A preliminary comparative treatment planning study for radiotherapy of age-related maculopathy.

PURPOSE: We present a comparative planning of different approaches for external radiotherapy in age-related maculopathies. MATERIALS AND METHODS: Calculated dose distributions and dose-volume histograms for (a) bilateral irradiation with 6 MV photons, (b) a single lateral-oblique beam using either photons, electrons or protons and (c) an anterior circular proton beam. RESULTS: For lateral photon or electron beams the dose to the lens is usually lower than 10% of the dose to the macula. The entrance doses for bilateral photon beams are about 50% which increase up to 100% at the orbital bone. About 5 mm of optic nerves are irradiated at the maximal dose while the optic chiasma is spared. A single photon beam gives 50% of the dose to the fellow eye. The electron beam spares the fellow eye but gives a rather inhomogeneous dose to the target volume. For a lateral proton beam, 4 mm of optic nerve receives 90% of the dose, the skin dose is at least 70% of the dose to the macula and the lens and the fellow eye are spared. An anterior proton beam gives 90% of the dose to 1 mm of optic nerve and the 50% isodose approaches the periphery of the lens. CONCLUSION: Doses to the critical structures can be dramatically diminished for all the techniques by reducing the beam size, but only if very precise set-up techniques are used. Proton beams are an attractive solution, but the impact of such a choice on the use of proton facilities and on the national health system should be carefully evaluated, as well as the risk of radio-induced secondary neoplasias.

Humans

Effectiveness of protons and argon ions in initiating lipid peroxidation in low-density lipoproteins.

In this study, human low-density lipoprotein (LDL) vesicles were irradiated with 73 MeV protons (LET of 1 keV/microm) and 11.4 MeV/nucleon argon ions (LET of 1.52 MeV/pm) and the effectiveness of charged particles in initiating peroxidation of LDLs was investigated. The LDL suspension (6 g/l) was exposed to protons and to argon ions in a dose range of 24 Gy to 2.4 kGy. Irradiations were carried out at the synchrocyclotron at the CPO and at the UNILAC of the GSI. After irradiation three chemical assays were used to study the progression of peroxidation of LDLs: the formation of conjugated dienes, the formation of thiobarbituric acid-reactive substances (TBARS) and the increase in the relative electrophoretic mobility of the LDLs. The results were compared with those obtained after gamma irradiation. For protons the yields of the peroxidation products were 10 times lower than after gamma irradiation. However, for doses below 200 Gy, protons appeared to be more effective than gamma rays in damaging the protein moiety, as deduced from the observed increase in the relative electrophoretic mobility of the LDLs. The irradiation with argon ions led to a negligible formation of peroxidation products, but an increase in the relative electrophoretic mobility of the LDLs was observed. The results are indicative of a lower yield of lipid peroxidation after irradiation with high-LET particles. In contrast, protons and argon ions appear to be more effective in inducing bulk protein and phospholipid damage than gamma rays.

Argon

Proton dosimetry comparison involving ionometry and calorimetry.

A comparison of the absorbed dose to tissue determined by various ionization chambers, Faraday cups, and an A-150 plastic calorimeter was performed in the 200 MeV proton beam of Orsay, France. Four European proton-therapy centers (Clatterbridge, UK, Louvain la Neuve, Belgium, and Nice and Orsay, France) participated in the comparison. An agreement of better than 1% was observed in the absorbed dose to A-150 measured with the different chambers of the participating groups. The mean ratio of the absorbed dose to A-150 determined with the calorimeter to that determined by the different ionization chambers in the different irradiation conditions was found to be 0.952 +/- 0.007 [1 standard deviation (SD)] according to the code of practice used by all the participating centers, based on Janni's tables of stopping powers and a value of 35.2 J/Coulomb for (W(air)/e)p. A better agreement in the mean ratio calorimeter/chamber, 0.985 +/- 0.007 (1 SD) is observed when using the proton stopping power ratio values recently published by the International Commission on Radiation Units and Measurements in Report no. 49. The mean ratio of these doses determined in accordance with the American Association of Physicists in Medicine protocol and using the new recommended stopping power tables becomes 1.002 +/- 0.007 (1 SD). Two Faraday cups agree in measured charge to within 0.8%; however, the calculation of dose is underestimated by up to 17%; compared with ion chamber measurements and seems to be very sensitive to measurement conditions, particularly to the distance to the collimator.

Calibration

[Initial results of proton therapy in choroidal melanoma at the d'Orsey Center for Proton Therapy; the first 464 cases].

PURPOSE: Retrospective analysis of the treatment of choroidal melanoma with protontherapy at the Centre de protonthérapie d'Orsay, France. PATIENTS AND METHODS: Between September 1991 and September 1995, 612 patients presenting with choroidal melanoma were treated by protontherapy in Orsay. Following initial management of the first 464 patients, results were analyzed, as were results after a 1-year follow-up for 305 patients, a 2-year follow-up for 169 patients, and a 3-year follow-up for 59 patients. RESULTS: Univariate analysis showed that the actuarial local recurrence rate was 5%, the 3-year survival rate 88%, and the overall metastasic rate 5%. The initial tumor volume was the most significant predictive factor for visual results and metastases. Multivariate analysis revealed that visual results were significantly related to the initial tumor volume, initial retinal detachment, and total dose delivered to the optic nerve and macula. CONCLUSION: Protontherapy of choroidal melanoma allows in most cases conservation of the eye without modification of survival. Visual results mainly depend on the site and size of the tumor.

Choroid Neoplasms

Proton dosimetry intercomparison.

BACKGROUND AND PURPOSE: Methods for determining absorbed dose in clinical proton beams are based on dosimetry protocols provided by the AAPM and the ECHED. Both groups recommend the use of air-filled ionization chambers calibrated in terms of exposure or air kerma in a 60Co beam when a calorimeter or Faraday cup dosimeter is not available. The set of input data used in the AAPM and the ECHED protocols, especially proton stopping powers and w-value is different. In order to verify inter-institutional uniformity of proton beam calibration, the AAPM and the ECHED recommend periodic dosimetry intercomparisons. In this paper we report the results of an international proton dosimetry intercomparison which was held at Loma Linda University Medical Center. The goal of the intercomparison was two-fold: first, to estimate the consistency of absorbed dose delivered to patients among the participating facilities, and second, to evaluate the differences in absorbed dose determination due to differences in 60Co-based ionization chamber calibration protocols. MATERIALS AND METHODS: Thirteen institutions participated in an international proton dosimetry intercomparison. The measurements were performed in a 15-cm square field at a depth of 10 cm in both an unmodulated beam (nominal accelerator energy of 250 MeV) and a 6-cm modulated beam (nominal accelerator energy of 155 MeV), and also in a circular field of diameter 2.6 cm at a depth of 1.14 cm in a beam with 2.4 cm modulation (nominal accelerator energy of 100 MeV). RESULTS: The results of the intercomparison have shown that using ionization chambers with 60Co calibration factors traceable to standard laboratories, and institution-specific conversion factors and dose protocols, the absorbed dose specified to the patient would fall within 3% of the mean value. A single measurement using an ionization chamber with a proton chamber factor determined with a Faraday cup calibration differed from the mean by 8%. CONCLUSION: The adoption of a single ionization chamber dosimetry protocol and uniform conversion factors will establish agreement on proton absorbed dose to approximately 1.5%, consistent with that which has been observed in high-energy photon and electron dosimetry.

Calibration

An experimental approach to the design of a scattering system for a proton therapy beam line dedicated to ophthalmological applications.

PURPOSE: The development of a scattering system for a proton therapy beam line dedicated to ophthalmological applications. METHODS AND MATERIALS: A protontherapy beam line has been developed for the treatment of uveal melanoma at the Orsay synchrocyclotron. The original 200 MeV proton beam is degraded to 76 MeV and the final beam characteristics (range, modulation, flatness, collimation) are obtained with beam modifiers in the treatment room. A passive scattering system is used to obtain a uniform dose distribution in the beam cross-section throughout 30 mm in diameter, with minimal losses in energy and dose rate. We have used an experimental approach for the scattering study. RESULTS: An elliptical ring shaped from 0.1-mm thick lead is the solution we have adopted for the scattering system. For a modulated beam, a flatness of 1% is obtained on transverse profiles. The energy loss introduced by this scatterer is only 0.5 MeV, with no appreciable change in the range over the treatment field. For an unmodulated beam, 21% of intensity is lost when the scatterer is used. The distal and the lateral dose fall-off (90-10%) for a modulated beam are 2.6 mm. These last values are independent of the range and the modulation currently used for the ophthalmic applications. CONCLUSION: A specific passive scattering system can be adapted to a particular beam emittance. A systematic experimental approach can easily be undertaken to obtain the scatterer adapted for small irradiation fields in proton therapy.

Equipment Design