PubMedDate not supplied
Lead levels in drinking water were measured by the standard U.S. Environmental Protection Agency (EPA)-approved atomic absorption spectroscopy-graphite furnace Method 239.2 and compared with determinations made with a newly available portable anodic stripping voltammetry (PASV) instrument. A standard curve was prepared at 2, 5, 10, 15, 20, 25, and 30 microg/L of lead. PASV instrument readings were lower than standard EPA method values, with a mean difference and standard error (SE) of the mean between the two of 1.538+/-0.588 microg/L (n = 7, p = 0.040, significant at the 95% confidence interval [CI]). First-flush drinking water samples collected from 144 water fountains/coolers were preserved with nitric acid. Total lead content was tested twice for 29 EPA method samples and 54 PASV instrument samples to determine the variation within each method; results were not significant at the 95% CI. Total lead content was determined for 144 samples by both methods. PASV instrument readings were lower than standard EPA method values (mean difference and SE of the mean 0.630+/-0.206 microg/L; n = 144, p = 0.0027, significant at the 95% CI). Mean and standard deviation of the 144 samples for the EPA method and the PASV instrument were 6.5+/-11.8 microg/L and 5.9+/-11.6 microg/L, respectively. Means were below the action level for lead of 15 ppb (microg/L), but some values were above the action level (18 [13%] using the EPA method; 20 [14%] using the PASV instrument). Retesting by EPA method showed two false positive PASV values. Results indicate that in some field situations the PASV instrument may prove useful due to its relatively low price, small size, ease of use, and quick readings.