Iseikonia and accommodation in the fellow eye in monocular IOL implantation.

In unilateral implantation of an intraocular lens (IOL), the main goal, for both distant and near vision, is iseikonia. After monocular IOL implantation, when the normal fellow eye accommodates for different distances, the changed magnification will result in aniseikonia. We have derived formulas for calculating the power of an iseikonizing IOL and of corresponding spectacles for both distant and near vision. Using the formulas, we can predict the degree of aniseikonia produced by accommodation in the normal eye. We suggest that accommodation in the fellow eye be avoided by using reading glasses.

Accommodation, Ocular↗

In vivo coronary angioscopy.

The feasibility of in vivo coronary angioscopy was tested utilizing a 1.8 mm angioscope in vessels where blood had been replaced by optically clear liquids, including a new perfluorocarbon emulsion. After trials in postmortem canine and human coronary arteries, in vivo intraluminal visualization was accomplished in the dog with a catheterization technique and in patients during open heart surgery. The results demonstrate the feasibility and potential clinical usefulness of direct visualization of intravascular anatomy and disease, analogous to endoscopy of other organ systems.

Animals↗

Obtaining a high-quality retinal image with a biconvex intraocular lens.

We investigated the quality of the retinal image provided by a conventional, plano-convex intraocular lens by calculating the point spread function using a wide-angle model of the eye. We then calculated the ratio of the front and back curvatures for a biconvex lens with the conventionally used refractive index (1.491) using Seidel's aberration formulas. The point spread function for this intraocular lens was better than that of the conventional plano-convex lens. THe calculated ratio, which depends on the refractive index and the depth of the anterior chamber, was 3.16 (refractive index, 1.491; anterior chamber, 3.5 mm). As the depth of the anterior chamber decreases, the best ratio increases whereas the total power of the intraocular lens decreases. If an asymmetric intraocular lens is used, the greater curvature must face the cornea.

Anterior Chamber↗

Macular photocoagulation. Optimal wavelength selection.

Relative merits of different laser wavelengths for macular photocoagulation are analyzed in terms of light scattering and light absorption in hemoglobin, melanin, and macular xanthophyll. Disadvantages of inner retinal damage in the macula caused by the blue-green argon photocoagulators in standard clinical use are discussed, along with advantages of argon green and krypton yellow light. Results of treatment of subretinal neovascularization in disciform macular degeneration are presented. A repetitive treatment, close follow-up protocol was employed with the objectives of closing neovasculature and preserving central visual field. In 74 successive patients treated with monochromatic argon green radiation, neovasculature was closed in 93%, central visual field loss was reduced or stabilized in 76%, and visual acuity was improved or stabilized in 70%. Results were analyzed in terms of location and size of the neovascular membrane, and it was found that best results were obtained with neovascularization farther than 600 microns from the foveola and smaller than 2 mm2. In 17 patients with neovascular membranes more than 600 microns from the foveola, neovasculature was closed in 100%, central visual field loss was reduced or stabilized in 94%, and visual acuity was improved or stabilized in 88%. Similar results were obtained with monochromatic krypton yellow radiation.

Humans↗

Retinal vessel photocoagulation: a quantitative comparison of argon and krypton laser.

Retinal arteries of owl monkeys were photocoagulated with single exposures to argon laser green light (514 nm), krypton laser green light (531 nm), and krypton laser yellow light (568 nm). The distribution of damage along the vessel in different retinal layers was characterized geometrically after serial sectioning of the histopathologic material. Krypton laser radiation (green and, to an even greater extent, yellow) produced measurably greater effects on retinal vessels and adjacent structures than on deeper retinal levels; argon laser radiation (green) produced greater effects on the pigment epithelium. Krypton yellow produced the greatest total effect. The location of damage to the retinal arteries and pigment epithelium differed, with the peak of the damage around the arteries being farther from the optic disc than was the peak of the pigment epithelial damage.

Animals↗

Wide-angle noncontact and small-angle contact cameras.

The theoretical wide-angle model of the human eye and the principle of separation of ophthalmoscopic illumination and observation apertures have been used to design two fundamentally different ophthalmoscopic cameras: a 90 degree wide-angle, noncontact camera and a high-resolution macula-disc camera. The chief advantage of the 90 degree noncontact camera is the increase in contrast achieved by eliminating from the retinal image light reflections off corneal and lenticular surfaces. This contrast increase is obtained in addition to a wide-angle field, while the resolution of current fundus cameras is maintained. The chief advantage of the macula-disk camera is that it records the first image of the patient's fundus and thus permits an actual increase in resolution of retinal detail not obtainable with current fundus cameras. This additional detail should greatly facilitate future clinical study and treatment of macular disease.

Contact Lenses↗

Image formation in fundus cameras.

Imaging in a fundus camera depends more on design of the system than on correction of the first fundus image as formed by the ophthalmoscopic lens. We show here that the designer may use the free parameters of the ophthalmoscopic lens (contact or noncontact) to correct the latter for observation and illumination of the fundus. In both contact and noncontact systems the fundus is illuminated by forming a ring of light on the patient's cornea around a central area (the corneal window) reserved for observation. On the first surface of the crystalline lens, the light also forms a ring which must accomodate the total entrance pupil (TEP) of the observation system in its middle and which is limited on the outside by the patient's iris. The restrictions that result from this situation define the entrance pupil of the bundle of rays that image the marginal point of the retina. The limits of this bundle are imposed by the choice of the angular field of view and by the size of the patient's pupil.

Fundus Oculi↗

Clinical trials with the Equator-Plus camera.

The Equator-Plus camera can photograph a 148-degree view of the fundus, measured from the nodal point of the eye. We tested this instrument clinically more than 800 times on over 700 eyes. The large field, which is three to four times that of any conventional fundus camera, is achieved by using a special contact lens as the front element in the camera optics. Photographs taken with the Equator-Plus camera can resolve a lesion as small as one-eighth of a disk diameter, if the contrast is good. Protruding tumors, choroidal or retinal detachment, and extensive disease of the choroid and fundus show up well with this technique. We obtained good pictures in 87% of the clinical tests. We have not observed complications of any kind after photography with the Equator-Plus camera.

Adolescent↗

Effect of the ocular media on the main wavelengths of argon laser emission.

(1) In owl monkeys, where no yellow staining of the media is present, there are clear-cut ophthalmoscopic and histologic differences in the lesions produced by equal amounts of argon green and blue light. (2) The yellow staining of the macula causes differences in lesion intensity and depth when different wavelengths are used for photocoagulation. (3) In humans the influence of age on this difference is apparent.

Adult↗

Variation of energy density in argon laser photocoagulation.

The power density at the retina provided by presently available laser photocoagulators is subject to variation. Three factors contribute to making the retinal image of the laser beam substantially different from the spot size indicated on the selector of the photocoagulator. These are the operator's variable accommodation, the astigmatism introduced by the contact lens, and the depth of focus of the observation system. The assumption that the same setting on the instrument regularly produces the same power density at the retina will lead to errors in comparing lesions. The safest procedure is to start from the threshold in every location.

Argon↗

Equator-plus camera.

A new fundus camera can photograph a field extending anterior to the equator (148 degrees from the nodal point). The problems of reflection from the surfaces of the crystalline lens are minimized by fiber optic illumination from carefully chosen locations on the cornea. The camera may be used with monochromatic light of various wavelengths and is equipped with interference filters.

Fiber Optic Technology↗