PubMed Health⌕ Search

Biomedical subjects

R Quintal

Publications and source records attributed to R Quintal.

3 recordsLinked to original sources

Comparisons of PBDE composition and concentration in fish collected from the Detroit River, MI and Des Plaines River, IL.

Polybrominated diphenyl ethers (PBDEs) were identified in fish collected from the Detroit River, MI and Des Plaines Rivers, IL. In the Detroit River fish, carp and large mouth bass, the congener patterns were dominated by the 2,2',4,4'-tetrabromo (BDE-47) congener, however, in Des Plaines River carp the dominant isomers were the heptabromo congeners BDE-181 and BDE-183 and lesser amounts of another heptabromo congener, BDE-190, and two hexabromo congeners, BDE-154 and BDE-153. Three possible sources exist for these less-commonly identified PBDE congeners: (a) waste discharge from manufacturing or discarded products near the river, (b) public owned treatment work (POTW) effluents which constitute more than 75% of the flow in the Des Plaines River, (c) or formation of these congeners by debromination of in-place deposits of decabromodiphenyl ether. Average concentration totals (sum of concentrations for seven of the dominant PBDE congeners) were similar on a wet weight bases for the carp (5.39 ng/g wet weight) and large mouth bass (5.25 ng/g) in the Detroit River samples; however, the bass were significantly higher, p = 0.01, when compared on a lipid basis (bass--163 ng/g vs. carp--40.5 ng/g lipid weight). Some of the PBDE congeners were positively correlated with increasing lipid levels in both fish species. Average total PBDE concentrations in the carp from the Des Plaines River (12.48 ng/g wet weight) were significantly higher, p = 0.01, than in carp from the Detroit River. The residues were isolated using standard organochlorine methods for fish and analyzed using gas chromatography/mass spectrometry-negative chemical ionization methods.

Animals↗

Pseudo failure of sensing in patients with universal pacemakers and junctional rhythms.

Ventricular fibrillation has been only rarely observed as a complication of cardiac pacing after the advent of demand pacemakers. Automatic AV sequential pacing (DDD) may provide the setting for ventricular fibrillation in patients with junctional rhythms. In this report we present two patients with implanted DDD pacemakers in whom ventricular pacemaker spikes were seen occurring on top of the T wave during episodes of junctional rhythm. This apparent lack of sensing of QRS complexes does not represent pacemaker malfunction, but rather, is the result of physiologic lack of sensing (blanking) which occurs 56 to 100 msec. following the output of the atrial and ventricular channels. During junctional rhythm when the atrial spike occurs at the beginning of a QRS complex the ventricular channel is blanked and does not sense the intrinsic ventricular activity and thus, ventricular output occurs during repolarization. Increasing the maximum pacemaker rate and decreasing the AV delay will reduce the chance occurrence of this phenomenon.

Aged↗

Myopotential interference with a DDD pacemaker--report of a case.

This case report describes myopotential interference with an automatic AV sequential pacemaker (DDD) producing repetitive firing of the ventricular channel up to a rate of 160 and occasional suppression of the ventricular output. A silastic boot which covered the pacemaker did not prevent sensing of electromyopotentials. The problem was partially corrected by decreasing the sensitivity of the ventricular channel. The ventricular response to myopotentials sensed by the atrial channel was limited by decreasing the maximum programmed rate to 120/minute.

Action Potentials↗