Pressure infusors: variability in delivered infusion pressure.

A pressure infusor was used to apply external pressures of 300, 250, 200, and 150 mm Hg to 500-ml plastic bags of saline. At constant applied pressure, 50-ml aliquots of saline were periodically withdrawn. The delivered pressures were less than the applied pressures, and the difference increased as the volume of saline within the bag decreased. This inherent pressure difference should be considered when the system is used with a continuous flush device.

Catheterization↗

A multiple-purpose automatic pressure regulator.

The Gairdner Hospital Automatic Pressure Regulator was developed to deliver a constant pre-set gas pressure for the continuous inflation of infusion pump cuffs or pneumatic tourniquets. It has been found to be economical, safe, reliable and easy to use. The device facilitates both continuous flushing of catheters and rapid intravenous infusions especially during emergency situations. It may also be used in other clinical situations where a continuous gas supply at a pre-set constant pressure is required.

Air Pressure↗