Models for studying the progression of hearing loss caused by noise.

A computer-controlled fixed-frequency Békésy audiometer has been used together with simplified mathematical models for hearing threshold measurements and analysis in 1,796 subjects from seven different noise environments with exposures varying from less than 80 dB(A) to greater than 95 dB(A). The methods of measurement and analysis have been used to determine hearing loss caused by noise for various age groups for men and women. It has also been possible to estimate the equivalent noise level to which workers are nowadays subjected under hearing protectors and the reduction in exposure level over the past 40 years. The results illustrate the effects of noise-reduction measures and the use of hearing protectors on the development of hearing loss suffered by workers in the actual Swedish industries.

Adult↗

Computerized audiometry--a facility to prevent hearing damages caused by noise.

A computerized audiometer has been developed to perform pure-tone manual or fixed-frequency Békésy audiometry. A number of statistical evaluation programs are part of the integrated audiometer. It means that even small but statistically certain changes in hearing thresholds can be revealed at an early stage.

Audiometry↗

High speed tympanometry and ipsilateral middle ear reflex measurements using a computerized impedance meter. A comparison with the results obtained by a conventional impedance meter.

By means of a new computerized impedance meter, middle ear pressure, tympanograms and ipsilaterally elicited middle ear reflex can be recorded and presented in 3 sec. Owing to the fast rate of air pressure change in the external ear canal it is not necessary to have a completely airtight seal around the ear probe when recording middle ear pressure, which saves time. Results obtained using the new device and those obtained by a conventional impedance bridge are compared.

Acoustic Impedance Tests↗

Comparison between hearing threshold measurements made with and without the aid of an ear speculum.

The hearing threshold levels of a small group of shipyard workers having differing degrees of hearing impairment were measured five times using the following audiometric techniques: fixed-tone Békésy audiometry with MX 41/AR cushions, fixed-tone Békésy audiometry with an ear speculum, and sweep-frequency Békésy audiometry with an ear speculum. From these data the mean standard deviations of the hearing threshold levels for the frequencies 0.5, 1, 2, 3, 4, 5, 6, 7 and 8 kHz were calculated, thus giving a measure of the accuracy of each method. The hearing threshold levels obtained by the three methods are compared, and the merits of data handling, expressing hearing threshold levels in pascals rather than decibels, are also discussed.