A study to determine whether cavitation occurs around dental ultrasonic scaling instruments.

The aim of this investigation was to determine if cavitation occurred around dental ultrasonic scalers and to estimate the amount of cavitation occurring. Three styles of tip (3 x TFI-10, 3 x TFI-3, 3 x TFI-1) were used, in conjunction with a Cavitron SPS ultrasonic generator (Dentsply, USA), to insonate terephthalic acid solution. The hydroxyl radical, [*OH], concentration, produced due to cavitation from the scaler tips, was monitored by fluorescence spectroscopy. Cavitational activity was enhanced at higher power settings and at longer operating times. The tip dimensions and geometry as well as the generator power setting are both important factors that affect the production of cavitation.

Dental Scaling↗

A new insight into the oscillation characteristics of endosonic files used in dentistry.

The aim of this study was to assess the oscillation characteristics of unconstrained endosonic files using a scanning laser vibrometer (SLV). Factors investigated included file vibration frequency and node/antinode location as well as the variation in file displacement amplitude due to increasing generator power setting. A 30 kHz Mini Piezon generator (Electro-Medical Systems, Switzerland) was used in conjunction with a #15 and #35 K-file. Each file was fixed in position with the long axis of the file perpendicular to the SLV camera head. The laser from the SLV was scanned over the length of the oscillating file for generator power settings 1 to 5 (minimum to half power). Measurements were repeated ten times. The fundamental vibration frequency for both files was 27.50 kHz. Scans of each file showed the positions of nodes/anti-nodes along the file length. The #15 file demonstrated no significant variation in its mean maximum displacement amplitude with increasing generator power, except at power setting 5, where a decrease in displacement amplitude was observed. The #35 file showed a general increase in mean maximum displacement amplitude with increasing power setting, except at power setting 4 where a 65% decrease in displacement amplitude occurred. In conclusion, scanning laser vibrometry is an effective method for assessing endosonic file vibration characteristics. The SLV was able to demonstrate that (unloaded) file vibration displacement amplitude does not increase linearly with increasing generator power. Further work is being performed on a greater variety of files and generators. Vibration characteristics of files under various loads and varying degrees of constraint should also be investigated.

Dentistry↗

Does cavitation occur around powered toothbrushes?

AIMS/OBJECTIVES: The aim of this investigation was to determine if cavitational activity occurred around powered toothbrushes using a chemical dosimeter system based on terephthalic acid (TA). MATERIAL AND METHODS: Five powered toothbrushes were used in this investigation: Braun Plaque Remover D8, Braun Oral-B 3D, Philips Jordan Sensiflex HX2520, Sonicare PS-1 and Sonicare Elite HX 7351/02. Each brush head was inserted into a conical flask containing 50 cm3 of aqueous TA solution. Brushes were operated for 10 and 20 min and a cuvette of the solution was placed in a fluorescence spectrometer (Perkin Elmer 3000). The fluorescence emitted at wavelength 425 nm, which is proportional to *OH radical concentration, was monitored. RESULTS: Any cavitational activity that may have been produced by the powered toothbrushes was below the limit of detection of the system (<10(-8) M) for the timescales investigated. CONCLUSIONS: This work has demonstrated that cavitational activity does not occur around powered toothbrushes. Operating the toothbrushes for periods up to 20 min resulted in no cavitational activity being detected.

Equipment Design↗

Displacement amplitude of ultrasonic scaler inserts.

OBJECTIVES: Scanning laser vibrometry is a noninvasive method of measuring the velocity, displacement amplitude and oscillation frequency of vibrating objects. The purpose of this study was to assess, using a scanning laser vibrometer (SLV), the performance of different designs of commercially available ultrasonic scaler generators by measuring the oscillatory characteristics of various scaler tips. METHODS: Four ultrasonic generators were tested (Cavitron SPS and Cavitron Select (Dentsply, York, PA, USA) and Piezon Master 400 and Mini Piezon (Electro-Medical Systems, Switzerland)) with various designs of scaler tip. The tips were positioned with their anterior surface perpendicular to the direction of the laser. A graduated scale, placed over the manufacturer's power dial, enabled incremental power setting selection. For each power setting, the laser beam from the SLV was scanned over the surface of the oscillating tip. RESULTS: The ranges of longitudinal displacement amplitudes (in micrometres) were as follows: Mini Piezon (P-tip): 12.90+/-1.44 to 44.03+/-7.80; Piezon Master 400 (P-tip): 16.02+/-2.66 to 35.85+/-5.29; Cavitron SPS (TFI-10 tip): 7.81+/-0.51 to 29.70+/-1.12; Cavitron Select (TFI-10 tip): 13.13+/-1.44 to 33.77+/-4.27; Cavitron SPS (TFI-3 tip): 5.50+/-0.46 to 31.35+/-3.62. CONCLUSIONS: This study shows that there are differences between commercially available generators and that tip movement varies between tips of the same style as well as between the generator and tip design. Users of ultrasonic scalers should be made aware of this inherent variability that may influence clinical procedures.

Dental High-Speed Technique↗

Ultrasonic scaler tip performance under various load conditions.

OBJECTIVES: The aim of this investigation was to assess the effect of loading on ultrasonic scaler tip displacement amplitude, using a scanning laser vibrometer. MATERIAL AND METHODS: Four generators were selected for this investigation including: Cavitron Select (Dentsply), Cavitron SPS (Dentsply), Piezon Master 400 (EMS) and Mini Piezon (EMS). The insert designs used with these generators included the TFI-10 and TFI-3 (with the Dentsply generators) and P-tip (with the EMS generators). The insert tips were contacted against tooth surfaces with loads of 0.25, 0.5 and 1.0 N. The scanning laser vibrometer was then used to scan the vibrating scalers at various generator power settings. RESULTS: The tips generally showed an increase in displacement amplitude with increasing power setting, although this increase was often nonlinear. Displacement amplitudes of like tips were generally found to be significantly different from each other at all loads. For individual inserts, loading caused a significant difference in tip response from the unloaded situation and also between loads. CONCLUSIONS: This investigation has demonstrated that there is a high variability associated with dental ultrasonic scaler inserts. Clinicians and researchers should be aware of this variability as it may affect both clinical procedures and future research.

Analysis of Variance↗

Vibration characteristics of ultrasonic scalers assessed with scanning laser vibrometry.

OBJECTIVES: Scanning laser vibrometry is a non-invasive method of accurately measuring the vibratory characteristics of oscillating objects. The aim of this study was to observe, using a scanning laser vibrometer (SLV), the vibration patterns of dental ultrasonic scaler tips and to assess the effects of water flow rate and power setting on these patterns whilst operating the tips in an unloaded environment. METHODS: A 30kHz ultrasonic scaler (TFI-10, Dentsply) was fixed in position and a laser beam from the SLV was focused onto the tip. The laser, guided by a virtual measurement grid, was scanned over the oscillating tip surface. Scans were taken with the laser beam perpendicular to the long axis of the front face of the tip. RESULTS: Oscillation frequencies and the displacement amplitude at the unconstrained end of the tip were measured for various power/water settings. Vibration nodal positions were recorded for the various settings and were found to occur approximately 4mm from the free end of the tip. At low and medium power settings, tip displacement amplitude was reduced by increased water flow. At high power settings, combined with a high flow rate, the water leaves the body of the instrument as a jet. This left the tip relatively unconstrained, allowing it to oscillate at increased displacement amplitudes. CONCLUSIONS: This study shows that the SLV is able to accurately characterise the movement of oscillating ultrasonic scaler tips. The tips are affected by power setting and water flow rates.

Analysis of Variance↗