PubMed Health⌕ Search

Biomedical subjects

Paul A Jursinic

Publications and source records attributed to Paul A Jursinic.

6 recordsLinked to original sources

Initial experience of FDG-PET/CT guided IMRT of head-and-neck carcinoma.

PURPOSE: The purpose of this study is to evaluate the impact of (18)F-fluorodeoxyglucose positron emission tomography (FDG-PET) fused with planning computed tomography (CT) on tumor localization, which guided intensity-modulated radiotherapy (IMRT) of patients with head-and-neck carcinoma. METHODS AND MATERIALS: From October 2002 through April 2005, we performed FDG-PET/CT guided IMRT for 28 patients with head-and-neck carcinoma. Patients were immobilized with face masks that were attached with five fiducial markers. FDG-PET and planning CT scans were performed on the same flattop table in one session and were then fused. Target volumes and critical organs were contoured, and IMRT plans were generated based on the fused images. RESULTS: All 28 patients had abnormal increased uptake in FDG-PET/CT scans. PET/CT resulted in CT-based staging changes in 16 of 28 (57%) patients. PET/CT fusions were successfully performed and were found to be accurate with the use of the two commercial planning systems. Volume analysis revealed that the PET/CT-based gross target volumes (GTVs) were significantly different from those contoured from the CT scans alone in 14 of 16 patients. In addition, 16 of 28 patients who were followed for more than 6 months did not have any evidence of locoregional recurrence in the median time of 17 months. CONCLUSION: Fused images were found to be useful to delineate GTV required in IMRT planning. PET/CT should be considered for both initial staging and treatment planning in patients with head-and-neck carcinoma.

Adult↗

Measurement of head scatter factors of linear accelerators with columnar miniphantoms.

The measurement of linear accelerator head scatter factors or in-air output factors, Sc, with columnar miniphantoms is refined in this work. Columnar miniphantoms are constructed from water equivalent materials: solid water and M3, and materials with higher mass density and atomic number: copper and lead. The change in the value of Sc from a 4-cm X 4-cm to a 40-cm X 40-cm field is different by 22% +/- 3%, 18% +/- 2%, and 10% +/- 3% for 6, 15, and 23 MV x rays, respectively, when measured with water equivalent or lead miniphantoms of 10 gm/cm2 depth. Based on measurements of transmission factors in solid-water miniphantoms of different depths, it is demonstrated that the beam energy spectra decreases in energy with increased field size. These changes in beam energy spectra alter the transmission and scatter of radiation and buildup of the dose in the miniphantom even if the miniphantom is made of water-equivalent material. These changes underlie the alteration in Sc when measured by miniphantoms fabricated from materials of different atomic number. It is shown that miniphantoms designed with a depth just adequate to stop contamination electrons will minimize these distortions due to transmission and scatter of radiation and buildup of dose in the miniphantom. Use of a miniphantom constructed from water-equivalent material with a depth appropriate for the x-ray energy being measured is the preferred method for determining Sc for dosimetry in water.

Absorptiometry, Photon↗

Effect of image uncertainty on the dosimetry of trigeminal neuralgia irradiation.

OBJECTIVE: Our objective was to quantify the uncertainty in localization of the trigeminal nerve (TGN) with magnetic resonance imaging (MRI) and computed tomography (CT) and to determine the effect of this uncertainty on gamma-knife dose delivery. METHODS: An MR/CT test phantom with 9, 0.6-mm diameter, copper rings was devised. The absolute ring positions in stereotactic space were determined by the angiographic module of the LGP software. The standard deviation, sigma, in the difference between the absolute and MR-measured or CT-measured coordinates of the rings was determined. The trigeminal nerve in 52 previously treated patients was contoured and expanded by 1sigma and 2sigma margins to model the uncertainty in the location of the nerve. For gamma-knife treatment, a single isocenter was used and was located at the distal cisternal portion of the trigeminal nerve root. Irradiation methods included a 4-mm collimator, 90 Gy to isocenter and a 4&8-mm collimator, 70 Gy to isocenter. A patient outcome survey that sampled pain relief and morbidity was done. RESULTS: The MR coordinate sigma was 0.7 mm left-right, 0.8 mm anterior-posterior, and 0.6 mm superior-inferior, and the CT coordinate sigma was 0.4 mm left-right, 0.2 mm anterior-posterior, and 0.2 mm superior-inferior. A 45% higher dose line covered the TGN with the 4&8-mm method. No significant increase in pain reduction or morbidity occurred. CONCLUSIONS: The uncertainty of target location by MRI is more than twice that found in CT imaging. The 4&8-mm collimator method covers the trigeminal root cross section with a higher isodose line than does the 4-mm method. This higher dose did not significantly reduce pain or increase morbidity.

Humans↗

Dosimetric advantages of IMRT simultaneous integrated boost for high-risk prostate cancer.

PURPOSE: A sequential two-phase process, initial and boost irradiation, is the common practice for the radiotherapy management of high-risk prostate cancer. In this work, we explore the feasibility of using intensity modulated radiation therapy (IMRT) simultaneous integrated boost (SIB), a single-phase process, to simultaneously deliver high dose to the prostate and lower dose to the pelvic nodes. In addition, we introduce the concept of voxel-equivalent dose for the comparison of treatment plans. METHODS AND MATERIALS: The SIB is designed to deliver the same dose (e.g., 45 Gy, 25 x 1.8 Gy) as the conventional method to the pelvic nodes and to deliver higher doses to prostate in the same 25 fractions (i.e., hypofractionation). The equivalent uniform dose (EUD) was used to determine suitable SIB fractionations that deliver the biologically equivalent doses to prostate. For tumor, the EUD was estimated based on the linear quadratic (LQ) model. The most recent LQ parameters derived from clinical data for prostate cancer were used. The sensitivity of LQ parameters was evaluated. The EUD for normal tissue was computed based on the widely used Lyman model. To be able to consider biologic effectiveness spatially (e.g., voxel by voxel), we propose a new concept, termed the voxel-equivalent dose (VED). The calculation of VED was similar to that for EUD, except that it was done within a voxel. To demonstrate dosimetric feasibility and advantages of the proposed IMRT SIB, we have performed a retrospective planning study on selected patient cases using commercial IMRT and three-dimensional (3D) planning systems. Four treatment scenarios were considered: (1) the conventional 3D plan for initial whole-pelvic irradiation and subsequent conventional 3D boost plan for prostate gland, (2) the conventional 3D plan for initial whole-pelvic irradiation and subsequent IMRT boost plan for prostate, (3) IMRT plan for initial whole-pelvic irradiation and subsequent IMRT boost plan for prostate, and (4) IMRT SIB. EUDs and VED-based dose-volume histograms for prostate, pelvic nodes, small bowel, rectum, bladder, and other tissue for all 4 scenarios were compared. RESULTS: A series of equivalent hypofractionation regimens suitable for the IMRT SIB were obtained for high-risk prostate cancer. For example, the conventional treatment regimen of 42 x 1.8 Gy (EUD = 75.4 Gy) would be equivalent to a SIB regimen of 25 x 2.54 Gy. From the comparison of 3D VED dose distributions and dose-volume histograms between the SIB and the conventional two-phase irradiation, we found that the SIB offers better or equivalent dose conformity to prostate and pelvic nodes and better sparing to the critical structures. For example, for the 4 treatment scenarios with a prostate EUD of 75.4 Gy, the corresponding rectal EUDs are 67.1 (3D + 3D), 65.6 Gy (3D + IMRT), 63.7 Gy (IMRT + IMRT), and 62.0 Gy (SIB). CONCLUSIONS: A new IMRT simultaneous integrated boost strategy that irradiates prostate via hypofractionation while irradiating pelvic nodes with the conventional fractionation is proposed for high-risk prostate cancer. Compared to the conventional two-phase treatment, the proposed SIB technique offers potential advantages, including better sparing of critical structures, more efficient delivery, shorter treatment duration, and better biologic effectiveness.

Dose Fractionation, Radiation↗

Distortion of magnetic resonance images used in gamma knife radiosurgery treatment planning: implications for acoustic neuroma outcomes.

OBJECTIVE: To quantify the image distortion of our series of acoustic neuromas treated with gamma knife radiosurgery. STUDY DESIGN: Retrospective chart and digital radiographic file review with quantitative assessment of gamma knife treatment plans. SETTING: Tertiary referral center. PATIENTS: Patients undergoing gamma knife radiosurgery for the treatment of acoustic neuromas. INTERVENTION: Gamma knife radiosurgery. MAIN OUTCOME MEASURES: Gamma knife treatment plans containing magnetic resonance images were reviewed at each axial, sagittal, and coronal slice. The length of the greatest displacement of the treatment plan was measured and the volume of the treatment plan that fell outside of the internal auditory canal calculated. Known clinical measurements of audiometric, vestibular, facial, and trigeminal nerve functions were then compared with current measurements of tumor size. RESULTS: Twenty-two of the 23 patients had measurable image shifts on the axial images. The range of the image shift was 0 to 5.8 mm, with a mean shift of 1.92 +/- 1.29 mm (+/- standard deviation). Tumor volumes of the treatment plan that fell outside of the internal auditory canal ranged from 0 to 414 mm, with a mean of 90.5 mm. The mean percentage that fell outside of the internal auditory canal was 16.7% of total tumor volume (range, 2.4-77.6%). We could not draw any consistent correlations between degree of image shift and continued tumor growth or objective examination values. CONCLUSION: We have demonstrated a small but potentially significant shift in the treatment plan of gamma knife radiosurgery when based on magnetic resonance images. Although the image shift does not seem to affect the growth of the acoustic neuromas or auditory or facial nerve function, longer term follow-up is required to fully appreciate the true impact of this image shift.

Artifacts↗

A 2-D diode array and analysis software for verification of intensity modulated radiation therapy delivery.

An analysis is made of a two-dimensional array of diodes that can be used for measuring dose generated in a plane by a radiation beam. This measuring device is the MapCHECK Model 1175 (Sun Nuclear, Melbourne, FL). This device has 445 N-type diodes in a 22 x 22 cm2 2-D array with variable spacing. The entire array of diodes is easily calibrated to allow for measurements in absolute dose. For IMRT quality assurance, each beam is measured individually with the beam central axis oriented perpendicular to the plane of diodes. Software is available to do the analytical comparison of measurements versus dose distributions calculated by a treatment planning system. Comparison criteria of percent difference and distance-to-agreement are defined by the operator. Data are presented that show the diode array has linear response when beam fluence changes by over 300-fold, which is typical of the level of modulation in intensity modulated radiation therapy, IMRT, beams. A linear dependence is also shown for a 100-fold change in monitors units delivered. Methods for how this device can be used in the clinic for quality assurance of IMRT fields are described. Measurements of typical IMRT beams that are modulated by compensators and MLCs are presented with comparisons to treatment planning system dose calculations. A time analysis is done for typical IMRT quality assurance measurements. The setup, calibration, and analysis time for the 2-D diode array are on the order of 20 min, depending on numbers of fields. This is significantly less time than required to do similar analysis with radiographic film. The 2-D diode array is ideal for per-plan quality assurance after an IMRT system is fully commissioned.

Algorithms↗