PubMed Health⌕ Search

Biomedical subjects

Otto K Harling

Publications and source records attributed to Otto K Harling.

7 recordsLinked to original sources

The MIT user center for neutron capture therapy research.

Neutron capture therapy (NCT) research encompasses a wide range of preclinical and clinical studies needed to develop this promising but complex cancer treatment. Many specialized facilities and capabilities including thermal and epithermal neutron irradiation facilities, boron analysis, specialized mixed-field dosimetry, animal care facilities and protocols, cell culture laboratories, and, for human clinical studies, licenses and review board approvals are required for NCT research. Such infrastructure is essential, but much of it is not readily available within the community. This is especially true for neutron irradiation facilities, which often require significant development and capital investment too expensive to duplicate at each site performing NCT research. To meet this need, the NCT group at the Massachusetts Institute of Technology (MIT) has established a User Center for NCT researchers that is already being accessed successfully by various groups. This paper describes the facilities, capabilities and other resources available at MIT and how the NCT research community can access them.

Animals↗

Microdosimetric intercomparison of BNCT beams at BNL and MIT.

Microdosimetric measurements have been performed at the clinical beam intensities in two epithermal neutron beams, the Brookhaven Medical Research Reactor and the M67 beam at the Massachusetts Institute of Technology Research Reactor, which have been used to treat patients with Boron Neutron Capture Therapy (BNCT). These measurements offer an independent assessment of the dosimetry used at these two facilities, as well as provide information about the radiation quality not obtainable from conventional macrodosimetric techniques. Moreover, they provide a direct measurement of the absorbed dose resulting from the BNC reaction. BNC absorbed doses measured within this study are approximately 15% lower than those estimated using foil activation at both MIT and BNL. Finally, an intercomparison of the characteristics and radiation quality of these two clinical beams is presented. The techniques described here allow an accurate quantitative comparison of the physical absorbed dose as well as a measure of the biological effectiveness of the absorbed dose delivered by different epithermal beams. No statistically significant differences were observed in the predicted RBEs of these two beams. The methodology presented here can help to facilitate the effective sharing of clinical results in an effort to demonstrate the clinical utility of BNCT.

Boron Neutron Capture Therapy↗

Boron neutron capture therapy: cellular targeting of high linear energy transfer radiation.

Boron neutron capture therapy (BNCT) is based on the preferential targeting of tumor cells with (10)B and subsequent activation with thermal neutrons to produce a highly localized radiation. In theory, it is possible to selectively irradiate a tumor and the associated infiltrating tumor cells with large single doses of high-LET radiation while sparing the adjacent normal tissues. The mixture of high- and low-LET dose components created in tissue during neutron irradiation complicates the radiobiology of BNCT. Much of the complexity has been unravelled through a combination of preclinical experimentation and clinical dose escalation experience. Over 350 patients have been treated in a number of different facilities worldwide. The accumulated clinical experience has demonstrated that BNCT can be delivered safely but is still defining the limits of normal brain tolerance. Several independent BNCT clinical protocols have demonstrated that BNCT can produce median survivals in patients with glioblastoma that appear to be equivalent to conventional photon therapy. This review describes the individual components and methodologies required for effect BNCT: the boron delivery agents; the analytical techniques; the neutron beams; the dosimetry and radiation biology measurements; and how these components have been integrated into a series of clinical studies. The single greatest weakness of BNCT at the present time is non-uniform delivery of boron into all tumor cells. Future improvements in BNCT effectiveness will come from improved boron delivery agents, improved boron administration protocols, or through combination of BNCT with other modalities.

Animals↗

Treatment planning and dosimetry for the Harvard-MIT Phase I clinical trial of cranial neutron capture therapy.

PURPOSE: A Phase I trial of cranial neutron capture therapy (NCT) was conducted at Harvard-MIT. The trial was designed to determine maximum tolerated NCT radiation dose to normal brain. METHODS AND MATERIALS: Twenty-two patients with brain tumors were treated by infusion of boronophenylalanine-fructose (BPA-f) followed by exposure to epithermal neutrons. The study began with a prescribed biologically weighted dose of 8.8 RBE (relative biologic effectiveness) Gy, escalated in compounding 10% increments, and ended at 14.2 RBE Gy. BPA-f was infused at a dose 250-350 mg/kg body weight. Treatments were planned using MacNCTPlan and MCNP 4B. Irradiations were delivered as one, two, or three fields in one or two fractions. RESULTS: Peak biologically weighted normal tissue dose ranged from 8.7 to 16.4 RBE Gy. The average dose to brain ranged from 2.7 to 7.4 RBE Gy. Average tumor dose was estimated to range from 14.5 to 43.9 RBE Gy, with a mean of 25.7 RBE Gy. CONCLUSIONS: We have demonstrated that BPA-f-mediated NCT can be precisely planned and delivered in a carefully controlled manner. Subsequent clinical trials of boron neutron capture therapy at Harvard and MIT will be initiated with a new high-intensity, high-quality epithermal neutron beam.

Adult↗

A physical dosimetry intercomparison for BNCT.

An intercomparison of physical dosimetry methods used at the Massachusetts Institute of Technology (MIT) and Brookhaven National Laboratory was completed to enable retrospective analysis of BNCT trials. Measurements were performed under reference conditions pertinent to clinical irradiations at the epithermal neutron beam facility of the Brookhaven Medical Research Reactor (BMRR) using procedures developed at MIT during similar trials. Thermal neutron flux was determined from gold foil activation experiments and good agreement was found between the depth profiles measured in-phantom by the two groups. At a depth of 3.5 cm where the measured flux is greatest, the ratio of the MIT/BMRR measurements is 1.01+/-0.10 if the same reporting procedures are applied. Photon and fast neutron absorbed dose rates were assessed using ionization chambers with separate graphite and A-150 plastic walls. Measurement of the in-phantom photon depth dose component agreed favorably with that previously reported by the BMRR group using thermoluminescent dosimeters. At a depth of 3.5 cm the ratio of the MIT measurements to those made by the BMRR group was 0.89+/-0.12. In-air measurements of the fast neutron and photon absorbed dose rates agreed within the limits of experimental uncertainty. Additional studies were performed in the ellipsoidal water phantom regularly used for beam characterizations at MIT. No significant differences in the thermal neutron flux measured in either a solid PMMA cube or an ellipsoidal shaped water phantom were observed on the central axis of the beam. This study confirms the reproducibility and uniformity of dosimetry measurements performed by the two independent groups undertaking BNCT trials in the USA and provides the physical data necessary to compare BMRR treatment protocols with those applied at Harvard-MIT.

Biophysical Phenomena↗

Fission reactor neutron sources for neutron capture therapy--a critical review.

The status of fission reactor-based neutron beams for neutron capture therapy (NCT) is reviewed critically. Epithermal neutron beams, which are favored for treatment of deep-seated tumors, have been constructed or are under construction at a number of reactors worldwide. Some of the most recently constructed epithermal neutron beams approach the theoretical optimum for beam purity. Of these higher quality beams, at least one is suitable for use in high through-put routine therapy. It is concluded that reactor-based epithermal neutron beams with near optimum characteristics are currently available and more can be constructed at existing reactors. Suitable reactors include relatively low power reactors using the core directly as a source of neutrons or a fission converter if core neutrons are difficult to access. Thermal neutron beams for NCT studies with small animals or for shallow tumor treatments, with near optimum properties have been available at reactors for many years. Additional high quality thermal beams can also be constructed at existing reactors or at new, small reactors. Furthermore, it should be possible to design and construct new low power reactors specifically for NCT, which meet all requirements for routine therapy and which are based on proven and highly safe reactor technology.

Animals↗

A critical examination of the results from the Harvard-MIT NCT program phase I clinical trial of neutron capture therapy for intracranial disease.

A phase I trial was designed to evaluate normal tissue tolerance to neutron capture therapy (NCT); tumor response was also followed as a secondary endpoint. Between July 1996 and May 1999, 24 subjects were entered into a phase I trial evaluating cranial NCT in subjects with primary or metastatic brain tumors. Two subjects were excluded due to a decline in their performance status and 22 subjects were irradiated at the MIT Nuclear Reactor Laboratory. The median age was 56 years (range 24-78). All subjects had a pathologically confirmed diagnosis of either glioblastoma (20) or melanoma (2) and a Karnofsky of 70 or higher. Neutron irradiation was delivered with a 15 cm diameter epithermal beam. Treatment plans varied from 1 to 3 fields depending upon the size and location of the tumor. The 10B carrier, L-p-boronophenylalanine-fructose (BPA-f), was infused through a central venous catheter at doses of 250 mg kg(-1) over 1 h (10 subjects), 300 mg kg(-1) over 1.5 h (two subjects), or 350 mg kg(-1) over 1.5-2 h (10 subjects). The pharmacokinetic profile of 10B in blood was very reproducible and permitted a predictive model to be developed. Cranial NCT can be delivered at doses high enough to exhibit a clinical response with an acceptable level of toxicity. Acute toxicity was primarily associated with increased intracranial pressure; late pulmonary effects were seen in two subjects. Factors such as average brain dose, tumor volume, and skin, mucosa, and lung dose may have a greater impact on tolerance than peak dose alone. Two subjects exhibited a complete radiographic response and 13 of 17 evaluable subjects had a measurable reduction in enhanced tumor volume following NCT.

Adult↗