-the-bag fixation positions the IOL optic in the best possible position to create a barrier effect. The lowest PCO rates were generally noted with the acrylic and modern silicone IOL designs.

Acrylic Resins↗

Pathological examination of an explanted phakic posterior chamber intraocular lens.

PURPOSE: To report the clinicopathological correlation of an explanted phakic posterior chamber intraocular lens (PPC IOL) and to study the conformation of this lens implanted into human eyes obtained postmortem. SETTING: Center for Research on Ocular Therapeutics and Biodevices, Department of Ophthalmology, Storm Eye Institute, Medical University of South Carolina, Charleston, South Carolina, USA. METHODS: Three silicone PPC IOLs were explanted. One lens, explanted from a 38-year-old woman with complicated cataract, was submitted for pathological analysis. In addition to gross and scanning electron microscopy (SEM), the lens was studied after experimental implantation in human eyes obtained postmortem using frontal, posterior, and side-view techniques. RESULTS: Although gross and SEM showed that the IOL was well polished, examination of the lens in human cadaver eyes showed it was oversized and poorly fixated. It was relatively bulky in its anterior-posterior dimension and revealed evidence of significant contact with the iris and crystalline lens. CONCLUSIONS: This study illustrates many pitfalls to be avoided in the design of a plate PPC IOL. The lens in this report was too large, and instead of ciliary sulcus fixation, it showed poor fixation through the zonules onto the posterior face of the pars plicata. Present and future PPC IOLs should be submitted for similar preclinical studies to clarify the type and site of fixation.

Adult↗