Comment on "A method for dose determination in computerized tomography" by M. G. Yalcintas and O. Nalcioglu.
Explore the source record for details and available documents.
Biomedical subjects
Publications and source records attributed to R A Jucius.
Explore the source record for details and available documents.
Due to the x-ray field geometry, exposure measurements on computed tomographic (CT) scanners require modification of widely used measurement techniques. Ionization chambers have geometries which do not allow measurement of narrow rotating x-ray fields. However, by utilizing the partial volume effect, an ion chamber can be calibrated to measure narrow x-ray fields with a spatial resolution of down to 1 mm. The number of scans required to reconstruct a complete profile depends on the resolution required, and is typically on the order of 10 to 20 scans. This is accomplished using a series of attenuating sleeves over the ionization chamber. The chamber and sleeves are calibrated against a standard chamber using large field geometries. Comparisons with TLD measurements are shown to verify the accuracy of the technique.