PubMed Health⌕ Search

Biomedical subjects

S Goddon

Publications and source records attributed to S Goddon.

4 recordsLinked to original sources

Optimal mean airway pressure during high-frequency oscillation: predicted by the pressure-volume curve.

BACKGROUND: A number of groups have recommended setting positive end-expiratory pressure during conventional mechanical ventilation in adults at 2 cm H2O above the lower corner pressure (P(CL)) of the inspiratory pressure-volume (P-V) curve of the respiratory system. No equivalent recommendations for the setting of the mean airway pressure (Paw) during high-frequency oscillation (HFO) exist. The authors questioned if the Paw resulting in the best oxygenation without hemodynamic compromise during HFO is related to the static P-V curve in a large animal model of acute respiratory distress syndrome. METHODS: Saline lung lavage was performed in seven sheep (28+/-5 kg, mean +/- SD) until the arterial oxygen partial pressure/fraction of inspired oxygen ratio decreased to 85+/-27 mmHg at a positive end-expiratory pressure of 5 cm H2O (initial injury). The PCL (20+/-1 cm H2O) on the inflation limb and the point of maximum curvature change (PMC; 26+/-1 cm H2O) on the deflation limb of the static P-V curve were determined. The sheep were subjected to four 1-h cycles of HFO at different levels of Paw (P(CL) + 2, + 6, + 10, + 14 cm H2O), applied in random order. Each cycle was preceded by a recruitment maneuver at a sustained Paw of 50 cm H2O for 60 s. RESULTS: High-frequency oscillation with a Paw of 6 cm H2O above P(CL) (P(CL) + 6) resulted in a significant improvement in oxygenation (P < 0.01 vs. initial injury). No further improvement in oxygenation was observed with higher Paw, but cardiac output decreased, pulmonary vascular resistance increased, and oxygen delivery decreased at Paw greater than P(CL) + 6. The PMC on the deflation limb of the P-V curve was equal to the P(CL) + 6 (r = 0.77, P < 0.05). CONCLUSION: In this model of acute respiratory distress syndrome, optimal Paw during HFO is equal to P(CL) + 6, which correlates with the PMC.

Animals↗

Repetitive high-pressure recruitment maneuvers required to maximally recruit lung in a sheep model of acute respiratory distress syndrome.

OBJECTIVE: To compare the effects of two different recruitment maneuvers repeated multiple times on gas exchange lung injury, hemodynamic, and lung mechanics. DESIGN: Randomized prospective comparison. SETTINGS: Animal research laboratory. SUBJECT: Nineteen fasted Hampshire sheep. INTERVENTIONS: In 15 27-kg sheep with saline lavage lung injury, we compared the effects of two recruitment maneuvers: 40 cm H2O continuous positive airway pressure for 60 secs and 40 cm H2O positive end-expiratory pressure with 20 cm H2O pressure control, rate 10 breaths/min, inspiratory to expiratory ratio 1:1 for 2 mins. Each recruitment maneuver was repeated four times, every 30 mins after a 30-sec ventilator disconnection. An additional group received no recruitment maneuvers. Animals were assigned randomly to the three groups and ventilated with 20 cm H2O positive end-expiratory pressure, pressure control 15 cm H2O, rate 20 breaths/min, inspiratory to expiratory ratio 1:1, and Fio2 1.0 between recruitment maneuver periods. MEASUREMENTS AND MAIN RESULTS: Significant and marked increases in Pao2 were observed in the pressure control recruitment maneuver group but only after the second recruitment maneuver. In both the control group and continuous positive airway pressure groups, Pao2 did not significantly increase after any recruitment maneuver compared with baseline injury. There was a significant decrease in cardiac output immediately after some continuous positive airway pressure recruitment maneuvers and a significant increase in mean pulmonary artery pressure in both continuous positive airway pressure and pressure control groups immediately after recruitment maneuvers, but these changes resolved within 10 mins. There were no marked histologic differences between groups and no volutrauma. CONCLUSION: In this model, maximal lung recruitment was obtained with 40 cm H2O positive end-expiratory pressure and 20 cm H2O pressure control applied repetitively every 30 mins for 2 mins without physiologic or histologic harm. Multiple recruitment maneuvers in some animals were required for maximum effect.

Analysis of Variance↗

Partial liquid ventilation ventilates better than gas ventilation.

Partial liquid ventilation (PLV) improves oxygenation in several models of lung injury. However, PLV has only been compared with conventional gas ventilation (GV) with low PEEP. Both PLV and GV can markedly improve oxygenation when PEEP is set above the lower corner pressure (Plc) on the inspiratory pressure-volume (P-V) curve of the total respiratory system. We questioned if the use of PEEP set above the Plc during PLV and GV would result in similar gas exchange. Lung injury was induced in 12 sheep by saline lavage before randomization to PLV (n = 6) or GV (n = 6). Animals in the PLV group were filled with perflubron (22 ml/kg) until a meniscus at the teeth was observed. Both groups were then ventilated with pressure control (FI(O(2)), 1.0; rate, 20/min; I:E, 1:1) and PEEP (1 cm H(2)O above the Plc on the inspiratory P-V curve). Peak inspiratory pressure (PIP) was limited to 35 cm H(2)O. Animals were ventilated for 5 h and then killed for histologic examinations. All 12 animals survived the 5-h ventilation period. After increasing PEEP above Plc, Pa(O(2)) increased significantly (p < 0.01) in both the GV and the PLV groups, but it did not differ significantly between groups (p = 0.86) at any time during the experiment. Pa(CO(2)) and VD/VT in GV increased markedly throughout the experiment after increasing PEEP (p < 0.001), but there was no significant change in Pa(CO(2)) in PLV (p = 0.13). Mean arterial blood pressure, mean pulmonary artery pressure, pulmonary artery occlusion pressure, and central venous pressure, increased and SVR decreased in GV (p < 0.05). The extent and the severity of lung injury in the dependent regions was greater in the GV group (p < 0.05). Both PLV and GV improved oxygenation, but PLV resulted in better ventilation than GV while preserving lung structure when PEEP was set 1 cm H(2)O above the Plc and PIP limited to 35 cm H(2)O.

Animals↗

The relationship between gas delivery patterns and the lower inflection point of the pressure-volume curve during partial liquid ventilation.

STUDY QUESTION: To determine whether a positive end-expiratory pressure (PEEP) level equivalent to the lower inflection point (LIP) could be identified by evaluation of the airway pressure, flow (f1. gif" BORDER="0">), and volume vs time waveforms during partial liquid ventilation (PLV). DESIGN: Prospective application of PEEP during PLV in a healthy animal model. SETTING: University hospital animal laboratory. PARTICIPANTS: Five healthy sheep weighing 30 kg each. INTERVENTIONS: The sequential application of 0 to 20 cm H(2)O PEEP in 2.5-cm H(2)O steps during PLV with both pressure and volume ventilation. MEASUREMENTS: Analysis of the pressure, volume, and f1. gif" BORDER="0"> waveforms as PEEP is sequentially increased. RESULTS: At 0 cm H(2)O PEEP, VT was markedly reduced compared with PEEP VT at > or = 7.5 cm H(2)O (p < 0.05) in pressure control ventilation (PCV), and peak inspiratory pressure minus PEEP was markedly increased compared with PEEP at > or = 5.0 cm H(2)O (p < 0.05) in volume control ventilation. At 10 cm H(2)O PEEP, all waveforms began to stabilize, and no significant differences in any variable assessed were measured at > 12.5 cm H(2)O PEEP. CONCLUSIONS: The application of PEEP during PLV markedly alters airway waveforms. Low PEEP decreases VT in PCV and increases airway pressure in VCV. The PEEP level equal to the LIP during PLV can be grossly estimated from airway waveforms. PEEP at > or = 10 cm H(2)O is needed to normalize gas delivery to functional residual capacity in the uninjured lung that is partially filled with perfluorocarbon.

Air Pressure↗