Application of an electrochemical detector to the determination of procarbazine hydrochloride by high-performance liquid chromatography.

An amperometric flow-through detector with a carbon paste working electrode was utilized as a high-performance liquid chromatographic (HPLC) detector to determine procarbazine hydrochloride, an antineoplastic agent, in both buffer solution and biological fluids. The HPLC system included an amino-cyano stationary phase and an aqueous (pH 7)-methanolic mobile phase which enabled the separation of procarbazine from its only electroactive degradation product, N-isopropyl-alpha-(2-methylhydrazono)-p-toluamide. The electrochemical detector, with an approximate limit of detection of 2 ng procarbazine injected, was 20 times more sensitive to procarbazine than a typical UV detector. The low dead volume (1 microliter) and superior selectivity of the electrochemical detector enabled the HPLC determination of procarbazine in untreated human urine and plasma.

Chromatography, High Pressure Liquid↗

Automated constant-current coulometric assay system for ascorbic acid and sodium ascorbate.

The performance of an automated constant-current coulometric system for the assay of ascorbic acid and sodium ascorbate is described. After loading, it is capable of analyzing 25 samples and printing out the titer values with no operator attention for 2.5 hr. Under optimum conditions, ascertained by evaluating various electrochemical parameters, the accuracy and precision (95% ts) were found to be +/- 0.3%.