Delayed pacemaker erosion due to electrode seal defects.

Over a 7-year period, follow-up data were available on 163 patients who underwent 209 pacemaker-related operations for initial insertion or revision of previously implanted units. During the follow-up period, 16 pacemaker generator units were remoaved from 16 pacemaker generator units were removed from 12 patients solely because of the development of local skin erosion. This represents a 7.7% incidence of pacemaker generator erosion. Tree units were removed within a 3-month interval following pacemaker implantation and were associated with positive bacterial cultures, theraby indicating an infectious cause. Findings at removal of the remaining 13 pacemaker generators included extensive fibrosis with chronic inflammation, fibroblastic proliferation, and granulation tissue, which was most prominent at the site of insertion of the pacemaker lead into the pacemaker generator unit. Granulation tissue frequently led from this area to the area of skin attenuation and erythema. Inspection of the pacemaker generator units and electrodes demonstrated fluid ingress and tissue ingrowth toward the electrode due to inadequate seal mechanisms. Inadequate seal design of the pacemaker generator and electrode junction leads to tissue ingrowth with fibroblastic proliferatiog chronic inflammation, and eventual pacemaker erosion.

Aged↗

Focus on: Newark Beth Israel Medical Center Biomedical Engineering Department.

This paper describes the Biomedical Engineering Department of Newark Beth Israel Hospital. This hospital's extensive involvement in research, heart transplantation, and special technologies has placed unusual demands on the Biomedical Engineering Department because of the large volume of routine and specialized equipment involved. The 12-person Department is responsible for servicing over 2700 pieces of equipment in 91 various hospital departments, and also offers its services to other local hospitals. Established in 1970, the Department uses a computerized biomedical equipment database program to manage data for repairs, preventive maintenance, and electrical safety.

Academic Medical Centers↗