Modelling tissue heating with ferromagnetic seeds.

Interstitial hyperthermia using ferromagnetic seeds demands accurate treatment planning: the seed characteristics and implant geometry must be determined prior to the treatment. A new, finite difference based, seed modelling method is presented. The seed, together with all its surrounding (non-tissue) layers is described as one unit, independent of the tissue grid. The calculation of the seed-tissue interaction is based on the local seed temperature and several tissue temperature samples in the direct vicinity. All the layers between the seed and the surrounding tissue are taken into account in this interaction calculation. The presented implementation describes the analytical solution of the modelled steady-state configurations very accurately. The separation between tissue and seed allows easy assessment of the resulting seed temperature profile which is essential to the optimization of the seed characteristics in treatment planning. The thermal effect due to blood flow in the modelled tissue volume surrounding the seed can be accounted for by inclusion of a heat sink term as well as by inclusion of realistic discrete vasculature.

Algorithms↗

Dose uniformity of ferromagnetic seed implants in tissue with discrete vasculature: a numerical study on the impact of seed characteristics and implantation techniques.

The results from simulations with a new three-dimensional treatment planning system for interstitial hyperthermia with ferromagnetic seeds are presented in this study. The thermal model incorporates discrete vessel structures as well as a heat sink and enhanced thermal conductivity. Both the discrete vessels and the ferroseeds are described parametrically in separate calculation spaces. This parametric description has the advantage of an arbitrary orientation of the structures within the tissue grid, easy manipulation of the structures and independence from the resolution of the tissue voxels (tissue calculation space). The power absorption of the self-regulating seeds is according to empirical data. The thermal effects of an unlimited number of thin layers surrounding the seed (coatings, catheters) can be modelled. The initial calculations have been performed for an array of 12 identical ferromagnetic seeds in a tissue volume with a computer generated artificial vessel network spanning four vessel generations in both the arterial and venous tree. The heterogeneously distributed large isolated vessels impair the temperature distribution significantly, indicating the limited accuracy of continuum models. Simulations with different types of ferromagnetic seeds have confirmed that the efforts of previous studies to optimize the self-regulating temperature control and the implantation techniques of the ferroseeds will improve the homogeneity of the temperature distribution in the target volume. Multifilament seeds implanted in brachytherapy needles and tubular seeds appear to be the most favourable configurations. The division of long seeds into shorter segments with the appropriate Curie temperature will further improve the homogeneity of the temperature distribution without increasing the average temperature in the volume of interest. Given the proper thermal tissue data, the model presented in this study will prove to be a useful tool in making choices for the implant geometry, seed spacing and Curie temperature.

Biophysical Phenomena↗

Power absorption and temperature control of multi-filament palladium-nickel thermoseeds for interstitial hyperthermia.

In interstitial hyperthermia using ferromagnetic seeds, multi-filament seeds have gained interest because of a more effective power absorption than solid seeds. Palladium-nickel (PdNi) seeds composed of filaments with diameters in the range from 0.1 to 1.0 mm (maximally 90 filaments) have been investigated to find the conditions for optimal power absorption and temperature control. Magnetic and calorimetric experiments have shown that a decreasing filament radius results in a more effective power absorption. The power absorption approaches a common asymptote for high field intensities at all filament diameters. This asymptotic behaviour can be understood as a consequence of the approach of saturation magnetization of PdNi. The sharpness of the transition at the Curie temperature, which is a measure for the quality of temperature control, improves as the magnetic field strength increases, but it is limited by the asymptote of the power absorption. When the asymptote has been reached the quality of temperature regulation of a seed can only be improved by increasing the amount of PdNi, e.g. by increasing the number of filaments. Calculations of the power absorption, using the generally applied theory based on a linear relation between the magnetization of PdNi and the magnetic field strength, do not correspond quantitatively with experimental results for seeds having an induction number smaller than the 'optimal value' of 2.5. For these seeds the measured heat production is larger than the calculated one.

Calorimetry↗

The development of PdNi thermoseeds for interstitial hyperthermia.

Magnetic induction heating of thermoseed implants can be used to produce highly localized hyperthermia in deep-seated tumors. Automatic temperature control throughout the tumor can be achieved by the self-regulating character of ferromagnetic seeds, which corrects for local variations in heat loss due to blood perfusion. An increased sharpness of the ferromagnetic transition at the Curie temperature, Tc, improves the performance of self-regulating control. This was realized for palladium-nickel alloys by a "cold working" procedure preceded and followed by annealing. Palladium-nickel seeds with a predetermined Tc were produced, showing a sharp decrease at Tc of the magnetic susceptibility and the heat production.

Humans↗

The effect of catheters and coatings on the performance of palladium-nickel thermoseeds: evaluation and design of implantation techniques.

In the development of materials for self-regulating thermoseeds much effort is put in improvement of the self-regulating temperature control mechanism of the seeds. The catheters and coatings which are needed to implant the seeds or to guarantee biocompatibility, generally impair the optimized performance of the ferromagnetic seeds. The influence of various coatings on the performance of PdNi seeds has been investigated by means of one-dimensional modelling and calorimetric experiments. Implantation using thin walled catheters is acceptable provided that the catheters are filled with water to assure good thermal coupling. Air layers inside catheters should be avoided as they reduce the sharp gradient of the heat production at the Curie temperature significantly. An alternative for the application of catheters is to insert the seeds into metallic needles. The effect of shielding by the metal needle can be minimized by driving the seed into its saturated state using a high magnetic field strength. The thermal interaction between the seed and surrounding tissue can also be enhanced by placing PdNi, e.g. tubular, on the outside of the catheter or brachytherapy needle. An additional advantage of this new design is an increase in the heat production and the quality of temperature control due to an increase in the amount of PdNi. For permanent implantation seeds can be coated with an inert metal, ceramics or plastic. The performance of the seeds is not affected by any of the coatings if certain conditions are met. For plastic coatings the thickness of the coating has to be very thin, preferably < or = 20 microns, to avoid thermal isolation.

Biocompatible Materials↗

Three-dimensional temperature control of palladium-nickel thermoseeds: a computer aided and experimental evaluation.

The capability of self-regulating thermoseeds to compensate for nonuniform cooling along their longitudinal axis has been investigated in this study. For this purpose a quasi three-dimensional computer model has been developed. Calculations of the temperature profile in tissue with nonuniform heat loss demonstrated a clear improvement in the longitudinal temperature control of PdNi seeds compared to constant power seeds. Further, two strategies for improved control of nonuniform cooling along the longitudinal axis of ferromagnetic seeds have been investigated: (1) application of a 'normal' undivided seed; and (2) division of a long seed in smaller segments of which each segment is able to respond more directly to local variations in the temperature distribution. Calculations with the quasi three-dimensional model showed that the loose segments do respond more directly to their close proximity. However, the equilibrium temperature of a segment in an area with high local blood flow will be relatively low due the limited heat production of PdNi thermoseeds. In the undivided seeds the high thermal conductivity of PdNi causes some levelling of the longitudinal temperature gradient in the seed. In addition, calorimetric experiments have shown that the heat production of a segmented seed is less effective because of a demagnetizing field. Also, the absence of PdNi between the segments reduces the heat production of the seed.

Animals↗