PubMed HealthSearch

Biomedical subjects

P Metcalfe

Publications and source records attributed to P Metcalfe.

At least 19 recordsLinked to original sources

Rapid phenotyping of HPA-1a using either diabody-based hemagglutination or recombinant IgG1-based assays.

BACKGROUND: The HPA-1 system is carried on the beta3 integrin. HPA-1a (Zw(a), Pl(A1)) is immunogenic in an HPA-1b homozygote (HPA-1b1b). In pregnancy, 1 of 365 women forms anti-HPA-1a, which causes severe thrombocytopenia in 1 in 1100 neonates. Identification of women at risk of forming anti-HPA-1a and the screening of donors to obtain HPA-1a-negative platelets for therapy need reliable, low-cost, automated assays. STUDY DESIGN AND METHODS: A diabody with dual specificity for HPA-1a x D and an IgG1 anti-HPA-1a have been constructed by the use of the genes encoding the first anti-HPA-1a fragment. With these reagents, two complementary HPA-1a phenotyping assays have been developed. RESULTS: This diabody was used in a simple hemagglutination technique to perform HPA-1a phenotyping on soluble glycoprotein IIb/IIIa from EDTA plasma samples. Over 1000 unselected donors have been correctly HPA-1a-phenotyped by use of the diabody. The human recombinant IgG1 anti-HPA-1a was produced in a rat myeloma cell line and was fluorescein labeled for use in a whole-blood flow cytometric HPA-1a phenotyping assay. This IgG1 anti-HPA-1a shows a clear differential between HPA-1a-positive and HPA-1a-negative platelets at nM antibody concentrations. CONCLUSIONS: The two recombinant reagents described are highly suitable for screening and confirmatory HPA-1a phenotyping. They permit rapid determination of the HPA-1a phenotype and are amenable to automation.

Antigens, CD

Accounting for treatment delays when treating highly proliferative tumours.

This study was undertaken to investigate the possibility of increasing the dose per fraction or increasing the number of fractions to account for treatment delays occurring during radiotherapy treatments for highly proliferative tumours. The linear quadratic model with time was used to determine the difference in biological effective dose (BED) for the original schedule and the schedule including a treatment delay. Tables of extra fractions and extra dose per fraction required to account for a number of possible delays have been determined. It has been shown that for tumours with very short potential doubling times it is best to deliver the extra dose as an increase in dose per fraction rather than an increase in the number of fractions, while for tumours with moderately short potential doubling times (above 7 days) the reverse is true. The equivalent uninterrupted schedules, which would have delivered the same effects to the tumour, have also been determined.

Cell Division

Perturbation of radiotherapy beams by radiographic film: measurements and Monte Carlo simulations.

Radiographic film is an established practical tool used in the measurement of the dose distribution for radiotherapy purposes. The accuracy and reproducibility of film optical density as an indicator of dose has been associated with several factors including photon energy, processing conditions and film plane orientation. Few studies have investigated the factors causing variability in film dosimetry, due to the difficulty of separating the individual contributions. The effect that a sheet of radiographic film in a water phantom has on its response to a 6 MV photon and a cobalt-60 teletherapy beam, when orientated perpendicular and parallel to the beam central axis, is reported. Monte Carlo generated spectra were used to calculate collision kerma (Kcoll) for water and film elements. Measured and calculated results indicate a potential over-response at 25 cm depth of the order of 14 +/- 2.4% and 18 +/- 6.0% respectively for 6 MV photons and 15 +/- 3.4% and 32 +/- 4.5% respectively for a cobalt beam. For film exposed parallel as compared to perpendicular to the central axis of the beam, the calculated results suggest an explanation in terms of the predominantly forward directed secondary electrons for the measured difference in film response at depth. It is proposed that the difference in response of the parallel as compared to perpendicular exposed film be due to the predominantly 'upstream' photon interactions giving rise to energy deposition in film. The simulations indicate that the variation with depth of relative energy imparted in film and water elements correlates with the observed variation in film response with depth.

Energy Transfer

Standard effective doses for proliferative tumours.

This study was undertaken to investigate the treatment schedules used clinically for highly proliferative tumours, particularly with reference to the effects of fraction size, fraction number and treatment duration. The linear quadratic model (with time component) is used here to compare non-standard treatment regimens (e.g. accelerated and hyperfractionated schedules), currently the focus of randomized trials, with each other and some common 'standard regimens'. To ensure easy interpretation of results, two parameters known as proliferative standard effective dose one (PSED1) and proliferative standard effective dose two (PSED2) have been calculated for each regimen. Graphs of PSED1 and PSED2 versus potential doubling time (Tp) have been generated for a range of fractionation regimens which are currently under trial in various randomized studies. From these graphs it can be seen that the highly accelerated schedules (such as CHART) only show advantages for tumours with very short potential doubling times. Calculations for most of the schedules considered showed at least equivalent tumour control expected for the trial schedule compared with the control arm used and these values agree quite well with clinical results. These calculations are in good agreement with clinical results available at present. The greater the PSED1 or PSED2 for the schedule considered the greater the tumour control, which can be expected. However, as has been seen with clinical trials, this higher cell kill also results in higher acute effects which have proved too great for some accelerated schedules to continue.

Cell Division

HPA genotyping by PCR-SSP: report of 4 exercises.

BACKGROUND AND OBJECTIVES: Polymerase chain reaction with sequence specific primers (PCR-SSP) is widely used for the determination of the alleles encoding the human platelet antigens (HPA)-1 to 5. In order to evaluate and improve performance with this technique, four exercises were organised during 1996-1998. MATERIALS AND METHODS: Coded DNA samples were distributed from the National Institute for Biological Standards and Control (NIBSC) as follows: exercise one, 18 samples; two, 12 samples; three, 6 samples, and four, 4 samples. RESULTS: Performance improved over the four exercises following the adoption of a consensus protocol and the re-design of one primer. The percentage of incorrect results in each exercise was as follows: exercise one, 9%; two 3.2%, three, 0.8%, and four, 0.3%. CONCLUSION: The modified PCR-SSP protocol is a reliable method for genotyping HPA-1 to 5 in reference laboratories. DNA-based HPA genotyping has an important role in platelet immunology and further exercises will be included in the bi-annual platelet immunology exercises organised by NIBSC.

Antigens, Human Platelet

Activation during preparation of therapeutic platelets affects deterioration during storage: a comparative flow cytometric study of different production methods.

Three different separation methods, all using centrifugation, are routinely used to prepare therapeutic platelet concentrates from human donor blood. Platelet concentrates derived from platelet-rich plasma (PRP-PC), buffy coat (BC-PC) and apheresis (AP-PC) were investigated at the end of production, and over an 8 d storage period. Change in platelet surface markers were measured by flow cytometry, using fluorescein-conjugated antibodies to fibrinogen, P-selectin (CD62P), GPIIb-IIIa (CD41), GPIb alpha (CD42b) and GPV (CD42d), and fluorescein-conjugated Annexin V was used to measure expression of anionic phospholipid. All concentrates showed some changes during preparation but PRP-PC underwent the greatest changes with significantly higher levels of P-selectin (P<0.001) and bound Annexin V (P=0.001) than AP-PC or BC-PC, and lower levels of GPIb alpha (P=0.002) and GPV (P<0.001). These changes were attributable to component separation rather than venesection. These markers all continued to change on storage with a strong positive correlation between the changes seen during production and those after 5 d storage. PRP-PC continued to show the greatest changes whereas BC-PC showed the least. Fibrinogen was bound to 40-50% of platelets in all preparations and this did not alter significantly on storage whereas total expression of GPIIb-IIIa remained unchanged throughout. There was no evidence that the platelet surface changes were thrombin-mediated and leucocyte depletion of BP-PC by filtration had no effect on the changes. It is proposed that the deterioration of platelet concentrates during storage may be related to activation occurring during preparation. 'Whole blood' flow cytometry using a panel of fluorescein-labelled reagents provides an informative method for evaluating platelet concentrates.

Annexin A5

Interlaboratory variation in the detection of clinically significant alloantibodies against human platelet alloantigens.

This report describes the results of eight workshop exercises which were designed to test the proficiency of laboratories in the detection of antibodies to human platelet antigens (HPA). Detection of the most clinically significant alloantibody, anti-HPA-1a, is adequate. However, despite improvements in consistency of test results between laboratories over the last 3 years, there is still a high probability that clinically significant antibodies against other HPA alloantigens will not be detected.

Antigens, Human Platelet

HPA genotyping by PCR sequence-specific priming (PCR-SSP): a streamlined method for rapid routine investigations.

We describe a streamlined method for the simultaneous identification of alleles of the human platelet antigens (HPA) 1-5. The method employs the polymerase chain reaction with sequence specific primers (PCR-SSP). Although PCR-SSP has been applied to HPA genotyping, all methods previously described have required different reaction mixes and PCR conditions. We have designed a set of sequence-specific primers for HPA 1-5 which react optimally under identical reaction and PCR conditions. Comparative testing with reference samples gave 100% concordance. The advantages of this method include speed; accuracy; smaller sample requirements and no reliance on human typing sera or platelet integrity. The method also has the potential to be applied to amniotic fluid. Simplified DNA techniques will lead to more extensive and proficient platelet antigen typing. This will facilitate accurate laboratory diagnosis of alloimmune thrombocytopenia and the provision of HPA-matched blood products.

Alleles

The validity of using radiographic film for radiotherapy dosimetry.

Radiographic film is routinely used to obtain dosimetric information about therapy treatment beams. One source of inaccuracy is the variability of the chemical processing of the radiographic film. Several processors, film types and two sources (light and x-rays) were used to investigate the reproducibility of film processing for valid film dosimetry. Particular attention is paid to films commonly used in radiation therapy. The results suggest that for this series of film, given the same exposure, the variability in processing may result in an error of +/- 3.3% in optical density, which would lead to an error of +/- 4.4% in indicated dose. This level of inaccuracy is typical for both point and relative dose estimates. These results indicate that film should not be used as a point or relative dosimeter, unless the combination of film type, exposure, processing and reading have been specifically validated.

Equipment Design

Multicentre dosimetry study of mantle treatment in Australia and New Zealand.

PURPOSE: To determine the difference between expected and measured dose for patients prescribed a mantle treatment for Hodgkin's disease and estimate the range of dose at critical sites and between different treatment centres. METHODS AND MATERIALS: Twenty three radiotherapy centres were surveyed with regard to the accuracy of dose delivery to a custom-built upper-torso phantom. Thermoluminescent dosimeters were used to monitor the delivered dose at sites such as mid-plane, spinal cord, neck, axilla and lung. RESULTS: The intended dose to the phantom at each centre was 1 Gy to central axis mid-plane. Of the centres surveyed, the median measured dose to this region was 0.96 Gy with a minimum of 0.92 and a maximum of 1.00 Gy. Median dose to the axilla region was low (0.90 Gy) whereas median dose to the blocked lung and neck region were higher than expected, 0.18 Gy and 1.25 Gy, respectively. The 95% confidence interval on the reported relative dose using the 4000 thermoluminescence dosimetry readings in this study was +/- 1.5% CONCLUSION: In this controlled experiment, using conventional methods to calculate dose, there was a surprising variability in the dose delivered at the central axis mid-plane position. This was traced to lack of uniformity in the use of equivalent squares to calculate the output factor. The measured doses to axilla and lung are explained by photon and electron scattering effects. Centres, where dose compensation was included, had a superior dose homogeneity in the neck. The off-axis dose calculations depend on computer planning software but the magnitude of these differences is secondary to that of central axis mid-point dose differences. Improved consistency of dose calculation techniques between centres would enable more reliable dose response evaluation from multicentre clinical studies of Hodgkin's disease.

Australia

Antenatal screening for fetal alloimmune thrombocytopenia: the results of a pilot study.

Feto-maternal incompatibility for the human platelet antigen HPA-1a is an important cause of severe fetal thrombocytopenia. The incidence is 1 in 1000-2000 pregnancies, which is more common than other conditions for which screening is presently carried out. Antenatal diagnosis and management are now available, but only for subsequent siblings following diagnosis of a previously affected infant. This study describes a pilot prospective screening programme for the antenatal detection of fetomaternal alloimmune thrombocytopenia (FMAIT) due to HPA-1a incompatibility. 3473 women were typed for HPA-1a using a method designed for large-scale typing. 71 women found to be HPA-1a negative were further tested for HLA-DR52a as a risk factor for alloimmunization. All women were monitored for the development of anti-HPA-1a throughout pregnancy and a cord full blood count was taken at delivery. Two affected pregnancies were found and treated: a singleton pregnancy was treated antenatally and a twin pregnancy after delivery. The study showed that screening for FMAIT could be established within the pre-existing antenatal red cell serology programme. It was concluded that screening should be based on platelet typing and offered regardless of parity. Further stratification, combining DR52a typing and HPA-1a antibody screening, although focusing on the group of women at greater risk, may not identify all affected pregnancies. Confirmation of the diagnosis and severity of FMAIT continues to depend on fetal blood sampling during pregnancy or cord blood samples after birth.

Adult

Human platelet antigen-2 and -3 genotyping by PCR-SSP.

Allele-specific PCR using sequence specific primers (PCR-SSP) is a simple and reliable technique to detect point mutations in genes. We have developed a PCR-SSP to enable the detection of a C-T mutation at position 482 of the GPIb gene and a T-G mutation at position 13,962 in exon 26 of the GPIIb gene. These point mutations are at the basis of the HPA alloantigens 2a, 2b and 3a, 3b respectively. One primer of each primer set has a 3' nucleotide complementary to the DNA sequence coding for one allele. PCR product is only produced when the corresponding DNA is present and thus the genotype is determined by the presence or absence of a band in agarose electrophoresis of PCR products. A second set of primers in the same reaction yields a product regardless of the HPA genotype to control the efficiency of the PCR amplification. The HPA-2 and -3 genotypes determined in this way were in strict concordance with those established by conventional genotyping using PCR followed by restriction enzyme digestion (PCR-ASRA). PCR-SSP is a rapid and reliable technique that can be used for the determination of alleles which code for platelet alloantigens.

Antigens, Human Platelet

The Windsor Clinic Alcohol Withdrawal Assessment Scale (WCAWAS): investigation of factors associated with complicated withdrawals.

A new scale for measuring the severity of the alcohol withdrawal syndrome has been developed. The Windsor Clinic Alcohol Withdrawal Assessment Scale (WCAWAS) has more objective criteria than its parent scales and has high inter-rater reliability and concurrent validity. This article describes the development of the WCAWAS and its use in an investigation of which factors were associated with complicated withdrawals in 142 consecutive patients admitted for detoxification. Multivariate analyses showed that drinking behaviour variables and MCV were successful at predicting complicated withdrawals. The WCAWAS was not successful at predicting which patients developed convulsions. Further studies are required to see if the WCAWAS is more successful at predicting which patients will develop visual hallucinations. The WCAWAS is useful as a teaching aid and in following the clinical course of patients and their response to treatment.

Adult

A simplified method for large-scale HPA-1a phenotyping for antenatal screening.

A simplified method for large-scale HPA-1a phenotyping of platelets was developed for use in an antenatal screening programme for fetomaternal alloimmune thrombocytopenia (FMAIT). The test was based on the MAIPA assay, which was modified for antigen-typing with a well-characterized anti-HPA-1a reagent. The resulting assay gave reliable results, was inexpensive and allowed testing of large batches using semiautomated equipment.

Antibodies, Monoclonal

Relative importance of immune and non-immune causes of platelet refractoriness.

In this prospective study, 26 consecutive patients being treated for haematological malignancies receiving standard (i.e. non-leucocyte-depleted) blood components were observed for the development of refractoriness to platelet transfusions. One hundred and sixteen of the 266 (44%) platelet transfusions failed to produce a satisfactory response. In 102/116 (88%), the poor response was in the presence of non-immune factors known to be associated with platelet refractoriness. Non-immune factors were present alone in 78/116 (67%), and in combination with immune factors in a further 24/116 (21%). Immune factors (HLA and platelet-specific antibodies) were present during 29/116 (25%) of unsuccessful platelet transfusions. Statistical analysis confirmed that platelet refractoriness was significantly associated with the presence of non-immune factors. The non-immune factors associated with refractoriness were often multiple, most frequently a combination of fever, infection and antibiotic therapy. This study provides evidence that immune mechanisms were not the predominant cause of platelet refractoriness in the patient population studied. It also suggests that measures for the prevention of HLA alloimmunisation, such as leucocyte depletion, may have a limited impact in reducing the incidence of refractoriness to platelet transfusions.

Acute Disease

Antenatal management of severe feto-maternal alloimmune thrombocytopenia: HLA incompatibility may affect responses to fetal platelet transfusions.

In feto-maternal alloimmune thrombocytopenia (FMAIT), severe hemorrhage, particularly intracranial haemorrhage (ICH), may occur before delivery. Management strategies to prevent ICH in high-risk pregnancies include maternal administration of intravenous Ig with or without steroids and fetal platelet transfusions. This report describes a patient who lost three fetuses with ICH because of FMAIT due to anti-HPA-1a. ICH occurred earlier in successive pregnancies (at 28, 19, and 16 weeks of gestation) despite maternal treatment with intravenous Ig and steroids from 14 weeks of gestation in the third pregnancy. The fourth pregnancy was managed by administering weekly intraperitoneal injections of Ig to the fetus from 12 to 18 weeks of gestation. At 18 weeks, there was no evidence of ICH, but the fetal platelet count was only 12 x 10(9)/L. Serial fetal platelet transfusions were started, but there were poor responses because of immune destruction of the transfused platelets by maternal HLA antibodies. There were improved responses to transfusions prepared from the mother and from HLA-compatible HPA-1a-negative donors. At 35 weeks of gestation, a normal infant was delivered by Caesarean section after 20 platelet transfusions. There was prolonged thrombocytopenia in the baby for 15 weeks after birth, probably due to transfer of HPA-1a antibodies in the transfusions of unwashed maternal platelets. The optimal management of pregnancies likely to be severely affected by FMAIT is still evolving. Intensive management was successful in this case, but a successful outcome cannot be guaranteed in severely affected cases. This is the first time that HLA incompatibility has been found to complicate fetal transfusion therapy.

Adult

Investigation of the tissue equivalence of gels used for NMR dosimetry.

The transition of Fe2+ to Fe3+ in Fricke solution after irradiation results in a change of NMR proton relaxation times in agarose gels which can be used for the dosimetry of ionizing radiation. The main advantage of this system is the possibility of observing dose distributions in 3 dimensions in a medium which is supposed to be tissue equivalent. The aim of the present study was to quantify parameters which determine the tissue equivalence of NMR dosimetry gels. Electron densities and effective atomic numbers were calculated for gels with varying iron, sulphur and agarose concentration. The Hounsfield CT numbers so derived agree well with the CT numbers measured on a clinical CT scanner (effective photon energy 70.7 keV). The Hounsfield CT number of 7.2 +/- 1.5 (n = 9) measured for a 1.5% agarose gel doped with 0.5 mM ammonium ferrous sulphate and 125 mM sulphuric acid compares well with the calculated one of 8 +/- 5. Relaxation times were measured from a series of MR images obtained on a 1.5 T clinical MR scanner. The observed change in 1/T1 of the gel with dose was found to be linear up to 10 Gy (0.084 s-1 Gy-1). No difference in dose response for 10 Gy delivered by four different superficial radiation qualities (HVT = 1.4-7.5 mm Al) could be observed. These findings and the calculated effective atomic number of 7.46 demonstrate the close tissue equivalence of this agarose gel which makes it an ideal tool for the investigation of low energy therapeutic x-rays.

Ferrous Compounds